CHAPTER 2
Principles of Implant Design (The Paradigm):
Design Parameters
2.1 Functional Performance of the Device (and Attachment to Tissue)
2.1.1 Mechanical
2.1.1.1Strength and Modulus of Elasticity:
Modulus Matching
2.1.1.2 Tribol ogy
2.1.2.3 Kinematics
2.1.2 Chemical
2.1.2.1 Drug Delivery/Controlled Release System
2.1.2.2 Matrices to Facilitate Tissue Regneration
2.1.3 Attachment Vehicle
2.1.3.1 Grouting Agents
2.1.3.2 Topography/Porosity
2.1.3.3 Surfaces/Coatings for Chemical Bonding of Tissue
2.2 Effects of the Implant on the Body ("Biocompatibility")
2.2.1 Local Effects
2.2.2 Systemic Effects
2.3 Effects of the Body on the Implant (Degradation of the Device)
2.3.1 Corrosion of Metals
2.3.2 Degradation of N onabsorbable Polymers
2.3.3 Degradation of Absorbable Polymers
Paradigm for Implant Design
Function of the Device
Mechanical
Chemical
Electrical
Thermal
Optical
Effects of the
Device on the
Body
Mechanical
Interpenetrating
(Tension, Shear, Compress.)
-Chain entanglement
-ECM in interconnecting
pores (porosity)
Interdigitating
(Shear, Compress.)
-ECM in surface
irregularities (rugosity)
Chemical
Protein adsorption/
binding
Mineral deposition
Nonreactive
("Inert")
Hydrophobic
Hydrogel
Effects of Body on Implant
Resistence to Degradation?
Yes(Permanent); No(Absorbable)
Mechanical
-Fracture (Strength)
-Wear (Abrasive, Fatigue)
Chemical
-Wear(Adhesive)
-Corrosion
-Oxidation
-Hydrolysis
-Enzymolysis
-Dissolution/precipitation
Bioadhesion?
(i.e.,Tissue Bonding?)
(Surface)
Yes No
Surface Bulk
Stress Shielding
2.1 IMPLANT DESIGN CRITE RIA: FUNCTION/PERFORMANCE OF THE DEVICE
A. Mechanical
1. Kinematics, mechanics (strength and modulus), and tribology of total joint
replacement prostheses.
2. Pumping devices (e.g ., artificial heart).
B. Chemical
1. Dissolution characteristics (e.g ., drug delivery systems)
2. Sensing characteris tics (e.g ., "intelligent" drug delivery systems)
C. Optical
Optical characteristics of implants for the eye (e.g ., intraocular lens) and transparent
wound covering materials
D. Electrical
Characteristics for sensing and sending electrical si gnals (e.g ., cardiac pacemaker
electrode tips)
E. Thermal
2.2 IMPLANT DESIGN CRITE RIA: EFFECTS OF THE IMPLANT ON THE BODY
(BIOCOMPATIBILITY/SA FETY)
A. Chemical
1. Molecules/Ions Released (Toxicity)
a. Direct effects of molecules/ions on cells
1. Co, Cr, Ni, Si, AL, V
2. Polymer fragments?
b. Binding of molecules/ions to proteins to form complexes that elicit adverse
biological response
1. Metal ion binding to existing antibodies to elicit immune response (these
metal ions act as haptens)
2. Metal ion- protein complexes that facilitate transport of metal ion through the
cell membrane
2. Alteration of Adsorbed Macromolecules Causing them to have an Adverse Effect
a. Change in conformation of adso rbed proteins (e.g ., on hydrophobic
surfaces) causing them to be immunogenic
b. Cleavage/fragmentation of adsorbed proteins such as complement molecules
thus activating the alternative pathway of the immune response
B. Mechanical: Alteration on Strains in Surrounding Tissue ("Modulus Mismatch")
1. "Hard Tissue"
High modulus (stiff) implants in bone result in an alteration of the distribution of
strain in surrounding bone (i.e ., stress transfer is altered). Bone in areas in which
strain has been reduced decreases in mass (i.e., becomes more porous- atrophies
osteopenia) due to "stress shielding." Bone in areas of increased stress can increase
in thickness and density (i.e ., less porous) and thereby hypertrophy.
2. "S oft Tissue"
Elastic behavior (e.g ., stiffness) of artificial blood vessels affects their performance
with respect to endothelialization and integrity of the anastomosis.
C. Electrical
1. Electrically conducting implants might short out "strain generated potentials" in
surrounding tissues (e.g., bone)?
2. Electrical currents produced by the device (e.g ., pacemakers) could adversely affect
cells.
D. Thermal
1. Heat resulting from exothermic polymerization reactions (e.g., PMMA- "bone
cement") can cause tissue necrosis.
2. Thermal conductivity and heat capacity ( i.e ., thermal diffusivity) could affect how
heat generated by implants (e.g ., functional heat from artificial joints) is dissipated?
2.3 IMPLANT DESIGN CRITE RIA: EFFECTS OF THE BODY ON THE IMPLANT
(DEGRADATION)
A. Chemical
1. Corrosion of metals: metal ion release due to an anodic (reduction) reaction
a. Pitting and crevice corrosion and "concentration cell" effect at sites of depleted
oxygen
b. Ga lvanic corrosion due to contact of dissimilar metal. The more reactive metal
(in the Galvanic series) becomes the anode.
c. Stress corrosion due to accelerated metal ion release at a crack tip where the
strain is high.
d. The oxide "passiv ation" layer reduces potential for corrosion.
e. Corrosion facilitates cracks initiation and thereby weakens the device.
f. Ranked according to their potential for corrosion:
Stainless steel > Co- Cr alloy > Ti alloy.
2. Oxidation of polymers
Oxidation of polyethylene results in chain scission and a reduction in the average
molecular weight. This causes increases in the density, modulus of elasticity, and
percent crystallinity. Oxidation can be determined by detecting the carbonyl groups
that are formed.
3. Hydrolysis of polymers
Ester linkages (e.g ., polylactic and polyglycolic acid) are attacked by water leading
to chain scission.
4. Water absorption
Water absorption can lead to an alteration in the mechanical pro perties of certain
hydrophilic thermoplastic polymers (e.g ., polysulfone).
5. Lipid absorption
Absorption of lipid by certain hydrophobic polymers (e.g ., polydimethyl siloxane).
6. Dissolution
The effect of water and pH in dissolving certain substances (e.g ., calcium
phosphates)
7. Precipitation
Deposition of calcium salts (calcification).
8. Enzymolysis
Natural polymers (e.g ., collagen) used as an implant materials can undergo
degradation as a result of the action of enz ymes (e.g ., collagenase).
B. Mechanical: Mechanical loading applied by the body can lead to wear (erosion)
and fatigue fracture of the device.
1. Fatigue testing of implants.
2. Wear due to rubbing of tissue ( viz ., bone) against the device.