Neuman, M.R. “Biomedical Sensors”
The Electrical Engineering Handbook
Ed. Richard C. Dorf
Boca Raton: CRC Press LLC, 2000
114
Biomedical Sensors
114.1 Introduction
114.2 Physical Sensors
114.3 Chemical Sensors
114.4 Bioanalytical Sensors
114.5 Applications
114.6 Summary
114.1 Introduction
Any instrumentation system can be described as having three fundamental components: a sensor, a signal
processor, and a display and/or storage device. Although all these components of the instrumentation system
are important, the sensor serves a special function in that it interfaces the instrument with the system being
measured. In the case of biomedical instrumentation a biomedical sensor (which in some cases may be referred
to as a biosensor) is the interface between the electronic instrument and the biologic system. There are some
general concerns that are very important for any sensor in an instrumentation system regarding its ability to
effectively carry out the interface function. These concerns are especially important for biomedical sensors,
since the sensor can affect the system being measured and the system can affect the sensor performance. Sensors
must be designed so that they minimize their interaction with the biologic host. It is important that the presence
of the sensor does not affect the variable being measured in the vicinity of the sensor as a result of the interaction
between the sensor and the biologic system. If the sensor is placed in a living organism, that organism will
probably recognize the sensor as a foreign body and react to it. This may in fact change the quantity being
sensed in the vicinity of the sensor so that the measurement reflects the foreign body reaction rather than a
central characteristic of the host.
Similarly, the biological system can affect the performance of the sensor. The foreign body reaction might
cause the host to attempt to break down the materials of the sensor as a way to remove it. This may, in fact,
degrade the sensor package so that the sensor can no longer perform in an adequate manner. Even if the foreign
body reaction is not strong enough to affect the measurement, just the fact that the sensor is placed in a warm,
aqueous environment may cause water to eventually invade the package and degrade the function of the sensor.
Finally, as will be described below, sensors that are implanted in the body are not accessible for calibration.
Thus, such sensors must be extremely stable so that frequent calibrations are not necessary.
Biomedical sensors can be classified according to how they are used with respect to the biologic system.
Table 114.1 shows that sensors can range from noninvasive to invasive as far as the biologic host is concerned.
The most noninvasive of biomedical sensors do not even contact the biological system being measured. Sensors
of radiant heat or sound energy coming from an organism are examples of noncontacting sensors. Noninvasive
sensors can also be placed on the body surface. Skin surface thermometers, biopotential electrodes, and strain
gauges placed on the skin are examples of noninvasive sensors. Indwelling sensors are those which can be placed
into a natural body cavity that communicates with the outside. These are sometimes referred to as minimally
invasive sensors and include such familiar sensors as oral-rectal thermometers, intrauterine pressure transduc-
ers, and stomach pH sensors. The most invasive sensors are those that need to be surgically placed and that
Michael R. Neuman
Case Western Reserve University
? 2000 by CRC Press LLC
? 2000 by CRC Press LLC
C
ARDIAC
M
ONITOR
he basic method of assessing heart function is thermodilution, a procedure that involves insertion
of a catheter into the pulmonary artery and is demanding in terms of cost, equipment, and skilled
personnel time. For monitoring astronauts in flight, NASA needed a system that was non-invasive
and considerably less complex.
In 1965, Johnson Space Center contracted with the University of Minnesota to explore the then-known
but little-developed concept of impedance cardiography (ICG) as a means of astronaut monitoring. A
five-year program led to the development of the Minnesota Impedance Cardiograph (MIC), an electronic
system for measuring impedance changes across the thorax that would be reflective of cardiac function
and blood flow from the heart’s left ventricle into the aorta. ICG clearly had broad potential for hospital
applications but further development and refinement was needed. A number of research institutions and
medical equipment companies launched development of their own ICGs, using MIC technology as a
departure point. Among them were Renaissance Technologies, Inc., Newtown, Pennsylvania, and Drexel
University of Philadelphia, who jointly developed the IQ System. The system provides a simple, repeatable,
non-invasive way of assessing cardiac function at dramatically reduced cost. The IQ System is in wide
use in hospital intensive care units, emergency rooms, operating rooms, and laboratories in the U.S. and
abroad.
IQ has two basic elements: the non-invasive, disposable patient interface known as IQ-Connect and
the touch screen monitor, which calculates and displays cardiac output values and trends. The hardware
design of the original MIC was retained but IQ has advanced automated software that features the signal
processing technology known as TFD (Time Frequency Distribution). TFD provides three-dimensional
distribution of the hemodynamic signals being measured, enabling visualization of the changes in power,
frequency, and time. This clinically proven capability allows IQ to measure all cardiac events without
using estimation techniques required in some earlier systems. (Courtesy of National Aeronautics and
Space Administration.)
The IQ-Connect interface electronically measures impedance changes across the thorax to reflect heart function.
(Photo courtesy of National Aeronautics and Space Administration.)
T
involve some tissue damage associated with their installation. For example, a needle electrode for picking up
electromyographic signals directly from muscles; a blood pressure sensor placed in an artery, vein, or the heart
itself; or a blood flow transducer positioned on a major artery are all examples of invasive sensors.
We can also classify sensors in terms of the quantities that they measure. Physical sensors are used in
measuring physical quantities such as displacement, pressure, and flow, while chemical sensors are used to
determine the concentration of chemical substances within the host. A subgroup of the chemical sensors that
are concerned with sensing the presence and the concentration of biochemical materials in the host are known
as bioanalytical sensors, or sometimes they are referred to as biosensors.
In the following paragraphs we will look at each type of sensor and present some examples as well as describe
some of the important issues surrounding these types of sensors.
114.2 Physical Sensors
Physical variables associated with biomedical systems are measured by a group of sensors known as physical
sensors. A list of typical variables that are frequently measured by these devices is given in Table 114.2. These
quantities are similar to physical quantities measured by sensors for nonbiomedical applications, and the devices
used for biomedical and nonbiomedical sensing are, therefore, quite similar. There are, however, two principal
exceptions: pressure and flow sensors.
The measurement of blood pressure and blood flow in humans and other animals remains a difficult problem
in biomedical sensing. Direct blood pressure measurement refers to evaluation of the blood pressure using a
sensor that is in contact with the blood being measured or contacts it through an intermediate fluid such as a
physiologic saline solution. Direct blood pressure sensors are invasive. Indirect blood pressure measurement
involves a sensor that does not actually contact the blood. The most familiar indirect blood pressure measure-
ment is the sphygmomanometer cuff that is usually used in most medical examinations. It is a noninvasive
instrument. Until recently, the primary sensor used for direct blood pressure measurement was the unbonded
strain gauge pressure transducer shown in Fig. 114.1. The basic principle of this device is that a differential
pressure seen across a diaphragm will cause that diaphragm to deflect. This deflection is then measured by a
displacement transducer. In the unbonded strain gauge sensor a closed chamber is covered by a flexible
diaphragm. This diaphragm is attached to a structure that has four fine gauge wires drawn between it and the
chamber walls. A dome with the appropriate hardware for coupling to a pressure source covers the diaphragm
on the side opposite the chamber such that when the pressure in the dome exceeds the pressure in the chamber,
the diaphragm is deflected into the chamber. This causes two of the fine wires to stretch by a small amount
while the other two wires contract by the same amount. The electrical resistance of the wires that are stretched
increases while that of the wires that contract decreases. By connecting these wires, or more correctly these
unbonded strain gauges, into a Wheatstone bridge circuit, a voltage proportional to the deflection of the
diaphragm can be obtained.
In recent years semiconductor technology has been applied to the design of pressure transducers. Silicon
strain gauges that are much more sensitive than their wire counterparts are formed on a silicon chip, and
micromachining technology is used to form this portion of the chip into a diaphragm with the strain gauges
integrated into its surface. This structure is then incorporated into a plastic housing and dome assembly. The
entire sensor can be fabricated and sold inexpensively so that disposable, single-use devices can be made. These
TABLE 114.1Classification of
Biomedical Sensors According
to Their Interface with the
Biologic Host
Noninvasive Noncontacting
Body surface
Invasive Indwelling
Implanted
TABLE 114.2Physical Variables Sensed by
Biomedical Sensors
Displacement, velocity, acceleration (linear and angular)
Temperature
Force (weight and mass)
Pressure
Flow
Radiant energy (optical)
? 2000 by CRC Press LLC
have the advantage that they are only used on one patient and they do not have to be cleaned and sterilized
between patients. By using them on only one patient, the risk of transmitting blood-borne infections is
eliminated.
In biomedical applications pressure is generally referenced to atmospheric pressure. Therefore, the pressure
in the chamber of the pressure transducer must be maintained at atmospheric pressure. This is done by means
of a vent in the chamber wall or a fine bore, flexible capillary tube that couples the chamber to the atmosphere.
This tube is usually included in the electrical cable connecting the pressure transducer to the external instru-
mentation such that the tube is open to the atmosphere at the cable connecter.
In using this sensor to measure blood pressure the dome is coupled to a flexible plastic tube, and the dome
and tube are filled with a physiological saline solution.
1
As described by Pascal’s Law, the pressure in the dome,
and hence against the diaphragm, will be the same as that at the tip of the tube provided the tip of the tube is
at the same horizontal level as the dome. Thus by threading the tube into a blood vessel, an invasive procedure,
the blood pressure in that vessel can be transmitted to the dome and hence the diaphragm of the pressure
transducer. The pressure transducer will, therefore, sense the pressure in the vessel. This technique is known
as external direct blood pressure measurement, and the flexible plastic tube that enters the blood vessel is known
as a catheter. It is important to remember that the horizontal level of the blood pressure transducer dome must
be the same as that of the tip of the catheter in the blood vessel to accurately measure the pressure in that vessel
without adding an error due to the hydrostatic pressure in the catheter.
In addition to problems due to hydrostatic pressure differ-
ences between the chamber and the dome, catheters introduce
pressure errors as a result of the dynamic properties of the
catheter, fluid, dome, and diaphragm. These properties as well
as air bubbles in the catheter, or obstructions due to clotted
blood or other materials, introduce resonances and damping.
These problems can be minimized by utilizing miniature
pressure transducers fabricated using microelectronic semi-
conductor technology that are located at the tip of a catheter
rather than at the end that is external to the body. A general
arrangement for such a pressure transducer is shown in Fig. 114.2. As with the disposable sensors, strain gauges
are integrated into the diaphragm of the transducer such that they detect very small deflections of this dia-
phragm. Because of the small size, small diaphragm displacement, and lack of a catheter with a fluid column,
these sensors have a much broader frequency response, give a clearer signal, and do not have any hydrostatic
pressure error.
1
It must be pointed out that the use of such a sensor is not limited to blood pressure measurement. The strain gauge
pressure sensor can be used to measure the pressure of any fluid to which it is appropriately coupled.
FIGURE 114.1 An unbonded strain gauge pressure transducer.
FIGURE 114.2A catheter tip pressure transducer.
? 2000 by CRC Press LLC
Although the indwelling catheter tip pressure transducer appears to solve many of the problems associated
with the external pressure transducer, there are still important problems in pressure transducer design that
need to be addressed. Long-term stability of pressure transducers is not very good. This is especially problematic
for venous pressure measurements which are carried out at relatively low pressure. Long-term changes in
baseline pressure require pressure transducers to be frequently adjusted to be certain of zero pressure. While
this can be done relatively easily for external and indwelling pressure transducers, there is no way to carry out
this procedure for implanted transducers, since there is not a way to establish zero pressure at the sensor. Thus
devices that have very low long-term baseline drift are essential for implantable applications.
The packaging of the pressure transducer also represents a problem that needs to be addressed. Packaging
must both protect the transducer and be biocompatible. It also must allow the appropriate pressure to be
transmitted from the biologic fluid to the diaphragm. The amount of packaging material required should be
kept at a minimum so as not to substantially increase the size of implantable or indwelling sensors. Furthermore,
the material must be mechanically stable so that it does not swell or contract, since this will most likely change
the baseline pressure seen by the sensor. These problems need to be overcome before miniature pressure
transducers can be used reliably in implantable applications.
114.3 Chemical Sensors
There are many biomedical problems where it is
necessary to know the concentration of a particular
substance in a biological sample. Chemical sensors
provide the interface between an instrument and
the specimen to allow one to determine this con-
centration. These sensors can be used on a biolog-
ical specimen taken from the host and tested in a
laboratory, or they can be used for in vivo measure-
ments either as noninvasive or invasive sensors, the
latter being the most frequently used.
There are many types of chemical sensors used
in biomedical instrumentation. Table 114.3 lists
some general categories of sensors. Electrochemical
and optical sensors are most frequently used for biomedical measurements both in vivo and in vitro. An example
of an electrochemical sensor is the Clark electrode illustrated in Fig. 114.3. This consists of an electrochemical
cell separated from the specimen being measured by an oxygen-permeable membrane. The cell is driven at a
fixed potential of 600 mV, and under these conditions the following reaction occurs at the noble metal cathode:
O
2
+ 4e
–
+ H
2
O ? 4OH
–
FIGURE 114.3 The Clark electrode, an amperometric electrochemical sensor of oxygen.
TABLE 114.3Classifications of Chemical Biomedical Sensors
1.Electrochemical
a.Amperometric
b.Potentiometric
c.Coulometric
2.Optical
a.Colorimetric
b.Emission and absorption spectroscopy
c.Fluorescence
d.Chemiluminescence
3.Thermal methods
a.Calorimetry
b.Thermoconductivity
4.Nuclear magnetic resonance
? 2000 by CRC Press LLC
This reaction involves the reduction of molecular oxygen that diffuses into the cell through the oxygen-
permeable membrane. Since the other components of the reaction are in abundance, the rate of the reaction
is limited by the amount of oxygen available. Thus, the rate of electrons used at the cathode is directly related
to the available oxygen. In other words, the cathode current is proportional to the partial pressure of oxygen
in the specimen being measured.
The electrochemical cell is completed by the silver anode. The reaction at the anode involves forming the
low-solubility salt, silver-chloride, from the anode material itself and the chloride ion contained in the electro-
lyte. The cell is designed so that these materials are also in abundance so that their concentration does not
affect the sensor performance. This type of sensor is an example of an amperometric electrochemical sensor.
Another type of electrochemical sensor that is frequently used in biomedical laboratories is the glass pH
electrode illustrated in Fig. 114.4. The acidity or alkalinity of a solution is characterized by its pH. This quantity
is defined as
pH = – log
10
[H
+
]
where [H
+
] is the activity of the hydrogen ions in solution, a quantity that is related to the concentration of
the hydrogen ions. This sensor only works in an aqueous environment. It consists of an inner chamber
containing an electrolytic solution of known pH and an outer solution with an unknown pH that is to be
measured. The membrane consists of a specially formulated glass that will in essence allow hydrogen ions to
pass in either direction but will not pass other chemical species. If the concentration of hydrogen ions in the
external solution is greater than that in the internal solution, there will be a gradient forcing hydrogen ions to
diffuse through the membrane into the internal solution. This will cause the internal solution to have a greater
positive charge than the external solution so that an electrical potential and, hence, an electric field will exist
across the membrane. This field will counteract the diffusion of hydrogen ions due to the concentration
difference and so an equilibrium will be eventually established. The potential across the membrane at this
equilibrium condition will be related to the hydrogen ion concentration difference (or more accurately the
activity difference) between the inner and outer solutions. This potential is given by the Nernst equation
where E is the potential measured, R is the universal gas constant, T is the absolute temperature, n is the valence
of the ion, and a
1
and a
2
are the activities of the ions on each side of the membrane. Thus the potential measured
across the glass membrane will be proportional to the pH of the solution being studied. At room temperature
FIGURE 114.4 A glass electrode pH sensor.
E
RT
nF
a
a
=-
?
è
?
?
?
÷
ln
1
2
? 2000 by CRC Press LLC
the theoretical sensitivity of the electrode is approximately 60 mV/pH. It is not practical to measure the potential
across the membrane directly and so reference electrodes, sensors that can be used to measure electrical potential
of an electrolytic solution, are used to contact the solution on either side of the membrane to measure the
potential difference across it. The reference electrodes and the glass membrane are incorporated into the
structure shown in Fig. 114.4 known as a glass pH electrode. This is an example of a potentiometric measure-
ment made using an ion-selective membrane.
There are other types of ion-selective membrane potentiometric chemical sensors that are used for biomedical
applications. The membranes of these sensors determine the ion being sensed. The membrane can be based
upon glass or a polymeric material such as polyvinyl chloride, but the key component is the substance that is
added to the membrane that allows it to selectively pass a single ion.
Important problems in the development of chemical biomedical sensors are similar to those discussed above
for the pressure sensor. Issues of long-term stability and packaging are critical to the success of a chemical
sensor. The package is even more critical in chemical sensors than it was in pressure sensors in that the package
must protect portions of the sensor that require isolation from the solutions being measured while it provides
direct contact of the chemically sensitive portions of the sensor to the solution. The maintenance of a window
through the package for this contact represents a critical aspect of sensor development. Frequent calibration is
also necessary for chemical sensors. Just about every type of chemical sensor requires some sort of calibration
using a standard solution with known concentration of the analyte being sensed. The best calibration method
is a two-point procedure where two standards are used to establish the slope and the intercept of the calibration
line. Some chemical sensors have stable slopes but need to be calibrated in terms of the baseline or intercept.
In this case a single-point calibration can be used.
114.4 Bioanalytical Sensors
A special class of sensors of biological molecules has evolved in recent years. These bioanalytical sensors take
advantage of one of the following biochemical reactions: (1) enzyme-substrate, (2) antigen-antibody, or (3)
ligand-receptor. The advantage of using these reactions in a sensor is that they are highly specific for a particular
biological molecule, and sensors with high sensitivity can be developed based upon these reactions. The basic
structure of a bioanalytical sensor is shown in Fig. 114.5. There are two principal portions of the sensor. The
first contains one component of the biological sensing reaction such as the enzyme or the antibody, and the
second component involves a means of detecting whether the biological reaction has taken place. This second
portion of a bioanalytical sensor is made up of either a physical or chemical sensor that serves as the detector
FIGURE 114.5 A generalized bioanalytical sensor.
? 2000 by CRC Press LLC
of the biological reaction. As illustrated in Fig. 114.5, this detector can consist of an electrical sensor such as
used in electrochemical sensors, a thermal sensor, a sensor of changes in capacitance, a sensor of changes in
mass, or a sensor of optical properties.
An example of a bioanalytical sensor is a glucose sensor. The first portion of the sensor contains the enzyme
glucose oxidase. This enzyme promotes the oxidation of glucose to glucuronic acid and consumes oxygen in
the process. Thus, by placing an oxygen sensor along with the glucose oxidase in the bioanalytical sensor, one
can determine the amount of glucose oxidized by measuring the amount of oxygen consumed. An even better
approach is to have two identical sensor structures in the same package. The only difference is that only one
of the sensors contains the enzyme. When there is no glucose present, both sensors will measure the same
oxygen partial pressure. The presence of glucose, however, will cause the sensor with the glucose oxidase to
have a reduced partial pressure of oxygen due to the oxygen consumption of the reaction. By making a
differential measurement of oxygen partial pressure with both sensors, other factors that can cause an apparent
change in oxygen partial pressure such as temperature will have a much lower effect than if a single sensor was used.
Stability problems are important for bioanalytical sensors, especially those that are used for long-term
measurements. Not only are the stability issues the same as for the physical and chemical sensors, but they are
also related to preservation of the biological molecules used in the first stage of the sensor. These molecules
can often be degraded or destroyed by heat or exposure to light. Even aging can degrade some of these molecules.
Thus, an important issue in dealing with bioanalytical sensors is the preservation of the biochemical components
of the sensor. Not all biochemical reactions are entirely reversible, and so the bioanalytical sensors based on
them will not be reversible as well. This may be acceptable for some applications but not for others and must
be taken into consideration in choosing a bioanalytical sensor.
114.5 Applications
Biomedical sensors and instrumentation are used in biomedical research and patient care applications. In terms
of patient care, sensors are used as a part of instruments that carry out patient screening by making measure-
ments such as blood pressure using automated apparatus. Specimen analysis is another important application
of biomedical sensors in patient care. This can include analyses that can be carried out by the patients themselves
in their homes such as is done with home blood glucose analyzers. Instrumentation based upon biomedical
sensors can be used in the physician’s office for carrying out some chemical analyses of patient specimens such as
urinalysis or elementary blood chemistries such as serum glucose and electrolytes. Sensors also are a part of large
multicomponent automatic blood analyzers used in the central clinical laboratory of major medical centers.
Another application for biomedical sensors is in patient monitoring. Sensors represent the front end of
critical care monitors used in the intensive care unit and in the operating and recovery rooms. Measurements
cover a wide range of biomedical variables such as continuous recordings of blood pressure and transcutaneous
measurement of the partial pressure of carbon dioxide in the blood. The performance of these instruments is
strongly dependent on biomedical sensors. Patient monitoring can also be carried out in the various clinical
units of the hospital. Devices such as ambulatory cardiac monitors that allow patients to be observed while
they are free to move around if they desire are becoming important in clinical care in “step-down” units for
patients who have completed their stay in the intensive care unit. Patient monitoring has even made its way
into the home. Home cardiorespiratory monitors are thought to have some potential value in identifying infants
at risk of sudden infant death.
114.6 Summary
Sensors serve an important function in biomedical instrumentation systems in that they provide the interface
between the electronic instrument and the biologic system being measured. Very often the quality of the instrument
is based upon the quality of the sensor at the instrument’s front end. Although electronic signal processing has been
developed to a high level, the signals are no better than the quality of the sensors that provide them. Although there
have been many advances in biomedical sensor technology, many problems remain. Biomedical sensors will continue
to be an import area for research and development in biomedical engineering.
? 2000 by CRC Press LLC
Defining Terms
Amperometric sensor: An electrochemical sensor that determines the amount of a substance by means of
an oxidation-reduction reaction involving that substance. Electrons are transferred as a part of the
reaction, so that the electrical current through the sensor is related to the amount of the substance seen
by the sensor.
Analyte: The substance being measured by a chemical or bioanalytical sensor and instrumentation system.
Bioanalytical sensor: A special case of a chemical sensor for determining the amount of a biochemical
substance. This type of sensor usually makes use of one of the following types of biochemical reactions:
enzyme-substrate, antigen-antibody, or ligand-receptor.
Biomedical sensor: A device for interfacing an instrumentation system with a biological system such as a
biological specimen or an entire organism. The device serves the function of detecting and measuring
in a quantitative fashion a physiological property of the biologic system.
Chemical sensor: The interface device for an instrumentation system that determines the concentration of
a chemical substance.
Noninvasive sensor: The interface device of an instrumentation system that measures a physiologic variable
from an organism without interrupting the integrity of that organism. This device can be in direct contact
with the surface of the organism or it can measure the physiologic quantity while remaining remote from
the organism.
Physical sensor: An interface device at the input of an instrumentation system that quantitatively measures
a physical quantity such as pressure or temperature.
Potentiometric sensor: A chemical sensor that measures the concentration of a substance by determining
the electrical potential between a specially prepared surface and a solution containing the substance being
measured.
Related Topics
56.1 Introduction ? 56.2 Physical Sensors ? 56.3 Chemical Sensors ? 56.4 Biosensors ? 56.5 Microsensors
References
R.S.C. Cobbold, Transducers for Biomedical Measurements: Principles and Applications, New York: John Wiley,
1974.
B.R. Eggins, Biosensors: An Introduction, Chichester; New York: John Wiley, 1996.
D.G. Fleming, W.H. Ko, and M.R. Neuman, Eds., Indwelling and Implantable Pressure Transducers, Cleveland:
CRC Press, 1977.
L.A. Geddes, The Direct and Indirect Measurement of Blood Pressure, Chicago: Year Book Medical Publishers,
1970.
L.A. Geddes, Electrodes and the Measurement of Bioelectric Events, New York: John Wiley, 1972.
W. G?pel, J. Hesse and J.N. Zemel, Sensors; A Comprehensive Survey, Weinheim, Germany:
VCH Verlagsgesellschaft, 1989.
A.H. Hall, Biosensors, Englewood Cliffs, N.J.: Prentice Hall, 1991.
J. Janata, Principles of Chemical Sensors, New York: Plenum Press, 1989.
M.R. Neuman, R.P. Buck, V.V. Cosofret, E. Lindner, and C.C. Liu, “Fabricating biomedical sensors with thin-
film technoogy, IEEE Engineering in Medicine and Biology Magazine, 13, 409–419, 1994.
R. Pallas-Areny and J.G. Webster, Sensors and Signal Conditioning, New York: John Wiley, 1991.
J.I. Peterson and G.G. Vurek, “Fiber-optic sensors for biomedical applications,” Science, vol. 224, pp. 123–127,
1984.
P. Rolfe, “Review of chemical sensors for physiological measurement,” J. Biomed. Eng., vol. 10, pp. 138–145, 1988.
J.G. Webster, Ed., Encyclopedia of Medical Devices and Instrumentation, New York: John Wiley, 1988.
O.S. Wolfbeis, Ed., Fiber Optic Chemical Sensors and Biosensors, Boca Raton, Fla: CRC Press, 1991.
? 2000 by CRC Press LLC
Further Information
Research reports on biomedical sensors appear in many different journals ranging from those that are concerned
with clinical medicine through those that are engineering and chemistry oriented. Three journals, however,
represent major sources of biomedical sensor papers. These are listed as follows:
The IEEE Transactions on Biomedical Engineering is a monthly journal devoted to research papers on bio-
medical engineering. Papers on biomedical sensors frequently appear, and the February 1986 issue was devoted
entirely to the topic of biomedical sensors. For more information or subscriptions, contact IEEE Service Center,
445 Hoes Lane, P.O. Box 1331, Piscataway, NJ 08855-1331.
The international journal Medical and Biological Engineering and Computing is published bi-monthly by the
International Federation for Medical and Biological Engineering. This journal also contains frequent reports
on biomedical sensors and related topics. Subscription information can be obtained from Peter Peregrinus Ltd.,
P.O. Box 96, Stevenage, Herts SG12SD, United Kingdom.
The journal Biomedical Instrumentation and Technology is published by the Association for the Advancement
of Medical Instrumentation. This bimonthly journal has reports on biomedical instrumentation for clinical
applications, and these include papers on biomedical sensors. Subscription information can be obtained from
Hanley and Belfus, 210 S. 13th Street, Philadelphia, PA 19107.
There are also several scientific meetings that include biomedical sensors. The major meeting in the area is
the international conference of the IEEE Engineering in Medicine and Biology Society. An extensive book or
CD ROM of extended abstracts for this meeting is published each year by the IEEE. Further information can
be obtained by contacting the IEEE at the address listed above.
? 2000 by CRC Press LLC