Transmittal No. 96-E1 (9/96)
Form FDA 2905a (6/92)
Pesticide Analytical Manual Vol. I SECTION 500
500–1
Table of Contents
page date
501: General Information
501 A: Principles 501-1 1/94
501 B: Equipment for GLC 501-1 1/94
Gas Chromatographic Components 501-1 1/94
Other Apparatus 501-2 1/94
Reagents and Gases 501-3 1/94
501 C: Residue Methodology for GLC Determination 501-3 1/94
Cleanup 501-4 1/94
Reagent Blanks 501-5 1/94
Choice of Solvent 501-6 1/94
501 D: Injection Techniques 501-6 1/94
Manual Injection 501-6 1/94
Autoinjectors 501-7 1/94
501 E: Reference Standards 501-8 1/94
References 501-8 1/94
Chapter 1
Regulatory Operations
Chapter 2
General Analytical
Operations and Information
Chapter 3
Multiclass
MRMs
Chapter 5
GLC
Chapter 4
Selective
MRMs
Chapter 6
HPLC
Transmittal No. 96-E1 (9/96)
Form FDA 2905a (6/92)500–2
SECTION 500 Pesticide Analytical Manual Vol. I
page date
502: Columns
502 A: Introduction 502-1 1/94
Column Specifications 502-1 1/94
Column Parameters 502-2 1/94
502 B: Packed Columns 502-6 1/94
Components of Packed Columns 502-6 1/94
Preparation of Packed Columns 502-8 1/94
Installation of Packed Columns 502-9 1/94
Conditioning of Packed Columns 502-10 1/94
Rejuvenation of Packed Columns 502-10 1/94
Criteria for Acceptable Packed Columns 502-11 1/94
Recommended Operating Procedures 502-12 1/94
for Packed Columns
502 C: Open Tubular Capillary Columns 502-13 1/94
Column Description 502-14 1/94
Injection onto Capillary Columns 502-16 1/94
Capillary Column Systems 502-16 1/94
Installation and Conditioning of 502-20 1/94
Capillary Columns
Rejuvenation of Capillary Columns 502-21 1/94
Recommended Operating Procedure for 502-22 1/94
Wide Bore Columns (Isothermal)
Apparatus and Reagents 502-22 1/94
System Startup and Inspection 502-25 1/94
References 502-25 1/94
503: Detectors
503 A: Introduction 503-1 1/94
Definitions of Detector Characteristics 503-1 1/94
503 B: Electron Capture Detector 503-2 1/94
Principles 503-3 1/94
Design 503-3 1/94
Apparatus and Reagents 503-3 1/94
Detector Characteristics 503-4 1/94
Other Influences on Detector Performance 503-7 1/94
Recommended Operating Procedures 503-8 1/94
503 C: Flame Photometric Detector 503-9 1/94
Principles 503-9 1/94
Design 503-9 1/94
Apparatus and Reagents 503-10 1/94
Detector Characteristics 503-10 1/94
Other Influences on Detector Performance 503-11 1/94
Recommended Operating Procedures 503-12 1/94
Troubleshooting 503-13 1/94
503 D: Electrolytic Conductivity Detector 503-14 1/94
Principles 503-14 1/94
Transmittal No. 96-E1 (9/96)
Form FDA 2905a (6/92)
Pesticide Analytical Manual Vol. I SECTION 500
500–3
page date
Design 503-14 1/94
ElCD-X 503-16 1/94
Principles 503-16 1/94
Apparatus and Reagents 503-16 1/94
Detector Characteristics 503-17 1/94
Other Influences on Detector Performance 503-19 1/94
Recommended Operating Procedures 503-21 1/94
System Suitability Test 503-21 1/94
Troubleshooting 503-22 1/94
ElCD-N 503-23 1/94
Principles 503-23 1/94
Apparatus and Reagents 503-23 1/94
Detector Characteristics 503-23 1/94
Other Influences on Detector Performance 503-24 1/94
Recommended Operating Procedures 503-25 1/94
System Suitability Test 503-25 1/94
Troubleshooting 503-25 1/94
General Precautions for ElCDs 503-26 1/94
503 E: Nitrogen/Phosphorus Detector 503-27 1/94
Principles 503-27 1/94
Design 503-27 1/94
Apparatus and Reagents 503-28 1/94
Detector Characteristics 503-28 1/94
Other Influences on Detector Performance 503-30 1/94
Recommended Operating Procedures 503-31 1/94
References 503-32 1/94
504: Quantitation
504 A: Introduction 504-1 1/94
504 B: Manual Quantitation 504-2 1/94
Measurement of Peak Height 504-2 1/94
Measurement of Area by Triangulation 504-2 1/94
504 C: Electronic Integration 504-3 1/94
504 D: Special Considerations for Complex 504-6 1/94
Chromatograms
BHC 504-6 1/94
Chlordane 504-8 1/94
PCBs 504-10 1/94
Toxaphene 504-14 1/94
References 504-15 1/94
505: Bibliography
General Texts 505-1 1/94
Inlets 505-1 1/94
Columns 505-1 1/94
Detectors 505-1 1/94
Transmittal No. 96-E1 (9/96)
Form FDA 2905a (6/92)500–4
SECTION 500 Pesticide Analytical Manual Vol. I
page date
Figures
502-a Polysiloxane Stationary Phases 502-2 1/94
502-b GLC Column Parameters 502-4 1/94
502-c Ferrules for Connecting Glass and Metal 502-9 1/94
502-d Capillary Column Cross-Section 502-15 1/94
502-e Inlet Adapters for Capillary Columns 502-17 1/94
502-f Van Deemter Curves 502-19 1/94
502-g Capillary Column Inlet System 502-23 1/94
503-a Two EC Detector Designs 503-3 1/94
503-b Reproducibility 503-5 1/94
503-c Dynamic Response Ranges 503-6 1/94
503-d Linear Response Range 503-6 1/94
503-e Effect of Detector Temperature 503-7 1/94
503-f Effect of Carrier Gas Flow Rate 503-7 1/94
503-g Single-Flame FPD 503-9 1/94
503-h Dual-Flame FPD 503-10 1/94
503-i Block Diagram of the ElCD 503-15 1/94
503-j ElCD Reactor and Conductivity Cell 503-15 1/94
503-k N/P Detector Components 503-28 1/94
503-l N/P Detector Configurations 503-29 1/94
504-a Manual Peak Measurement 504-2 1/94
504-b Triangulation of Peak on Sloping Baseline 504-3 1/94
504-c Technical Chlordane 504-8 1/94
504-d Chlordane, Heptachlor, Heptachlor Epoxide 504-9 1/94
504-e PCBs in Chinook Salmon 504-13 1/94
504-f Toxaphene 504-14 1/94
Tables
502-a: Common GLC Liquid Phases Used in Pesticide 502-3 1/94
Residue Determination
502-b: Operating Conditions for Packed Columns 502-13 1/94
504-a: Effects of Changing Electronic Integrator 504-6 1/94
Settings
504-b: Response of Two Detectors to Four 504-7 1/94
BHC Isomers
504-c: Weight Percent Factors for Individual Gas 504-12 1/94
Chromatographic Peaks in Aroclor Reference
Standards
Pesticide Analytical Manual Vol. I SECTION 501
501–1
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
501: GENERAL INFORMATION
Multiresidue methodology by definition requires determinative steps capable of
separating analytes from one another so each can be detected and measured
individually. Both gas-liquid chromatography (GLC) and high performance liquid
chromatography (HPLC) provide these capabilities, and both are used in modern
laboratories.
GLC has been the predominant determinative step in pesticide multiresidue
methodology for over 30 years. Because GLC involves interaction between a vapor
phase and liquid phase, its application is restricted to analytes that can be vapor-
ized without degradation. For heat-labile chemicals, HPLC offers a variety of alter-
native schemes for separating analytes according to chemical or physical charac-
teristics, but GLC’s relative simplicity and ruggedness cause it to remain the deter-
minative step of choice for residues to which it is applicable.
501 A: PRINCIPLES
Separation in GLC is achieved by differences in distribution of analytes between
mobile and stationary phases, causing them to move through the column at dif-
ferent rates and from it at different times [1]. A measured aliquot of solution is
injected into a gas chromatographic column through an inlet heated to a suffi-
ciently high temperature that analytes are vaporized. In this state, the flow of inert
gas that forms the mobile phase sweeps analytes through the column; retarding
this movement is the analyte’s solubilization in the liquid phase. During passage
through the column, analytes that were injected in the same solution separate
from one another because of their different vapor pressures and selective interac-
tions with the liquid phase [2]. When analytes elute from the column and enter
a detector, the detector responds to the presence of a specific element or func-
tional group within the molecule. The detector’s response causes a change in
electronic signal, which is proportional to the amount of residue; the signal is
amplified and recorded as a chromatogram.
Analytes are identified by the time it takes them to pass through a column of
specific liquid phase (retention time), at a specified temperature and gas flow.
Quantities are calculated from the detector response. Both retention time and
response are compared to values obtained for a reference standard solution in-
jected into the same system.
501 B: EQUIPMENT FOR GLC
Gas Chromatographic Components
The basic gas chromatograph consists of an inlet system, column, detector, elec-
tronic equipment to amplify the detector signal, and a recorder or other data-
handling device. Carrier gas(es), with appropriate pneumatic system(s), are also
integral to the GLC system. The inlet system, column, and detector are maintained
in temperature-controlled environments.
The following are desirable features in GLC hardware:
1) Inlet, column oven, and detector should be individually heated and
temperature-controlled. Temperature should be maintained to ±0.1° C.
501–2
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Pesticide Analytical Manual Vol. ISECTION 501
Control of detector temperature usually is not as critical but should be
well controlled, constant, and not affected by such things as line voltage
fluctuations.
2) Temperature readout should be available for column, detector, and inlet.
(Check accuracy of instrument temperature indicators with accurate py-
rometer.)
3) Instrument design should be simple enough to facilitate troubleshooting
and repairs. Design should permit easy removal or inspection of either
column or detector without affecting the temperature of the other.
4) System should be designed to prevent or minimize contact between sample
injection and any metal parts; system should be all-glass (or as near as
possible).
Several sizes of packed and open tubular capillary columns are used in
residue analysis, and hardware for inlet and column must accommodate
configurations that will be needed. Section 502, Columns, includes direc-
tions for adapting equipment.
5) Certain detectors may require multiple heated zones, including combus-
tion furnaces. For flexibility, designs that permit ready access for servicing
and maintenance are preferred. Section 503 provides details on various
detectors used in pesticide residue determination.
6) Electrical signal monitoring equipment is usually one of two designs: (1)
amplifier with 1 or 10 mV output, compatible with strip chart recorder,
and (2) amplifier with 1 or 10 V output, compatible with data processing
by either electronic integrator or computer. Other remote devices such as
autosamplers can be easily adapted to any of these systems.
Other Apparatus
Gas Regulators. Two-stage gas pressure regulators with stainless steel diaphragms
are required for all GLC determinations of trace residues. Regulators with a sec-
ondary stage maximum pressure of 80 psi are acceptable, but those with 200 psi
offer more flexibility. If a hydrogen purifier is used (below), the latter type of
regulator is required, because higher pressure is needed.
Gas lines that connect gas tanks to the chromatograph must be clean and free of
components that contain oil or gas-purgeable elastomers; “refrigeration grade”
copper (i.e., cleaned of all oil) is preferred. Tubing (even refrigeration grade)
should be sequentially rinsed with methylene chloride and acetone before use.
Plastic and nylon lines must be avoided to reduce the likelihood of air contami-
nating the gas.
Syringes. The most common syringes for injection of food extracts into a chro-
matograph are 5 and 10 μL fixed needle syringes with 22° bevel points; some other
sizes may be needed for special purposes. Hamilton syringes or equivalent are
available from all chromatography suppliers. Plunger “guides” are available as
options to minimize bending the plunger during injection.
Pesticide Analytical Manual Vol. I SECTION 501
501–3
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Some specialty products exist to facilitate injection and minimize aggravation, and
each has found favor with some analysts. For example, syringes with removable
needles permit replacement of needles on which “burrs” have formed that destroy
septa; removable needles with a “side port point” do not shred the septa as do
standard bevel point needles; and syringes with plungers and needles made of a
titanium alloy cannot be bent.
Reagents and Gases
Reagents associated with GLC include column liquid phases and solid supports,
gases used for mobile phase and for detector reactions, and certain other reagents
relevant to detector operation. Most of these reagents are discussed further in
pertinent sections of this chapter; only gases, including filters used to remove
contaminants from gas flow, are included in this introductory section.
Helium, hydrogen, and nitrogen are most commonly used as column carrier gases.
Purity is always critical to avoid damage to the column, and more stringent purity
requirements may be imposed by the detector. Purity specifications of the instru-
ment manufacturer should always be followed.
Helium and hydrogen requirements range from 99.999-99.9999% purity, depend-
ing on the detector. Even with the highest purity, oxygen traps, available from
chromatography suppliers, are recommended; traps that change color when per-
meated with oxygen are ideal for alerting the analyst to potential problems.
Purchase of ultra high purity helium and hydrogen may not be necessary if spe-
cially designed purifiers are used. Purifiers are available that permit use of com-
mercial grade gases (99.995%) at a much lower price, justifying the cost of the
purifier. Different purifiers are needed for helium and hydrogen; they are not
interchangeable. FDA has had successful experiences with:
hydrogen purifiers: Model 560, AADCO Instruments, Inc., Clearwater, FL;
Model 8372V, Consolidated Technologies, Inc., West Chester, PA
helium purifiers: Product # HP, Valco Instrument Co., Houston, TX;
Model 2-3800, Supelco, Bellefonte, PA
Nitrogen is used as a carrier gas only for packed columns (Section 502 B). Either
nitrogen or argon/methane (95+5 or 90+10) is also required as a carrier and/or
makeup gas for the electron capture detector (Section 503 B). Commercial grades
of these gases are acceptable if oxygen and moisture traps are used between the
gas tank and the chromatograph.
501 C: RESIDUE METHODOLOGY FOR GLC DETERMINATION
Applications of analytical methodology require consideration of many factors to
assure compatibility of method steps. The following factors related to extraction
and cleanup of food samples profoundly influence accuracy and reliability of GLC
determinative steps.
501–4
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Pesticide Analytical Manual Vol. ISECTION 501
Cleanup
Solvent extraction of pesticide residues also extracts food constituents (“co-extrac-
tives”) from the sample. Cleanup steps are included in residue analytical methods
to remove co-extractives that can interfere in the determinative step of the analysis
or cause damage to the column and/or detector.
For many years, predominant use of the nonselective electron capture (EC) detec-
tor caused justifiable concern about potential detector response to nonpesticidal
co-extractives. In addition, documented cases in which sample co-extractives dam-
aged GLC columns and caused subsequent breakdown of injected residues sup-
ported the need for extensive cleanup prior to GLC determination [3].
More recently, several factors have reduced emphasis on cleanup. The more selec-
tive GLC detectors now in use have decreased the likelihood that sample or re-
agent artifacts might be mistaken for pesticide residues. In addition, use of cap-
illary columns, which are more efficient than equivalent packed columns, result in
increased peak height response for the same amount of analyte. The amount of
extract injected can thus be reduced without changing the level of quantitation,
and this in turn reduces the likelihood of damage to the GLC system. Inlet liners
and adapters used with capillary columns (Section 502 C) also provide the column
with some degree of protection from damage caused by co-extractives. Finally,
there are many incentives to perform more analyses with the same or fewer re-
sources and to minimize the volume of solvents that must be purchased and
disposed of. These factors contribute to a trend toward performing only minimal
cleanup of sample extracts during routine surveillance analyses, with the intention
of cleanup with applicable step(s) if an extract is found to contain interfering
materials.
Despite these compelling reasons to reduce cleanup, GLC systems that are not
protected from co-extractives deteriorate faster than those into which only cleaned
up extracts are injected. The column and/or detector may be damaged by injec-
tion of insufficiently cleaned up samples, especially when the method and the
chromatograph are used repeatedly. Such detrimental effects can occur even when
the chromatogram appears to be clean enough for residue identification and
measurement. Experience with a variety of sample types should make the analyst
aware of these occurrences.
Detector response to sample co-extractives (artifacts) is still possible even with
element-selective detectors. Although a selective detector is less likely to respond
to chemically unrelated artifacts than the nonselective EC detector, artifacts con-
taining an element to which the detector responds can still interfere with residue
analysis. This occurs most often with nitrogen-selective detectors because of the
number of nitrogenous chemicals in foods, but it can occur with any detector.
Likelihood of interferences and potential for mistaken identity increase with de-
creasing cleanup.
Insufficiently clean extracts may also affect quantitative accuracy when determin-
ing residues that are polar or otherwise subject to adsorption by active sites in a
GLC column. Such chemicals usually exhibit poor chromatography when standard
solutions are injected, because adsorption delays or inhibits the chemical during
its passage through the column. Peak tailing and/or changes in retention times
are caused by adsorption. The net effect is an apparently diminished detector
Pesticide Analytical Manual Vol. I SECTION 501
501–5
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
response, which is especially evident if peak height measurements are used rather
than peak area.
In contrast, when an uncleaned extract containing the same analyte is injected
into the GLC system, co-extractives compete for the column’s active sites, and the
analyte moves through the column in a tighter chromatographic band. Analyte
concentration (per unit time) entering the detector thus increases, and detector
response (peak height) is greater. Quantitation by the usual practice (i.e., com-
parison of detector responses to residue and reference standard) results in calcu-
lation of an inaccurately high residue level, especially if peak heights are com-
pared. Quantitative accuracy can be improved for such chemicals by employing
more rigorous cleanup of the extract or by using a GLC column with fewer active
sites.
An appropriate balance is needed between efficiency in processing samples and
accuracy in determining residues. Every injected extract should be sufficiently
clean that it (1) does not jeopardize the column beyond the point that it can be
easily repaired; (2) does not introduce substances that will degrade co-injected or
subsequently injected residues; (3) does not foul any part of the detector, includ-
ing combustion tube, flame, radioactive source, etc.; (4) minimizes introduction of
artifacts to which the detector will respond; and (5) does not cause a dispropor-
tionate response enhancement of the residue in the extract.
Reagent Blanks
The analyst must ascertain that no interference from reagents and/or glassware
occurs during residue analysis. Scrupulous attention is required to eliminate all
such contaminants, and routine analysis of reagent blanks should be specified in
the laboratory quality assurance plan (Section 206).
Contaminants can be introduced from a variety of sources. Studies with the
EC detector have identified interferences from impure solvents, adsorbents, so-
dium sulfate, glass wool, Celite, blender gaskets, laboratory air filters, and polyeth-
ylene containers. The more nonselective the detector, the more likely it is to
respond to interferences introduced by reagents or the environment. A thorough
examination of the reagent blank is also necessary for methods that use a relatively
selective detector. One example demonstrated that chemicals extracted by petro-
leum ether from a polyethylene squeeze bottle caused response by both an EC and
a halogen-selective detector [3]. Contaminants can even be pesticides themselves,
present on glassware or microliter syringes used in prior analyses, or present in the
laboratory environment because of pest control treatment.
When interferences are discovered and the source(s) identified, every effort must
be made to reduce or eliminate the problem. Solvents can be purchased to meet
requirements or may be redistilled. Solids frequently can be washed and/or heated
prior to use. Section 204 provides purity tests and procedures for purifying certain
commonly used reagents; other reagent purification procedures are included in
pertinent method descriptions in Chapters 3 and 4. Sometimes the method cleanup
step removes interferences added to the sample during previous steps, but whether
this is accomplished must be determined by a complete investigation of the method
reagent blank.
Equipment should be washed thoroughly and rinsed with solvent as soon as pos-
sible after use. Syringe plungers and needles should be wiped with lint-free wipers
501–6
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Pesticide Analytical Manual Vol. ISECTION 501
dipped in an appropriate solvent (e.g., acetone), and the barrel should be cleaned
by drawing solvent through the needle and out the top by a vacuum applied to the
top. Particular care should be taken to assure elimination of residues from glass-
ware or syringes previously in contact with high concentrations of pesticides.
Choice of Solvent
The solvent in which the final extract is dissolved must be compatible with the
detector(s) in the GLC determinative step(s). The most basic requirement is that
the solvent not contain elements to which the detector responds. Specifically, no
amount of chlorinated solvent, such as methylene chloride, can remain in extracts
being examined by an EC or halogen-selective detector, and no trace of acetoni-
trile can be present in extracts examined with nitrogen-selective detectors.
Other effects besides element selectivity cause incompatibility between detectors
and solvents. For example, acetonitrile has an unexplained adverse effect on re-
sponse of the EC detector, and aromatic and halogenated solvents may increase
detector response of the N/P detector and eventually render it useless.
Solvent volatility must also be considered when using a detector that requires a
solvent venting time. For these detectors, the most volatile practical solvent in
which residues are soluble should be chosen to minimize length of venting time
and avoid potential loss of early eluting analytes.
Solvent volatility has another practical effect related to the ease with which the
extract can be concentrated. Final volume of concentrated extract must be suffi-
ciently small that the volume injected into the GLC system contains sufficient
equivalent sample weight necessary to reach the targeted level of quantitation
(Section 105). Sensitivity of a particular detector to residues of interest governs
how much sample equivalent must be injected, and column type and arrangement
limit the volume that can be injected. In cases where a very small final extract
volume is needed, or where the concentration step begins with a very large solvent
volume, practicality dictates the choice of a volatile solvent to minimize time needed
for concentration.
501 D: INJECTION TECHNIQUES
The technique used to inject extracts and reference standards into the chromato-
graph is critical to system performance. Improper syringe handling can lead to
myriad problems, including asymmetrical peak shapes and nonreproducible reten-
tion times or responses. Autoinjectors are increasingly used for residue determi-
nation, but manual injection is still practiced.
Manual Injection
If extracts and standards are injected manually, it is imperative that each analyst
develop and follow good technique in syringe handling and sample introduction.
This can be achieved through practice and care. Several methods presently in use
for filling syringes and injecting include:
1) A volume of solvent greater than or equal to needle volume is drawn into
the syringe, followed by a small amount of air. The extract (or reference
standard solution) is then drawn completely into the syringe barrel, where
Pesticide Analytical Manual Vol. I SECTION 501
501–7
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
its volume can be measured by reading both ends of the liquid. Injec-
tion is then made. The initial solvent flushes the extract or standard into
the chromatograph. This technique is referred to as the “solvent flush”
or “sandwich” technique.
2) The syringe is filled by drawing extract (or standard solution) com-
pletely into the barrel (i.e., none is left in the needle). Total volume is
measured by reading both ends of the liquid. Injection is made, with the
syringe removed quickly from the inlet. The syringe plunger is with-
drawn until whatever volume of liquid remains is completely in the
barrel of the syringe, where it is measured as before. The difference in
liquid volume before and after injection is the amount actually injected.
It is important when using this technique to remove the syringe from
the heated injection port as quickly as possible after injection to avoid
any evaporation of liquid remaining in the syringe.
3) The syringe is filled to the desired volume, the volume noted, and the
injection made. The volume measured is considered to be the volume
injected. This technique introduces error, because it ignores the volume
in the needle and the volume that remains after injection. The effective
error can be minimized by use of the same solvent for both sample
extract and standard solution and by injection of the same volume of
each.
Whichever injection technique is chosen, it must be performed reproducibly. Each
analyst should choose the injection technique he/she finds most reproducible and
use it routinely. Poor precision among chromatograms from repetitive injections
may be caused by faulty syringes or poor analyst technique, as well as by inappro-
priate solvents or inadequate sample cleanup. Volume of liquid in the syringe
should be measured by holding the syringe in the same manner each time while
looking toward a light background. The same injection technique must be used
for both the sample extract and the reference standard to which it will be com-
pared.
Choice of injection technique is not solely based on personal preference; type of
column being used (packed vs capillary) must also be considered. Any technique
described above can be applied when using packed columns. However, too much
solvent can overwhelm the small diameter capillary column, so injection volume
must be limited. Several inlet systems and injection options are used with capillary
columns to accommodate both column restrictions and volume requirements of
residue determination (Section 502 C). Consistently good capillary column results
have been achieved with manual injection and the solvent flush technique. The
syringe needle should remain in the inlet 1 sec for each μL injected to allow the
pressure surge from vaporization of solvents to dissipate.
The syringe manufacturer’s recommendations for use and care of the syringe
should be followed. Syringes must be kept free of traces of analyte. This should be
checked occasionally by injecting a volume of pure solvent; if the syringe is clean,
no peaks other than the solvent peak will appear.
Autoinjectors
The best injection performance is achieved using an autoinjector (also called
autosampler). Various commercially available autoinjectors can be interfaced with
501–8
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Pesticide Analytical Manual Vol. ISECTION 501
GLC systems. For normal use of autoinjectors, extracts and standard solutions are
placed in disposable glass vials with vapor-tight septum caps. The autoinjector wets
the syringe completely and removes air bubbles by pumping extract (or standard
solution) into the barrel. It then draws a precisely measured volume of solution
into the barrel and injects it into the chromatograph. Between injections, the
autoinjector flushes the needle with appropriate solvent to clean it. Beyond the
improved reproducibility achieved with autoinjectors, their use permits unattended
operation of the chromatograph and frees the chromatographer to perform other
tasks.
501 E: REFERENCE STANDARDS
Section 205 provides information on pesticide standards. The importance of reli-
able standard solutions to accurate pesticide analyses cannot be overemphasized.
Solvents used for GLC standard solutions are subject to the same requirements
and limitations listed above for extracts.
The quality assurance plan for analyses involving GLC determination should in-
clude routine injection of a mixed standard solution. The mixture should include
compounds normally used as markers for retention time and response and should
also include compounds prone to adsorption or degradation. Vulnerable com-
pounds serve as indicators of problems developing in the system; e.g., the presence
of p,p′-DDT in such a solution serves to alert the analyst when degradation to p,p′-
TDE occurs. GLC systems used for determination of organophosphorus or other
polar residues should be checked with a solution that includes, at a minimum,
methamidophos, acephate, and monocrotophos. Response to acephate may disap-
pear in systems that contain too much glass wool, and response to methamidophos
may not be seen if it elutes with the solvent front or if column packing is of poor
quality; both these situations can be avoided by monitoring the system with rou-
tine injection of an appropriate mixed standard. Frequency of injection of mixed
standard, at least twice during an 8-hr period, should be specified in the quality
assurance plan.
For best quantitative results, reference standards should be dissolved in the same
solvent that is used for the final sample extract. In addition, reference standards
should be injected within minutes of the sample containing the residue(s) to be
quantitated, and responses to residue and standard should match within ±25% for
accurate quantitation.
References
[1] Standard Practice for Gas Chromatography Terms and Relationships, ASTM E 355-
77, reapproved 1983, ASTM, Philadelphia, PA
[2] Jennings, W. (1987) Analytical Gas Chromatography, Academic Press, Orlando,
FL
[3] Burke, J.A., and Giuffrida, L.A. (1964) J. Assoc. Off. Agric. Chem. 47, 326-342
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 502–1
SECTION 502Pesticide Analytical Manual Vol. I
502: COLUMNS
502 A: INTRODUCTION
Separations among analytes in GLC are achieved within the column. Although
choice of detector dictates which class of analytes can be determined, individual
detection and measurement of multiple analytes would not be possible without the
separations provided by the column.
Columns are available in several different physical configurations, each of which
offers advantages and disadvantages to pesticide residue determination. The two
basic types of GLC columns currently used in pesticide residue determination are
(1) packed columns, in which liquid phase is immobilized as a film on particles
of fine mesh solid support and packed into 2-4 mm id columns, and (2) open
tubular capillary columns, in which liquid phase is immobilized as a film on the
interior walls of a capillary tube. Capillary columns are further distinguished by
internal diameter: wide bore (0.53 mm id), traditional (0.25-0.32 mm id), and
narrow bore (≤0.25 mm id). Each type of column requires unique hardware and
operating parameters.
In all GLC columns, identity of the liquid (stationary) phase is the primary factor
dictating what separations are achievable. Carrier gas (mobile phase) is also inte-
gral to GLC operation and must be included in any discussion of columns. How-
ever, only inert gases are used as carrier gases, so few options exist. Operating
parameters that affect column efficiency, including column temperature and car-
rier gas identity and flow rate, provide additional variables that can be adjusted to
achieve separations required for the analysis.
Column Specifications
Descriptions of GLC columns and operating conditions must specify the following:
type of column (packed or capillary); its length, in meters (or feet), and internal
diameter (id), in mm; identity and amount of liquid phase; identity of solid sup-
port, including pretreatments and mesh size (packed columns only); operating
temperature; and carrier gas identity and flow rate.
Liquid phases used in GLC are viscous materials able to be thinly dispersed on
solid support or on an internal column wall. Many different liquid phases are
available, but relatively few are in routine use for pesticide residue determination,
because pesticide residues usually either chromatograph on one of these phases
or are not amenable to GLC. The chemical structure of the most common phases
consists of a polysiloxane backbone with various substituent groups; Figure 502-a
illustrates several of these.
Liquid phase polarity, important to its separation capabilities, varies with polarity
and concentration of substituent group(s) on the polysiloxane. Thus, in terms of
polarity, methyl<5% phenyl<cyanopropylphenyl<50% phenyl<cyanopropyl. The
100% methyl-substituted phase, least polar of those in Figure 502-a, is best suited
to separation of nonpolar analytes; it has been used for many years as a general
purpose phase for a wide variety of pesticide residues. The phase with 50%
cyanopropylphenyl-substitution is the most polar of those shown and is a better
choice for more polar analytes.
Pesticide Analytical Manual Vol. ISECTION 502
502–2
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Figure 502-a
Polysiloxane Stationary Phases
CH3
SiO
CH3
x
100% methyl
(DB-1)
CH3
SiO
CH3
x
x = y
C6H5
SiO
C6H5
y
50% phenyl, 50% methyl
(DB-17)
CH3
SiO
CH3
x
x = y
CH2
SiO
C6H5
y
50% cyanopropylphenyl, 50% methyl
(DB-225)
CH2
CH2
CN
Equivalent products suitable for different column configurations are commercially
available for most common liquid phases; Table 502-a lists some of these products.
Although the table refers to liquid phases themselves, most pesticide residue lab-
oratories no longer purchase liquid phases as materials for preparing columns
in-house. Instead, laboratories that use packed columns usually purchase them
prepacked or at least purchase packing material precoated with liquid phase.
Residue laboratories always purchase commercially prepared capillary columns.
The liquid phase for a particular analysis is selected to take advantage of differ-
ences in chemical and physical properties of the analytes involved. No one liquid
phase is universally applicable to the wide range of chemical and physical proper-
ties found in pesticide residues, so a variety of liquid phases of different polarities
should be available in a residue laboratory.
For packed columns, the amount of liquid phase, often called “liquid load,” is
described as a percentage, i.e., weight liquid phase × 100/(weight liquid phase +
weight solid support). For open tubular columns, the amount of liquid phase is
described as film thickness (μm) of the layer of liquid phase bonded to the inter-
nal wall of the column.
GLC columns are always heated to a temperature at which analytes remain in
the vapor phase. Both isothermal and temperature-programmed operation are
possible. Use of capillary columns with temperature programming is becoming
increasingly common, but this operation will not be described further in this
chapter because FDA has not yet validated its use on an interlaboratory basis.
Maximum operating temperatures vary with specific stationary phases; information
on each is provided by the manufacturer. Increasingly polar stationary phases (e.g.,
cyanopropylphenyl) have significantly lower maximum operating temperatures than
nonpolar phases (e.g., 100% methyl). In use, maximum operating temperature is
usually 20° C higher for temperature programming than for isothermal work.
Column Parameters
The following column characteristics or parameters are commonly used to de-
scribe chromatographic behavior or to measure column performance; terminol-
ogy of these parameters is illustrated in Figure 502-b. Evaluation and comparison
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 502–3
SECTION 502Pesticide Analytical Manual Vol. I
Table 502-a: Common GLC Liquid Phases Used in Pesticide Residue
Determination
Equivalent Commercially Available Products
1
Basic Structure, Substitutions Capillary Open Tubular Packed
Polysiloxane, 100% methyl DB-1(ht), HP-1, HP-101, OV-101, OV-1,
007-1(MS), SP-2100 SP-2100, DC 200,
SPB-1, BP-1, CP-Sil 5CB, CP-Sil 5, SE-30
Ultra 1, RSL-150, RSL-160,
Rtx-1, SP-2100, CB-1, OV-1,
PE-1, SE-30, AT-1
Polysiloxane, 50% phenyl, DB-17 (ht), HP-17, PE-17, OV-17, OV-11,
50% methyl 007-17(MPS-50), AT-50, SP-2250, OV-22,
SP-2250, Rtx-50, RSL-300 DC-710
Polysiloxane, 50% cyanopropyl- DB-225, HP-225, OV-225, OV-225
phenyl, 50% methyl SP-2330, CP-Sil 43CB, RSL-500,
Rtx-225, BP-225, CB-225,
PE-225, 007-225, AT-225
Polysiloxane, 14% cyanopropyl- DB-1701, SPB-7, CP-Sil 19CB, OV-1701
phenyl, 86% methyl Rtx-1701, BP-10, CB-1701,
OV-1701, PE-1701, 007-1701
Polysiloxane, 5% phenyl, DB-5 (ht), HP-5, Ultra-2, OV-3, OV-73, CP-Sil 8
95% methyl OV-5, SPB-5, Rtx-5,
CP-Sil 8CB, RSL-2000,
BP-5, CB-5, PE-5, SE-52,
007-2(MPS-5), SE-54
Polysiloxane, 50% trifluoro- DB-210, RSL-400, SP-2401 OV-210, SP-2401,
propyl, 50% methyl OV-202, OV-215
Polyethylene glycol DB-WAX, HP-20M, Carbowax, Carbowax 20M,
Supelcowax 10, CP-WAX 52CB, Supelcowax 10
SUPEROX II, Stabilwax, BP-20,
CB-WAX, PE-CW
Diethylene glycol succinate No equivalent DEGS (no longer
produced)
1
Commercial codes for each material are related to their manufacturer:
007: Quadrex, New Haven, CT
AT, RSL, SUPEROX: Alltech Associates, Inc., Deerfield, IL
BP: SGE, Inc., Austin, TX
Carbowax: Union Carbide Corp.
CB, CP-Sil, CP-WAX: Chrompak International BV, Middleburg, The Netherlands
DC: Dow Corning Corp., Midland, MI
DB: J&W Scientific, Folsom, CA
DEGS: Analabs, Inc., New Haven, CT
HP and Ultra: Hewlett-Packard, Co., Wilmington, DE
OV: Ohio Valley Specialty Chemical Co., Marietta, OH
PE: Perkin Elmer Corp., Norwalk, CT
Rtx, Stabilwax: Restek Corp., Bellefonte, PA
SE: General Electric
SP, SPB, and Supelcowax: Supelco, Inc., Bellefonte, PA
Pesticide Analytical Manual Vol. ISECTION 502
502–4
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
of columns can be based on such
parameters. More detailed discus-
sion of parameters and conditions
affecting each are found in any
basic chromatography text, such as
those listed in Section 505, Bibli-
ography.
Retention Time. The most basic
measurement in chromatography
is retention time, the time between
sample introduction and elution
of the analyte, measured at the
peak maximum (t
r
in Figure 502-
b). Retention time is corrected for
the time required for a non-
retained solute to reach the de-
tector (tm), often called dead time
or holdup time. Corrected retention time (t′
r
) is the difference between t
r
and t
m
.
For practical convenience, the peak caused by the solvent is used as the nonretained
solute in pesticide residue determinations.
Analyte retention time depends on the extent to which the analyte is retained by
the particular stationary phase under a given set of conditions. Retention time is
constant when column temperature and carrier gas flow are constant, so this
characteristic is the GLC measurement that serves to identify the analyte; it can
be measured electronically in seconds or manually in mm from the resulting
chromatogram. Retention time measured from injection to peak maximum is
often called “absolute retention time.”
Absolute retention time is affected by many column conditions that can vary,
including amount of liquid phase, temperature, carrier gas flow rate, column
length, and system volume. Thus, absolute retention times are insufficiently repro-
ducible to list in tables of data intended to assist an analyst in identifying analytes.
Instead, “relative retention times” are calculated and listed, because they are
far more reproducible from day to day and among different instruments or labo-
ratories.
Relative retention time (rrt) of an analyte is the ratio of its corrected retention
time (t′
r
) to the corrected retention time of a “marker” (reference) compound.
The pesticide chlorpyrifos, molecular formula C
9
H
11
C
13
NO
3
PS, is used in this
manual as the marker compound for most systems, because it chromatographs
well and contains all the heteroatoms to which selective GLC detectors respond;
retention times relative to chlorpyrifos (rrt
c
) for many pesticides and related com-
pounds are listed in Appendix I, PESTDATA.
For the same (or equivalent) liquid phase, rrt of an analyte is independent of
column type (packed vs capillary), liquid load, column length, or carrier gas flow
rate change. The rrts for a particular liquid phase vary significantly only with
column temperature; rrt
c
s in Appendix I are valid only at the temperature speci-
fied for each column.
Capacity Factor. Capacity factor describes the retentive behavior of a sample com-
ponent relative to the “retentive behavior” of a nonretained component. The
Figure 502-b
GLC Column Parameters
Column parameters are calculated from measurements
on a chromatogram produced by the column.
Inject
Nonretained
solute
t
m
t′
r
t
r
W
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 502–5
SECTION 502Pesticide Analytical Manual Vol. I
capacity factor of an analyte depends only on the time the analyte spends in the
stationary phase, which is, chromatographically speaking, far more important than
the time spent in the mobile phase. The capacity factor (k) of an analyte is cal-
culated from analyte retention time as k = (t
r
-t
m
)/t
m
.
(Capacity factor should not be confused with “sample capacity,” which describes
the maximum amount (e.g., 50 ng) of an analyte that can be injected onto a
chromatograph before column overload occurs. Column sample capacity depends
on percent liquid load in packed columns and on column id and film thickness
in capillary columns.)
Selectivity. Stationary phase selectivity is simply defined as the ability of a phase
to differentiate between analytes in the same injection. The selectivity term is
technically not interchangeable with polarity [1]. A polar column may exhibit very
poor selectivity for a particular chemical species. In general, nonpolar stationary
phases exhibit greatest selectivity for nonpolar analytes, and polar stationary phases
exhibit greatest selectivity for polar analytes. Selectivity of a GLC system is defined
by both the stationary phase and the analytes. In the literature, selectivity (α) is
synonymous with separation factor, relative retention, and selectivity factor and is
calculated as k
B
/k
A
, where k
B
and k
A
are capacity factors of two adjacent peaks. In
this calculation, a is always ≥1.0, but a separation factor of 1.0 indicates that no
separation is possible in that system [2].
Resolution. Resolution is the degree of separation between two chromatographic
peaks and is related to time (capacity factor), selectivity, and efficiency. Consider-
able information about resolution and its related parameters is available in general
textbooks on chromatography. For practical purposes, however, it is enough to
know that optimizing selectivity by choice of stationary phase will optimize resolu-
tion. Despite the importance of column efficiency in analyzing complex samples,
especially at low levels, increasing efficiency will not solve all separation problems
and often will only increase analysis time. A different choice of stationary phase
may solve a resolution problem more easily than a longer column will. Resolution
is considered optimized when calculated k values range between 2-10.
Efficiency. In qualitative terms, column efficiency refers to the degree to which
injected analyte is able to travel through the column in a narrow band. Visually,
a more efficient column produces narrower, sharper peaks on the chromatogram.
The more efficient the column, the better able it is to resolve analytes that elute
close to one another. Greater efficiency results in greater signal-to-noise ratio and
hence increases sensitivity. Efficiency is measured quantitatively by calculating
theoretical plates according to the formula
N = 16 (RT/w)
2
where N = total theoretical plates, RT = absolute retention time in mm, and w =
width of peak base in mm, measured as the distance at the baseline between lines
drawn tangent to the two sides of the peak. The analyte on which theoretical
plates are calculated must be specified, because comparisons are only valid for
analytes eluting at the same absolute retention time. Column efficiency can also
be expressed as height equivalent of one theoretical plate (HETP), i.e., column
length (cm)/N; using this expression, smaller numbers represent more efficient
columns. Calculation of theoretical plates/column length permits comparisons of
different length columns.
Pesticide Analytical Manual Vol. ISECTION 502
502–6
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Basic GLC texts, such as those listed in Section 505, provide additional explana-
tions about theoretical plate measurements, qualitative effect of column efficiency
on peak shape, and practical means of improving peak shape. Column efficiency
is referred to in this chapter when discussing relative advantages of different types
of columns.
502 B: PACKED COLUMNS
During most of the over 30 years of GLC use in pesticide residue determination,
packed column GLC prevailed as the only practical option. During early develop-
ment of open tubular capillary columns, when only traditional capillaries were
available, packed columns offered distinct advantages in ease of use and capacity
for injection of larger volumes of extract. Current availability of wide bore capil-
lary columns has reversed the trend, however, and use of packed columns is
diminishing.
Packed columns still offer advantages in ease of installation; no additional inlet
adapters or other specialized hardware are needed to install packed columns into
chromatographs designed for packed column operation. Packed columns can also
still withstand repeated injections of extract better than capillary columns. How-
ever, recent improvements in inlet systems and operating parameters for wide
bore columns have increased their capacity for injected extract. Combined with
the innately greater efficiency and inertness of wide bore columns, these improve-
ments are encouraging the shift from packed to wide bore columns for routine
use.
Components of Packed Columns
Packed columns consist of packing material made by coating inert solid support
with a thin film of stationary liquid phase, glass or metal tubing to contain the
packing material, and silanized glass wool plugs used to hold the packing material
in place within the tubing.
Solid Support. The solid support in packed GLC columns provides a large inert
surface onto which the stationary liquid phase is deposited as a relatively uniform
thin film. Solid support should provide as large a surface area as possible and
should interact as little as possible with analytes. Desirable properties of solid
supports are large surface area per unit volume, chemical inertness at high tem-
peratures, mechanical strength, thermal stability, ability to be wetted uniformly by
a stationary liquid phase, and ability to hold a liquid phase strongly.
The most frequently used solid supports for GLC column packings are derived
from diatomaceous earth. The structure of the diatomaceous earth consists essen-
tially of three-dimensional lattices containing silicon with active hydroxyl and oxide
groups on the surface. Untreated diatomaceous earth has considerable surface
activity that must be reduced before it becomes a suitable support material. Sev-
eral techniques have been used to deactivate the surface activity of diatomaceous
earth. Most frequently, the diatomaceous earth is acid-washed and then silanized
with an agent such as dimethyldichlorosilane.
Different commercially available solid supports and even different lots of the same
support may have different surface areas or variations in inertness toward particu-
lar analytes. Unpredictable behavior among solid supports provided the impetus
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 502–7
SECTION 502Pesticide Analytical Manual Vol. I
for most laboratories to purchase precoated packing. Variations in solid support
activity are of greatest concern when determining pesticide residues that are dif-
ficult to chromatograph, because such analytes are easily adsorbed or degraded
during chromatography. Adsorption or degradation of an analyte on a poor qual-
ity solid support can affect the relative retention time of the analyte and the size
and shape of the resulting peak. The most inert solid support material available
should always be used to prepare column packings.
Chromatographic solid supports are available in a variety of mesh sizes. A support
material of 80/100 mesh contains particles that will pass through an 80-mesh
screen but not through a 100-mesh screen. Experiments have shown that column
efficiency improves as solid support mesh number increases (particle size decreases)
[3]. However, to maintain the same gas flow through a column, carrier gas pres-
sure must be increased as solid support particle size decreases. Mesh size of 100/
120 was shown to produce optimum efficiency for a 6' column of 4 mm id. Col-
umns 4-6' long and 2-4 mm id, filled with column packings prepared from 80/100
or 100/120 mesh solid supports, are routinely used for residue determination.
Liquid Phase Load. No matter what liquid stationary phase is used, liquid load
influences column efficiency and capacity (amount of sample extract that can be
injected onto the column). Packing materials with loads ranging from <1 to 5%
are routinely used for pesticide residue determination.
Liquid phase load can be varied without changing relative retention times of
compounds if the same column temperature is used. At the same column tempera-
ture and gas flow, a column with less liquid phase will allow compounds to elute
more quickly than a higher load column. Carrier gas flow can be lowered when
using columns with less liquid phase to permit compounds to elute at approxi-
mately the same time as from higher load columns operated at higher gas flows.
Laboratory observations indicate that compounds with a tendency to degrade on
or be adsorbed by a column are more likely to do so when a lower liquid phase
load is used, probably because the lower load is incapable of covering all solid
support active sites. In these cases, analyte retention time and peak size will be
affected, as described above. Residue analysts should be aware of the pitfalls of low
load columns when dealing with compounds that are easily degraded or adsorbed.
Column Tubing. Almost all columns used in pesticide residue determinations are
made from glass tubing. Although some gas chromatographs require metal col-
umns, so many problems occur with metal that they should be avoided. In the
past, new glass columns had to be cleaned and silanized in the laboratory to
remove any residual caustic materials and to deactivate the column. Today, most
glass columns are silanized by the manufacturer and are purchased ready to use.
Inadequate deactivation of glass columns can cause peak tailing due to adsorption
or degradation of the sample or standard on the active sites of the column itself.
Glass Wool. Glass wool for use in GLC columns must be silanized to prevent
compound adsorption; presilanized glass wool is available commercially or
silanization can be performed by the laboratory. A plug of silanized glass wool is
always used at the outlet (detector) end of a packed column to hold the packing
material in place. Glass wool can also be used in the inlet end of a packed column,
but opinions vary about the advisability of this practice.
Pesticide Analytical Manual Vol. ISECTION 502
502–8
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Used in the inlet end of a packed column, glass wool can cause adsorption or
degradation of certain sensitive compounds. Problems with normally stable com-
pounds can also occur when deposits of sample co-extractives collect on the glass
wool.
In particular, when deposits of fatty extracts accumulate at the top of the column,
analytes in subsequent injections can be partially trapped; errors in residue
quantitation result. Elimination of glass wool at the inlet end of the column ap-
pears to minimize this problem by allowing injected co-extractives to spread over
a portion of the column where subsequent analytes cannot be trapped so readily.
In other cases, glass wool in the inlet end of the column may prevent the rapid
deterioration of columns caused by injecting co-extractives from fatty foods or
other commodities that are difficult to clean up. Co-extractives trapped on the
glass wool plug can be eliminated by replacing the plug, an easier, quicker, and
less expensive process than replacing the packing material.
Choosing whether to use glass wool in the inlet end of the column appears to
depend on several factors, including type of packing material used, commodity
being analyzed, analytes of interest, type of detector, and method of analysis.
Experience will dictate when the advantages of glass wool in the column inlet
outweigh the disadvantages; a laboratory attempting to locate the source of prob-
lems in a GLC determination should definitely investigate the effects of glass wool
in the column inlet.
Preparation of Packed Columns
Acceptable techniques for packing empty GLC columns are designed to fill the
column with as much packing material as possible (i.e., to pack the material as
tightly as possible) while breaking the fewest particles. Column efficiency increases
with the amount of properly coated support in the column, and adsorption and
degradation problems are minimized when careful handling of the packing mate-
rial creates the fewest broken (active) sites.
Poor packing technique causes visible differences in column performance (effi-
ciency) and peak symmetry. Loosely packed columns or columns containing too
little column packing are inefficient and a cause of inadequate separations. On
the other hand, a column packed too tightly requires excessive carrier gas pres-
sure, which can result in the column becoming plugged with broken particles.
To pack a glass column:
? Insert about 1-2" silanized glass wool into detector end of column, far
enough from end to prevent packing material from extending into detec-
tor base where temperatures are usually much higher than column oper-
ating temperature.
? Use rubber tubing to connect detector end of column to vacuum source
(aspirator or vacuum pump); attach funnel with short piece of rubber
tubing to inlet end of column.
? Apply partial vacuum at detector end of column, and slowly add prepared
packing material through funnel.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 502–9
SECTION 502Pesticide Analytical Manual Vol. I
? Tap column gently while adding packing material, to settle it as tightly
as possible; do not use a vibrator to help settle packing.
? Continue to tap gently along entire length of column while adding
more packing, until column is full or within 1" of being filled at inlet
end.
To reuse a glass column:
? Remove old packing.
? Rinse empty column successively with 5% potassium hydroxide/metha-
nol and dilute hydrochloric acid.
? Rinse empty column thoroughly with successive portions water, alcohol,
and ethyl acetate to eliminate accumulations of liquid phase and/or
sample co-extractives from column walls.
? Dry empty column before repacking.
Installation of Packed Columns
In modern GLC equipment, glass columns, filled with packing material, are con-
nected directly to the detector (metal) and injector (metal), an arrangement
that eliminates dead space in the system. The availability of ferrules that are ther-
mally stable at high temperatures makes these glass-to-metal connections possible
and eliminates problems once associated with such connections. Ferrules with
these capabilities include those made from Vespel, graphite, or Vespel/graphite
mixture.
To install a glass column in the chromatograph:
? Slide stainless steel or brass (usually 1/4"
Swagelok) nuts onto detector and inlet
ends of column followed by ferrule, as
shown in Figure 502-c.
? Connect nuts on column to correspond-
ing hardware on injector and detector.
? Tighten each nut finger tight.
? To seal ferrules, alternately tighten
detector nut and injector nut using
standard wrench, following instructions
provided by ferrule manufacturer. Un-
evenly exerted pressure on either end
of column may twist and break it.
? Turn on carrier gas (30-60 mL/min) and flush with gas for about 20 min.
? After all oxygen has been flushed from column by flow of carrier gas
(and not before), turn on column oven to heat column.
Figure 502-c
Ferrules for Connecting Glass
and Metal
[Reprinted with permission of Supelco, Inc., adapted
from Supeltex M-1 Packed Column Ferrules data sheet
(1987).]
Glass column
Ferrule
Metal nut
Pesticide Analytical Manual Vol. ISECTION 502
502–10
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
? Check connections for leaks after establishing carrier gas flow.
During installation, always hold the bottom of the column for support. Do not
overtighten the nuts, which can force the column against the bottom of the injec-
tor/detector and break the column. Refer to the instruction manuals provided by
the instrument manufacturer for more specific instructions on column installa-
tion.
Conditioning of Packed Columns
Column “bleed” is degradation of stationary liquid phase that causes a background
signal as the detector responds to its presence. Column bleed occurs in all col-
umns and is not in itself a symptom of damage. However, excessive or increasing
bleed, seen as a rise in baseline, may be caused by damage to the column. Bleed
increases when the column is operated at higher temperatures, and damage may
be caused by operation at temperatures higher than allowed for a particular
stationary phase.
To minimize column bleed, newly packed columns must be conditioned before
they are connected to the detector; conditioning purges volatile components that
could contaminate the detector and produce an unsteady baseline. Column con-
ditioning involves heating the column above normal operating temperatures for
an extended period prior to its use. The column must not be connected to the
detector during conditioning. In most cases, a normal carrier gas flow is main-
tained during column conditioning. Excessively high conditioning temperatures
will shorten column life.
Minimizing column bleed by conditioning is essential to good operation. If liquid
phase is bleeding from the column, frequent detector cleaning will be necessary,
sensitivity of the GLC system will change, quantitative results will probably be
affected, baselines will drift, and good quality chromatograms will not be obtained.
“Stabilized” liquid phases are designed to be more thermally stable than their
nonstabilized equivalents, because they bleed less at normal operating conditions.
However, conditioning of stabilized packings is still required before use.
Conditioning procedures vary with the type of column packing and are provided
by the manufacturer in the literature supplied with the packing.
Rejuvenation of Packed Columns
Column deterioration during use is most often caused by inadequate cleanup
of samples injected onto the column (see Section 501 C). Extracts of materials
containing large amounts of fats or oils (e.g., dairy products, animal tissue, and fish
oils) are difficult to clean up thoroughly. Injection of excessive amounts of oily
extract can cause irreversible damage to a GLC column. Waxy or colored material
co-extracted from nonfatty foods may also damage the GLC column, but this effect
is not as readily apparent as that caused by oily co-extractives. Care should be
taken to minimize the amount of any co-extractive material injected, including the
use of additional or alternative cleanup techniques when original cleanup is inad-
equate.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 502–11
SECTION 502Pesticide Analytical Manual Vol. I
No matter how rigorous the sample cleanup, some accumulation of co-extractives
on the column will occur. To prevent column deterioration, the column must be
periodically cleaned. Packed columns are most often cleaned by removing up to
several inches of packing at the inlet end of the column and replacing it with new
(preferably conditioned) packing. To perform this operation:
? Turn off column oven heat and permit oven to cool.
? When oven is cool, turn off carrier gas and remove column.
? Remove contaminated packing with disposable pipet or other device,
and swab inside of glass column with acetone using pipe cleaner or
other appropriate device to remove fatty deposits or other matrix con-
tamination that have adhered to interior column wall.
? Add fresh packing to column, in same way described above for packing
new columns.
Criteria for Acceptable Packed Columns
Column performance must meet the following criteria for successful pesticide
residue determination. Exact performance will vary somewhat as the column ages,
but minimum criteria should be met through its lifetime; when the column no
longer meets these standards, it should be replaced.
Some of these criteria relate to careful column preparation and conditioning and
are important to check when the column is new. Others relate to the potential for
gradual column deterioration and contamination during use. Some other part of
the GLC system may be responsible for the system’s failure to meet criteria, so all
parts should be examined when the system is malfunctioning.
1) Chromatography of selected compounds should result in a single sym-
metrical peak with no breakdown. Endrin frequently chromatographs
as two or three peaks when columns are not satisfactory, and methoxy-
chlor breaks down to its olefin. DDT deteriorates to TDE or DDE or may
be lost entirely on a contaminated column. None of these conditions
should be tolerated.
2) Peak resolution of selected compounds should be complete. For ex-
ample, dieldrin and endrin can be separated from one another on most
columns that are performing well; a mixture of the two should be
chromatographed routinely to monitor changes in resolution as the
column ages.
3) Peak heights for several compounds should be reproducible when re-
petitive injections are made. Poor reproducibility (≥5%) can have sev-
eral causes external to the column: improper injection technique, a
faulty syringe, a faulty septum, or detector malfunction. Poor reproduc-
ibility can also indicate breakdown or adsorption of the compound on
the column. Compounds used to test the column for general acceptabil-
ity are those that may break down or be adsorbed by columns but can
be successfully chromatographed, such as endrin. When a column is
used to analyze for compounds that are hard to chromatograph, it should
first be checked with a compound such as endrin.
Pesticide Analytical Manual Vol. ISECTION 502
502–12
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Sometimes, injection of large concentrations of compounds that are
difficult to chromatograph may improve their chromatography. Some
pesticide chemicals may not chromatograph well until a column is more
thoroughly conditioned by prolonged use. If a compound shows tailing,
or little or no response, or if multiple peaks are obtained from injection
of a single standard of known purity, adsorption, degradation, or some
other column effect may be the cause. Chromatography of some com-
pounds may not be satisfactory until the column has been used exten-
sively.
4) Instrument response to varying amounts of a compound should be linear.
A nonlinear response can have many causes, but breakdown or adsorption
of the compound on the column may be indicated when the system is
linear for some compounds but not for others. It is especially important
to ascertain linearity for each compound of interest when the compounds
are difficult to chromatograph.
5) A 4 mm id packed column should have about 500 theoretical plates/
foot of column length, as measured on a peak produced by p,p′-DDT.
(Retention time of the peak used affects theoretical plate calculation, so
measurement of the p,p′-DDT peak at whatever time it elutes from an
individual column is an admitted oversimplification, but is adequate for
the purpose defined here.)
Theoretical plate counts <500 do not necessarily render a column unac-
ceptable, but performance of columns with <400 plates/foot should be
closely observed. Routine measurement of theoretical plates will alert the
analyst to unsatisfactory new columns or to deterioration of columns al-
ready in use and is recommended as a part of the routine check on
instrument performance.
Recommended Operating Procedures for Packed Columns
Each GLC determinative step in Chapters 3 and 4 is described in terms of its
specifications and operating conditions. Most of these describe wide bore capillary
columns, now recommended for routine use in pesticide residue determination;
only Sections 302 DG20-DG23 describe systems with packed columns, because the
DEGS column of those modules has no wide bore equivalent. However, most GLC
data (rrts and responses) included in Appendix I, PESTDATA, were developed
with packed columns during the many years in which they were in use. Table 502-
b provides operating conditions for packed columns useful in pesticide residue
determination.
Column liquid phase and temperature are dictated by the analytes being sought
in a particular method. Choice of carrier gas depends on the requirements of the
detector; in some cases, argon/methane is used to accommodate the
63
Ni electron
capture detector. Carrier gas flow rate is typically 30-60 mL/min. Injection volume
is typically 3-8 μL.
Columns must be protected from damage that can occur when the stationary
phase is exposed to oxygen at high temperature. Increased bleed of degradation
products from oxidation will occur, and the phase can be damaged permanently.
After any exposure to air, e.g., during septum change, the column should be
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 502–13
SECTION 502Pesticide Analytical Manual Vol. I
Table 502-b: Operating Conditions for Packed Columns
1
OV-101 OV-17 OV-225
Liquid load, % 5 3 3
Injector temperature, ° C 225 220 225
Target rrt
c
2
3.1 ± 0.06 3.5 ± 0.07 3.6 ± 0.06
(marker compound) (p,p′-DDT) (p,p′-DDT) ( p,p′-DDT)
2.56 ± 0.05 3.36 ± 0.07 3.9 ± 0.1
(ethion) (ethion) (ethion)
0.69 ± 0.02
(lindane)
Elution time, min
3
4 4 5.5
Conditioning
4
1° C/min to 250° 1° C/min to 250° 50° C;
250° C, ≥16 hr 250° C, ≥72 hr 2° C/min to 245°
245° C, ≥60 hr
Special requirements Do not use with
ElCD or N/P
1
All columns are: 1.8 m × 2 or 4 mm id; liquid phase coated on 80/100 mesh Chromosorb W HP, or equivalent.
2
Column temperature is 200° C, adjusted as needed to produce specified rrt
c
for marker compound.
3
Approximate elution time of chlorpyrifos with carrier gas (nitrogen, helium, or argon/methane, as required by
detector) at about 60 mL/min.
4
Conditioning performed with column disconnected from detector. Degas with 60 mL/min nitrogen for 1 hr;
temperature program as specified; hold at specified temperature with nitrogen flowing for specified time period.
checked for leaks and then flushed with carrier gas for 15-20 min before restoring
the column to operating temperature.
If it is necessary to change carrier gas tanks while the column remains at operating
temperature, interruption of column carrier gas flow can be avoided by turning
off secondary valve pressure, which is usually at 40-80 psi. While the gas flow
continues bleeding into the column, the main tank valve can be turned off and
the regulator moved to a new tank.
502 C: OPEN TUBULAR CAPILLARY COLUMNS
Capillary column GLC has existed almost as long as packed column GLC and is
now preferred for determining pesticide residues in foods. Capillary columns
provide greater inertness, chemical and thermal stability, column efficiency (and
thus system sensitivity), resolution, operating temperature range, and column-to-
column reproducibility than equivalent packed columns.
In addition to the nature of the liquid phase, many factors affect capillary column
performance and applicability, including bore size, film thickness, operating
temperature, column length, and carrier gas identity and flow rate. Most often, a
change in one column parameter improves some features of column performance
and diminishes others, so choice of column for a particular analysis is based on
an assessment of the most important feature(s).
Pesticide Analytical Manual Vol. ISECTION 502
502–14
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Capillary columns are available with internal diameters ranging from 0.050-0.53
mm. Efficiency increases as capillary column bore size decreases. Traditional (0.25-
0.32 mm id) and narrow bore (<0.25 mm id) capillary columns are noted for
extraordinary efficiency (≥5000 theoretical plates/m), which improves signal-to-
noise ratio and thus sensitivity. High efficiency also provides the improved resolu-
tion necessary for analyses of complex samples. However, sample capacity de-
creases with decreasing bore size, and columns become less forgiving of improper
handling. In addition, the low carrier gas flow rates used (≤0.9 mL/min for narrow
bore and ≤3 mL/min for traditional) require specialized flow control hardware.
For certain determinations, advantages offered by narrow bore columns outweigh
their disadvantages. Thin film, narrow bore capillary columns are ideal for special-
ized “ultra trace” determinations at levels of part per trillion and below, e.g., for
determination of dioxin residues. Once adjustments are made to accommodate
requirements of narrow bore columns related to gas flow, sample capacity, and
injection technique, they provide the ultimate efficiency, resolution, and sensitivity
needed for these determinations.
The low gas flows required with narrow bore capillary columns also make them
the best choice for use in certain instruments. For example, interfacing narrow
bore columns directly to mass spectrometers has become an industry standard,
because the low flow is compatible with the requirements imposed by vacuum
conditions within the spectrometer (≤1 mL/min maximum flow). Use of narrow
bore columns obviates the need to divert carrier gas before effluent reaches the
spectrometer.
In contrast, either wide bore (0.53 mm id) or traditional capillary columns are
preferred for routine pesticide residue determination, with wide bore the most
popular. Although wide bore columns are less efficient than narrow bore, they
offer greater sample capacity; depending on the film thickness, wide bore columns
may have sample capacities comparable to packed columns. Carrier gas flow of ≤6
mL/min is recommended for optimum efficiency, but if this results in excessively
long analysis time, the larger internal diameter of a wide bore column can accom-
modate 20-30 mL/min without generating excessive column head pressure. Wide
bore columns can be operated at these higher gas flows (“packed column condi-
tions”) without the specialized pneumatics required for low flow rates. Perfor-
mance of wide bore columns can be optimized by changes in carrier gas flow rates
and other system parameters, such as injection technique.
Column Description
Early open tubular capillary columns were made from glass, with liquid (station-
ary) phase coating the interior wall. These columns were fragile and subject to
significant liquid phase bleed. The columns assumed the shape of the “cage” on
which they were mounted and thus required straightening before inserting the
ends in inlets or detectors. A high degree of operator skill was necessary for their
use.
The disadvantages of capillary columns were minimized or eliminated when
several features were vastly improved. Columns are now made from fused silica, a
synthetic quartz, coated on the outside with polyimide, which makes them
rugged, flexible, and easy to handle. Stationary phases are now cross-linked poly-
mers bonded to the interior column wall, effectively eliminating column bleed.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 502–15
SECTION 502Pesticide Analytical Manual Vol. I
Figure 502-d
Capillary Column Cross-Section
Drawing (not to scale) of GLC capillary
columns; id of fused silica tubing ranges
from 0.05-0.53 mm.
Polyimide coating
Fused silica tubing
Cross-linked
polymer liquid
phase
Figure 502-d shows a cross-sectional view of
a typical modern open tubular capillary col-
umn.
Tubing for capillary columns is produced by
drawing fused silica through a furnace. The
exterior of the drawn capillary tubing is then
coated with a plastic polyimide coating and
the interior cleaned and deactivated. Exact
processes used by manufacturers are propri-
etary and beyond the scope of this chapter.
The resultant tubing is very flexible, rugged,
and reasonably inert and requires only mini-
mal care in handling. It is easily cut and may
be coiled around cages and flexed as neces-
sary for instrument connections. When re-
leased, the tubing straightens, simplifying
connections. Capillary columns are available in lengths from 10-60 m; 15 m or 30
m columns are usually used for determination of pesticide residues in foods.
Stationary phases are no longer simply coated on the interior walls. Individual
stationary phase polymer “strands” are cross-linked, and the cross-linked stationary
phase is covalently bonded to the deactivated interior wall of the tubing by pro-
prietary processes. Columns prepared in this manner are more thermally stable
than coated phases, so they can be operated at higher temperatures; they are also
more efficient. Cross-linked phases exhibit minimal bleed and resist being stripped
by solvent, to the degree that they can be rinsed with solvent to remove nonvolatile
contaminants. The process of cross-linking also facilitates preparation of thicker
films (i.e., 1.0-8.0 μm) that are otherwise difficult to prepare. Chemically, the
stationary phases are equivalent to those coated on solid support for packed col-
umn use, so relative retention times for analytes are essentially the same in equiva-
lent packed and capillary columns, as long as column temperature is the same [4].
Capillary columns are available with films ranging from 0.10-5.0 μm thick. Col-
umns with <0.32 mm id usually have film thickness of 0.10-1.0 μm, while those of
≥0.32 mm id have films 0.l-≥5.0 μm. Film thickness is proportional to sample
(analyte) capacity, i.e., thicker films accommodate more analyte without overload.
Theoretically, a 0.53 mm id column has a sample capacity of 53, 130, 530, and
2600 ng for film thicknesses of 0.1, 0.2, 1.0, and 5.0 μm, respectively; sample
capacity for a 0.25 mm id capillary column is about half as much for each film
thickness [2].
Column efficiency, however, is inversely proportional to film thickness. Thick film
columns are also more retentive than thin film columns, so retention times are
longer and analyte peaks broader on the former. Thick film columns are also
more susceptible to column bleed.
Because polar stationary phases (e.g., cyanopropylphenyl) are difficult to coat onto
column walls, they are usually only available in film thicknesses up to 1.0 μm. Polar
stationary phases tend to bleed more than their nonpolar counterparts even under
ideal conditions.
Film thickness for columns used in pesticide determination is normally 1.0 or 1.5
μm, which provides optimum balance between phase thermal stability, analyte
Pesticide Analytical Manual Vol. ISECTION 502
502–16
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
retention, and analyte column capacity. Relationships between film thickness and
column efficiency, thermal stability, analyte retention, and capacity are discussed
in detail in most modern GLC books (Section 505).
Each stationary phase has upper and lower temperature limits that define the
operating range, but only the upper limit is of concern in pesticide determination.
Operation at temperatures exceeding the upper limit accelerates phase degrada-
tion. Heating a column without carrier gas flow, or exposing it to any oxygen at
temperatures ≥100° C, even for short periods, can damage phases rapidly and
irreversibly.
Injection onto Capillary Columns
The small internal diameter of all capillary columns imposes specific requirements
on how injection is performed; the narrower the diameter, the more rigid the
requirements. (These injection options should not be confused with injection
techniques discussed in Section 501 D. That section covers choices between manual
and automatic injection and among various techniques for handling syringes. This
section refers to ways to accommodate injection and vaporization of solvent into
the restricted space available in capillary columns.)
Extensive research into means of introducing solutions onto capillary columns has
produced four major injection techniques, called split, splitless, on-column, and
direct. Each has advantages and disadvantages, and each has found uses in particu-
lar GLC applications. FDA studies, however, support recommendations that pesti-
cide residue GLC determinations be performed with direct injection, using a re-
tention gap, onto wide bore capillary columns [5]. This system eliminates or
minimizes problems such as band broadening, peak splitting, and intolerance to
variable injection volumes [6-9]. Direct injection involves introduction of the sample
into a hot, vaporizing inlet with total transfer (no splitting) of injected materials
onto the analytical column. GLC inlets designed for packed columns are easily
converted to use with direct injection; kits for this purpose are commercially avail-
able. Injection volumes of 0.5-6.0 μL are used with direct injection.
Direct injection is not suitable for use with narrow bore columns or low gas flows,
so references such as those in Section 505 should be studied for further informa-
tion on the other injection techniques not covered here.
Capillary Column Systems
The practical necessities of residue determination require that a minimum weight
of sample equivalent be examined by the determinative step. When capillary col-
umn GLC is used for determination, provision must be made to ensure that the
volume of extract needed for injection of this weight does not overwhelm the
capacity of the column. The following arrangements are required to accommodate
physical limitations imposed by capillary columns. Because requirements become
more stringent as internal diameter decreases, different recommendations may
apply to wide bore and traditional capillary columns.
Retention Gaps (Guard Columns). Use of a “retention gap” [10] is recommended
for capillary column GLC used in pesticide residue determination. A retention gap
is a segment of deactivated fused silica tubing (without stationary phase) that is
placed between the instrument inlet and the top of the capillary column; in effect,
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 502–17
SECTION 502Pesticide Analytical Manual Vol. I
it serves as an extension of the column inlet. Tubing 0.53 mm id and 1-5 m long
is commonly used; a length of 5 m is recommended for pesticide residue deter-
mination.
A retention gap serves at least two purposes: (1) It provides space for the injected
solution to vaporize and expand, thus permitting injection of solvent volumes (>1
μL) that could not otherwise be injected into capillary tubing; and (2) it provides
surface area for deposition of co-extractives, thereby protecting the analytical col-
umn from buildup of nonvolatiles that can cause loss of efficiency and analyte
decomposition or adsorption; in this role, the retention gap is often called a
“guard column.” Properly installed, a retention gap will not noticeably reduce
column efficiency.
Inlet Adapters. Direct injection of extracts and standard solutions onto capillary
GLC columns requires a glass adapter to minimize analyte contact with hot metal
surfaces. Adapter design has gradually evolved to meet the practical needs
of trace level determinations.
Figure 502-e shows three styles of inlet
adapters evaluated for use with direct
injection. Adapter 1, the straight tube
adapter, is simplest. A capillary column
or retention gap is inserted into the
bottom of this adapter with a stainless
steel reducing union. This adapter,
containing a small glass wool plug, was
successfully used with various wide bore
columns to determine pesticide resi-
dues in foods analyzed by the method
described in Section 302 [11]. How-
ever, this style adapter is not recom-
mended, because it allows exposure of
analytes to the hot metal reducing
union. In addition, injection of large
volumes can result in flashback of sol-
vent vapors and analytes into the in-
strument pneumatic systems.
Adapters designed with tapered
restrictors at the point where the col-
umn or retention gap connects are
preferable to straight tube adapters; this
design eliminates contact of analytes
with the hot metal reducing union.
Adapter 2 displays a commercially avail-
able direct flash injection liner [12]
with a nontapered restrictor at the top
and a tapered restrictor below for con-
nection to a column. The top restrictor
minimizes both flashback during injection and contact of analytes and solvent with
the septum area of the inlet. This adapter was successfully used with extracts from
Section 302 [13], and its performance was validated with an interlaboratory trial
involving similar extracts cleaned up with Florisil [14]. The only drawback with
this adapter is difficulty in cleaning.
Figure 502-e
Inlet Adapters for Capillary Columns
Inlet adapter used with capillary columns: (1)
straight, (2) adapter with restrictors, and
(3) adapter with disposable liner.
(2) (3)(1)
Disposable
liner
Pesticide Analytical Manual Vol. ISECTION 502
502–18
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
A major improvement in inlet adapters is the addition of an easily replaced
disposable liner. Originally, an adapter intended for use with on-column
injection (not pictured in Figure 502-e) was inverted and modified to include a
disposable Supelco PureCol? inlet liner, a small amount of column packing, and
a small glass wool plug for determination of organochlorine pesticides in fatty
foods [15]. Subsequently, during application to analysis of nonfatty foods, the
column packing material was found to be unnecessary. Successful application of
this adapter-liner combination led to commercial production of Adapter 3, de-
signed specifically for use with a replaceable liner, as shown.
Use of a liner protects the inlet adapter, because nonvolatile co-extractives deposit
on the liner rather than the adapter. A contaminated liner is easily replaced
without disturbing the connection between the adapter and the column; depend-
ing on instrument design, the liner is changed after removing the septum or after
removing the adapter-liner combination from the GLC inlet.
Chromatographic efficiency using any of these adapters will deteriorate with re-
peated injections of food extracts. Efficiency can be restored by removing the
adapter from the instrument, cleaning, and resilanizing. After resilanizing, the
adapter (without a column attached) should be heated overnight to normal inlet
temperature with 10 mL/min gas flow to remove excess silanizing reagent.
Septa. In GLC, injections are made by microliter syringes through septa made of
materials that permit passage of a needle and then reseal after the needle is
withdrawn. For troubleshooting purposes, chromatographers must be aware of the
problems that can be caused by septa. Each septum has a limited useful life, after
which it leaks and must be replaced. Leaking septa cause inaccuracies in
quantitation, problems with chromatography, and exposure of the system to air.
Materials from which septa are made can contribute to system bleed and/or can
become brittle with use. Shards from damaged septa can also pass into wide bore
columns and block gas flow.
Connections. Any adapter installed in the GLC inlet is sealed by means of a nut
and high temperature ferrule. Ferrules are available in various sizes, shapes, and
materials; GLC instrument manufacturers specify requirements for ferrules to be
used in each instrument. Typically, ferrules of 100% graphite are used, though
ferrules consisting of graphite and Vespel are also common and sometimes re-
quired (e.g., graphite/Vespel is used in GC-MS because graphite ferrules out-gas).
Analytical columns are connected to retention gaps with “low dead volume” or
“zero dead volume” butt connectors. Various styles are available from chromato-
graphic supply companies, including ferrule, adhesive, and “press-in” types. The
simplest and least expensive are the press-in types, in which each tube is pushed
into opposite ends of a flared connector to form a seal. Press-in connectors are
suitable for most applications and are ideal for connecting 0.53 mm retention
gaps to smaller diameter analytical columns. The “universal” style of press-in con-
nectors, i.e., those that connect tubing of any sizes, have been found to work best.
The two most critical connections in capillary GLC are those that connect the
analytical column to the inlet and to the detector. Connections not only must be
leak-free, but positioning is critical to optimum performance. Most manufacturers
of gas chromatographs provide detailed instructions for positioning the column
outlet at a specific location in detectors. These instructions must be followed
exactly. If the manufacturer’s instructions for proper positioning of capillary
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 502–19
SECTION 502Pesticide Analytical Manual Vol. I
column ends are not available, optimum positioning must be determined experi-
mentally.
Inlet column positioning is also critical but is simplified with inlet adapters or
liners that have tapered glass restrictors for column seating. The restriction is
positioned at the proper location in the inlet, and the column is inserted firmly
to form a seal between the polyimide coating and tapered fitting. Adapters or
liners with no tapered glass fitting require careful, precise measurements for proper
column positioning. Manufacturer’s instructions for positioning tubing in inlet
adapters or liners must be followed exactly.
Carrier Gases. Hydrogen and helium are the carrier gases of choice with capillary
columns, because their flow rates can be increased with less loss of efficiency than
is seen with nitrogen. The
van Deemter curves for ni-
trogen, helium, and hy-
drogen (Figure 502-f) dis-
play the effect on column
efficiency (HETP) of in-
creasing average linear ve-
locity (cm/sec, calculated
as column length in cm/
retention time in sec, of
an unretained peak) in a
typical capillary column.
Minimum HETP (i.e.,
maximum efficiency) for
nitrogen carrier gas occurs
at very low linear velocity
(flow rate) and over a
narrow range. Any in-
crease of flow causes a
substantial decrease in
column efficiency. Chro-
matography at a flow rate
required for usable col-
umn efficiency results in
unacceptably long analysis
time.
Compared to nitrogen, van Deemter curves for helium and hydrogen show great-
est efficiency at higher flow rates. Use of these gases at their optimum flow reduces
analyte elution time. The much shallower curves for these gases also demonstrate
that increasing the flow above optimum to further reduce analysis time results in
acceptable losses of efficiency. For these reasons, helium and hydrogen are com-
monly used as carrier gases for most capillary GLC applications.
Two different modes of operation, differentiated by carrier gas flow rate, are
possible with wide bore capillary columns. Maximum column efficiency is achieved
in “capillary column mode,” i.e., with carrier gas flows ≤6 mL/min. However, at
these low flow rates, chromatographic time is considerably longer than the time
to which pesticide analysts are accustomed with packed columns.
Figure 502-f
Van Deemter Curves
Effect of carrier gas flow rate on column efficiency for several
gases, measured using 30 m × 0.25 mm id column, 0.25 mm
film thickness.
1.2
1.0
0.8
0.6
0.4
0.2
10 20 30 40 50 60 70 80 90
Average Linear Velocity (cm/sec)
HETP or h (mm)
N
2
He
H
2
Pesticide Analytical Manual Vol. ISECTION 502
502–20
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Operation of wide bore columns in “packed column mode,” i.e., 10-25 mL/min
carrier gas flow, at 200° C (isothermal) combines the advantages of packed col-
umns’ faster elution with open tubular columns’ greater deactivation (fewer active
sites). In this mode, column efficiency equals or exceeds that of equivalent packed
columns. At the same time, polar pesticides that require polar liquid phases (e.g.,
DEGS) for packed column chromatography can be successfully chromatographed
on capillary columns, even with nonpolar liquid phases [13, 14]. Because analyte
relative retention times vary only with stationary phase and column temperature,
extensive data (Appendix I, PESTDATA) compiled over the years for packed col-
umns may be used for tentative identification of residues found with capillary
columns operated at the same temperature [4]. Directions for operation of a wide
bore capillary column in packed column mode, validated by interlaboratory study
[5], are presented below.
Carrier gas should be of the highest possible purity, because use of highest purity
gas will extend capillary column life. Moisture and oxygen traps should be used
for all gases.
Makeup Gases. Because modern GLC detectors are designed for optimum per-
formance at gas flows greater than those preferred for capillary columns, some
systems require additional “makeup gas” to be added before effluent enters the
detector. In addition to providing proper flow rate for optimum detector perfor-
mance, makeup gas efficiently sweeps analytes from the end of the column into
the detector. A gas different from the carrier gas may be used if the detector
requires a specific moderating gas, such as argon/methane or nitrogen for an
electron capture detector.
Makeup gas used for detector moderation should be of the purity level recom-
mended by the manufacturer. As with carrier gases, moisture and oxygen traps
should be used for all gases.
Installation and Conditioning of Capillary Columns
Regardless of the connection being made, proper cutting of the column and
retention gap is critical. Square, clean cuts minimize flow disturbances and allow
tubing of the same or different diameters to connect smoothly. Cleaving tools for
cutting polyimide-coated fused silica capillary tubing are available from most chro-
matography supply companies. After scribing the polyimide coating with a cleav-
ing tool, the tubing is cut by applying gentle pressure to bend the column oppo-
site where it was scribed. If properly scribed, the column will break cleanly at that
point. New cuts should be examined with a 10-20X magnifier to ensure that the
cut is square and clean with no ragged edges or chips in the polyimide coating.
Tubing should be recut if necessary to achieve a proper finish. All cuts should be
made after installing any ferrules (especially graphite) onto the tubing to elimi-
nate the possibility of small shaved ferrule particles being deposited in the end of
the tubing.
Two techniques can be used to facilitate marking critical measurements for proper
column positioning with either detectors or inlets. Water-soluble typewriter correc-
tion fluid can be used to mark columns at the desired length; the fluid does not
react with the polyimide coating and is easily removed. Alternatively, a small slice
of septum pushed onto the column can act as a marker and simultaneously hold
the nut and ferrule in place for easier maneuvering in the oven.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 502–21
SECTION 502Pesticide Analytical Manual Vol. I
After the inlet, column, retention gap, and detector have been connected, carrier
gas flow should be established and all connections thoroughly checked for leaks.
Every connection should be treated as the source of a potential leak, so connec-
tions should be minimized. Manufacturer’s instructions must always be followed
carefully to obtain leak-free connections. Soap solutions must never be used to
detect leaks, because they can be aspirated into the system; the resulting contami-
nation with nonvolatile materials can be removed only by rinsing the contami-
nated area. Electronic leak detectors are preferred. All leaks should be eliminated
prior to heating the column.
Capillary columns with cross-linked, bonded stationary phases do not require the
extensive conditioning of packed columns, because the stationary phases are more
stable than those in packed columns and less susceptible to bleed. Usually, purg-
ing the column thoroughly with carrier gas, then heating it to 20-30° C above the
maximum operating temperature for 1 hr is sufficient to condition the column.
Conditioning is performed with carrier gas flowing; the detector may be con-
nected during conditioning of capillary columns. It is critical that the upper tem-
perature limit for the column not be exceeded.
Rejuvenation of Capillary Columns
With time and use, nonvolatile residues accumulate in all capillary columns, re-
gardless of the use of retention gaps or other protective measures. Efforts to
improve deteriorated chromatography should always begin with removal of por-
tions of a contaminated retention gap or replacement of the retention gap. If the
analytical column is also contaminated and replacement of the retention gap is
insufficient to improve chromatography, a portion of the inlet end of the analyti-
cal column can be removed by cutting the column as previously described. If a
capillary column ≥5 m is used, removal of a relatively short segment does not
significantly affect its overall length or behavior, even if segments are removed
repeatedly.
When removal of a contaminated segment of column is insufficient to restore
appropriate chromatography, the column can be rinsed with solvents to help
remove accumulated residues; rinsing may be performed with the retention gap
attached. Column rinsing is possible because of the stability of the cross-linked,
bonded phase.
Kits for rinsing columns are commercially available. Most kits consist of a vial that
serves as a solvent reservoir; the vial has fittings for insertion of the column and
for connection to a gas supply that pressurizes the solvent. The detector end of the
column is inserted into the vial containing the rinse solvent, and gas pressure
forces solvent backward through the column. The column should be rinsed with
a sequence of solvents in order of decreasing polarity, starting with water and
ending with hexane. Each solvent should be miscible with the preceding one.
After the column is rinsed and excess liquid purged with gas flow, the column is
re-installed and purged with carrier gas at ambient temperature for 30-60 min.
The column should be conditioned by heating to 20° C above operating tempera-
ture for at least 15 min, but the upper temperature limit for the column must not
be exceeded.
Pesticide Analytical Manual Vol. ISECTION 502
502–22
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Recommended Operating Procedure for Wide Bore Columns (Isothermal)
The following procedure describes the setup and operation of a wide bore capil-
lary column in packed column mode. Both direct injection adapters (below) were
successfully validated [5, 14], and either may be used. Option 1 is recommended,
because the disposable liner can be easily replaced once it has become contami-
nated with injected co-extractives.
Apparatus and Reagents.
column, fused silica capillary column, bonded with one of the substituted
polysiloxane phases (Table 502-a), 30 m long × 0.53 mm id, 1.0 or 1.5 μm
film thickness
retention gap, deactivated fused silica tubing, 5 m long × 0.53 mm id
capillary column connectors, low dead volume or zero dead volume, suitable
for connecting analytical column to retention gap
direct injection adapter. Use new adapter as is without silanizing. Resilanize
used and cleaned adapters with 10% dimethyldichlorosilane/toluene; after
resilanization, heat to 240° C overnight with gas flow before use. Optional
adapters are:
1) Silanized direct injection adapter (Figure 502-e, Adapter 3), 1/4" od,
4 mm id (Cat. No. 210-1071, J&W Scientific, Folsom, CA). This adapter
is specially made to FDA specifications and is available with the special
order part number. Direct injection adapter is 220 mm total length. Inlet
end, 125-130 mm long measured from top of restrictor, has notches.
Column oven end descends 75 mm below flared end of tapered restrictor.
Column oven end may be shortened if desired; if cut, leave at least 20
mm tubing to attach adapter/column reducing union and lightly fire
polish cut end. With Option 1 only: column inlet liner, PureCol? for
4 mm id columns (Cat. No. 2-0540M, Supelco, Inc., Bellefonte, PA, or
equivalent)
2) Double restrictor adapter, 1/4” od (Figure 502-e, Adapter 2). Adapter
has two restrictors. Top restrictor allows passage of syringe needle into
chamber formed by two restrictors. Lower restrictor is tapered for con-
nection of column end into adapter. Silanize adapter prior to use if not
silanized by manufacturer or if adapter has been cleaned. These adapters
may be purchased cut to specified length or as longer version to be
cut by user.
pesticide grade silanized glass wool; see Section 204 for silanization
in laboratory; must be free of nitrogen, chlorine, phosphorus, or sulfur con-
taminants
capillary installation kit, required if chromatograph was designed for packed
column. Kit should include necessary fittings to attach retention gap to inlet
adapter and, if necessary, to detector, with provision for makeup gas at de-
tector connection.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 502–23
SECTION 502Pesticide Analytical Manual Vol. I
Reducing
union:
adapter to
retention gap
75
mm
Retention
gap
125
mm
PureCol
TM
liner
GC inlet
Septum
Inlet nut
& ferrule
Direct
injection
adapter
Glass
wool
Ring formed
when column
seals with
restrictor
Tapered restrictor
within adapter
Gow-Mac Leak Detector, available from chromatography suppliers
Magnifier, 20X
typewriter correction fluid, water based (if needed to mark correct positions
on capillary tubing)
capillary column cleaving tool
Instrument Setup. Connect apparatus according to following directions; review
instrument manufacturer’s instructions and adjust directions as necessary to ac-
commodate specific equipment. Figure 502-g shows the Option 1 inlet system
arrangement:
? Place ferrules and nuts on end(s) of retention gap and column.
? Cut ends of column and retention gap with capillary cleaving tool. It is
imperative that ends be cut after placement of ferrules to eliminate pos-
sibility of ferrule fragments becoming settled in tubing. Examine new
cuts with 20X magnifier to assure that ends are square and smooth.
Recut as necessary to obtain smooth, square ends.
Figure 502-g
Capillary Column Inlet System
Pesticide Analytical Manual Vol. ISECTION 502
502–24
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
? Clean ends of column and retention gap with wiper wetted with metha-
nol.
? Attach retention gap to column, using appropriate capillary column con-
nector.
? Secure retention gap to column cage.
? If using injection adapter (1), place small plug of glass wool in inlet end
and push down as far as restrictor, then place disposable PureCol? liner
in inlet end of adapter. If using injection adapter (2), use as is without
additional equipment.
? Gently insert adapter into instrument inlet or fitting until it touches top
of injector. Move it back about 1-2 mm and tighten nut. Exercise care
not to fracture end of adapter or liner when tightening nuts and ferrules
during installation; overtightening can cause leaks and deform expen-
sive hardware. If considerable effort is necessary to tighten fitting, it
may already be deformed and should be replaced. (Follow manufac-
turer’s instructions on how to tighten ferrules, check for leaks with
leak detector, then tighten in small increments, e.g., 1/4 turns, using
correct size wrench to obtain leak-free connection.)
? Attach reducing union (for connecting adapter to retention gap) to col-
umn oven end of adapter.
? Insert end of retention gap through reducing union and into adapter.
Push retention gap firmly into flared portion of restrictor until seal is
formed between polyimide coating of retention gap and adapter restrictor
wall. Formation of seal is evidenced by visible “ring” at contact point
between tubing and restrictor wall (Figure 502-g, enlarged area). Tighten
column nut on reducing union.
? If necessary, install makeup gas fitting to detector inlet using appropriate
length and diameter of silanized narrow bore (1 mm id) glass tubing.
Install column into detector as directed by detector manufacturer’s in-
structions (if available) for positioning column. See section on connec-
tions, above, for additional cautions about effects of column positioning.
? Use helium carrier gas to get best column efficiency and compatibility
with various detectors. For 30 m × 0.53 mm id columns, set initial flow to
20 mL/min.
? Use makeup gas as needed to accommodate optimum detector operation.
Nitrogen, helium, argon/methane, or other gases may be used, as re-
quired for proper detector operation. Adjust flow of makeup gas so that
total flow of carrier and makeup gases equals optimum gas flow recom-
mended by detector manufacturer. Makeup gas flows of 5-40 mL/min are
typical.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 502–25
SECTION 502Pesticide Analytical Manual Vol. I
System Startup and Inspection.
? After installation and establishment of initial carrier and makeup gas
flows, check all connections and fittings for leaks with electronic leak
detector. Do not heat system until it is leak-free and has been thoroughly
purged with carrier gas for 15-20 min to avoid damage from oxygen.
? Heat column to 230° C for 1 hr or until detector baseline has stabilized.
Reduce temperature to 200° C and recheck both carrier and makeup
gas flow rates. Adjust carrier gas flow and column temperature as nec-
essary to meet retention time and rrt
c
requirements for specific system
being used. Re-adjust makeup flow as necessary to maintain optimum
detector flow.
? Evaluate system for linearity, repeatability, and tolerance for varying
injection volumes. System should be linear over at least one full scale
deflection on integrator/recorder. For accurate quantitation, responses
to repetitive injections of standard reference material should have per-
cent relative standard deviation ≤5%. Limitations on injection volumes
must be determined for each system. Experience has shown that injec-
tion volumes of 1.0-6.0 μL are normally tolerated without adverse affects
on analyte response.
References
[1] Jennings, W. (1987) Analytical Gas Chromatography, Academic Press, Orlando,
FL
[2] Poole, C., and Poole, S. (1991) Chromatography Today, Elsevier, New York
[3] Bostwick, D.C., and Giuffrida, L. (1968) J. Assoc. Off. Anal. Chem. 51, 34-38
[4] Fehringer, N.V., and Walters, S.M. (1984) J. Assoc. Off. Anal. Chem. 67, 91-95
[5] Parfitt, C.H. (1994) J. AOAC Int. 77 92-101
[6] Grob, K., Jr., and Muller, R. (1982) J. Chromatogr. 244, 1185-1196
[7] Mehran, M.F. (1986) J. High Resolut. Chromatogr. Chromatogr. Commun. 9, 272-
277
[8] Seferovic, W., et al. (1986) J. Chromatogr. Sci. 24, 374-382
[9] Hinshaw, J.V., Jr. (1987) J. Chromatogr. Sci. 25, 49-55
[10] Grob, K., Jr. (1982) J. Chromatogr. 237, 15-23
[11] Pennington, L.J. (March 1986) “GLC Chromatographic Behavior of Organo-
phosphate Pesticides on Three Megabore Capillary Columns,” LIB 3017, FDA,
Rockville, MD
[12] Jennings, W., and Mehran, M.F. (1986) J. Chromatogr. Sci. 24, 34-40
Pesticide Analytical Manual Vol. ISECTION 502
502–26
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
[13] Saxton, W.L. (Feb. 1988) “Evaluation of Wide Bore Capillary Columns for
Luke Extracts Which Have Not Undergone a Column Cleanup,” LIB 3182,
FDA, Rockville, MD
[14] Gilvydis, D.M., and Walters, S.M. (1991) J. Assoc. Off. Anal. Chem. 74, 830-835
[15] Hopper, M.L. (1987) J. High Resolut. Chromatogr. Chromatogr. Commun. 10,
620-622
Pesticide Analytical Manual Vol. I SECTION 503
503–1
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
503: DETECTORS
503 A: INTRODUCTION
GLC detectors are devices that indicate the presence of eluted components in the
carrier gas emerging from the column. Depending on the way in which they
measure the quantity of the analytes, detectors are classified as differential concen-
tration, differential mass, or integral [1]. The electronic equipment associated
with the detector amplifies the signal and causes the response to be recorded.
Definitions of Detector Characteristics
Characteristics of detector operation that are critical to qualitative and quantita-
tive determination of residues include sensitivity, selectivity, and linearity. Certain
terminology is common to the discussions of these characteristics in different
detectors.
Sensitivity. Detector sensitivity refers to the relationship between amount of analyte
injected and response of the detector. Detector response is the change in mea-
sured detector signal that results from a change in amount of analyte present
within the detector volume; measured detector signal includes the amplification
provided by associated electronics. Sensitivity is often described by referring to the
smallest amount of a specific analyte that causes a measurable detector signal.
FDA methodology has traditionally specified detector sensitivity in terms of ng of
a specified compound that causes 50% full scale deflection (FSD) on a recording
or data system. That convention is continued in this chapter and in the determi-
native step descriptions in Chapter 3 methods.
Selectivity. Detectors must be selective to be suitable for use in determining any
trace residue, including pesticides. Selectivity refers to the detector’s preferential
response to one or more elements (“heteroatoms”) or functional groups that
might be present in analytes of interest. Response of the detector to these moieties
must far exceed its response to carbon, hydrogen, and oxygen if the resulting
chromatogram is to distinguish between residues and food co-extractives present
in the same extract.
Nonselective detectors, such as flame ionization (FID) and thermal conductivity
(TC), respond to solutes in proportion to the mass of each that elutes from the
column. Such detectors are impractical for most residue determinations.
Among detectors that are suitable for residue determination, selectivity to the
moiety of interest varies considerably. Probably no detector is completely “specific”
to one heteroatom or functional group; instead, degrees of selectivity can be
described in terms of the relative response of the detector to the same weight of
different compounds or moieties.
In practical terms, the greater the detector selectivity, the less sample cleanup is
needed (within the boundaries discussed in Section 501 C) and the greater the
inherent degree of confirmation that is provided by the determination. Conversely,
the less selective the detector to the type of analyte being detected, the greater the
precaution needed in preparing samples, avoiding reagent contamination, and
confirming residue identity.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–2
SECTION 503 Pesticide Analytical Manual Vol. I
Some relatively selective detectors are subject to interferences from co-extractives
that contain the heteroatom to which they respond, in particular those detectors
that respond to the presence of nitrogen or sulfur. Sulfur-selective detectors are
subject to considerable interference from organic disulfides present in foods such
as onions, rutabagas, and brassicas. Use of cleanup steps to remove or react
the interfering co-extractives may also cause the analyte(s) to be lost. Nitrogen-
containing compounds are extractable from all foods; for this reason, no detector
that selectively responds to elemental nitrogen can ever be totally successful
in analysis of foods for trace-level contaminants. Several HPLC determinative steps
have been developed for particular nitrogen-containing functional groups
(Sections 401 and 403), and these have been more successful because interference
from other forms of nitrogen is avoided.
Linearity. Use of a detector within its linear range is a prerequisite for the sim-
plified way in which residues are routinely quantitated in pesticide residue deter-
minations (Section 504). Terms associated with detector linearity are: dynamic
(response) range, over which a change in the amount of chemical present within
the detector volume produces a measurable change in detector response, and
linear (response) range, the portion of the dynamic range over which a change
in the amount of a chemical present within the detector volume produces a
proportional change in detector response. A detector’s linear range is the range of
analyte weight over which the sensitivity of the detector is constant to ±5.0%, as
determined from a linearity plot of response/weight vs log weight [2].
FDA laboratories evaluate the dynamic range of a detector and then operate in a
segment of that range that exhibits appropriate linearity. For added assurance that
quantitation is accurate, sufficient extract and reference standard solutions are
injected to cause detector responses to residue and standard to agree within 25%.
503 B: ELECTRON CAPTURE DETECTOR
The electron capture (EC) detector has been used for many years to analyze
foods for organohalogen pesticide residues. The earliest EC detector in com-
mon use had a
3
H (tritium) radioactive source; this was later replaced by detectors
using a
63
Ni source. The continued popularity of the EC detector results from
its high sensitivity to halogen and certain other moieties, as well as its ruggedness
and low maintenance needs. Its sensitivity makes it applicable to determination of
residues at the ppb and even ppt level, its wide dynamic response range facilitates
its use with automatic data systems, and its high operating temperature (≤400° C)
minimizes detector contamination by sample co-extractives and column bleed.
These advantages are offset by the EC detector’s low selectivity compared to other
detectors used in residue determination. When using the EC detector, appropriate
methodology precautions are necessary to prevent introduction of interferences
from food samples, reagents, or the environment (Section 501 C).
EC detectors used in FDA laboratories have
63
Ni sources and constant current,
variable frequency design. Several manufacturers produce and market such detec-
tors, all of which operate on the principles described below but vary in source
activity, cell volume, and geometry. Most EC detectors currently in FDA labo-
ratories are from Hewlett-Packard (HP), Wilmington, DE; Tremetrics, Inc.
(formerly Tracor, Inc.), Austin, TX; or Varian Associates, Sunnyvale, CA. Discus-
sions of detector characteristics in this section refer to detectors from these manu-
facturers.
Pesticide Analytical Manual Vol. I SECTION 503
503–3
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Principles
High energy beta particles, emitted by the
63
Ni source, collide with carrier gas
molecules to produce low energy electrons. These electrons are continually
collected at the cell anode by applying voltage pulses to the cell electrodes. Cell
current thus produced is measured and the pulse interval (frequency) adjusted to
maintain constant cell current. A standing pulse frequency describes the equilib-
rium condition that exists when only carrier gas is passing through the cell.
When molecules of an electrophilic substance enter the detector, electrons are
“captured” to a degree dependent on the amount and electron affinity of the
substance. As the electron supply is thus decreased, pulse frequency increases to
generate the exact number of electrons necessary to maintain the established
constant current. Change in frequency required to maintain constant cell current
is converted to voltage, and this signal is sent to the recording device as the
detector’s response to the analyte.
Design
Two basic differences exist in EC cell
design, one a pin-cup with
63
Ni plated
on the cell wall and an anode sus-
pended in the center of the cell cavity
from the top, and the other with
63
Ni
plated onto a cylinder aligned parallel
to the column and cell flow. The elec-
trode leads enter the cell cavity at right
angles to the source and gas flow.
Figure 503-a, diagrams A and B, dis-
play these respective designs.
Apparatus and Reagents
Section 501 B provides general infor-
mation on apparatus and reagents
required for GLC. Further materials
or specifications for this detector are
described below.
Radioactive Source: Special Require-
ments. The presence of radioactive
material in EC detectors brings them
under the authority of the Nuclear
Regulatory Commission (NRC). The following special procedures must be fol-
lowed by a laboratory with an EC detector:
1) Labeling. According to NRC regulations, each chromatograph contain-
ing a
63
Ni detector must have a label signifying the isotope, activity, and
date at which the activity was determined and the words “Caution: Ra-
dioactive Material.” The NRC will accept the manufacturer’s label on
the detector if it contains the necessary information. If this information
is not present, stick-on labels must be applied to the gas chromatograph.
Appropriate labels have been provided to FDA laboratories by FDA’s
Column
A
B
Makeup gas
Anode
63
Ni foil
1.0 cm
Figure 503-a
Two EC Detector Designs
A, pin-cup EC,
63
Ni plated on cell wall; B,
63
Ni plated
onto cylinder.
[Reprinted with permission of John Wiley & Sons, Inc., from
Detectors for Capillary Chromatography (Copyright ?1992)
Hill, H.H., and McMinn, D.G., ed., Chapter 5, by Grimsrud,
E.P., Figure 5.2, p. 86.]
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–4
SECTION 503 Pesticide Analytical Manual Vol. I
Winchester Engineering and Analytical Center (WEAC), Winchester, MA.
In addition, the Department of Health and Human Services (DHHS)
requires that a sign with the following notice be posted on each gas chro-
matograph: “This equipment contains a radioactive source registered with
the RSO (i.e., the WEAC Radiation Safety Officer) as required by license
from the NRC. Notify the RSO before removing the source from this
location or upon any change in custodial responsibility” [3].
2) Venting. In addition to the labeling requirement, DHHS also requires
that chromatographs with
63
Ni or other radioactive sources be vented
through plastic tubing into a chemical hood or room exhaust [3].
3) Wipe tests. Any laboratory with a
63
Ni EC detector is required, as part of
the licensing procedure, to perform a wipe test of all accessible exterior
parts of the detector each January and July. WEAC supplies FDA labora-
tories with cotton-tipped swabs for performing the wipe tests. Each detec-
tor is wiped with an alcohol-moistened swab using moderate pressure,
with particular emphasis on potential leak areas such as the outlet termi-
nus and joints. Each swab is returned to WEAC in a mailing tube so that
radiation removed from the detector exterior can be measured. A Certifi-
cate of Inspection for each detector is provided by WEAC and returned
with new swabs.
4) Cleaning. NRC licenses in effect in FDA laboratories permit use of
EC detectors containing
63
Ni but do not allow their dismantling and clean-
ing. All FDA
63
Ni EC detectors are shipped to the WEAC facility for clean-
ing [4].
Laboratories outside FDA must either make arrangements with a properly licensed
laboratory for detector cleaning services or obtain the appropriate NRC license.
Laboratories that have an existing license for use of
63
Ni EC detectors
may be able to obtain from NRC an amendment that permits cleaning. Proof that
the laboratory is capable of handling such materials safely is required before NRC
will grant such an amendment.
Carrier and Makeup Gas. EC detector manufacturers recommend the use of
argon/methane (95+5 or 90+10) for greatest detector linear range; however, ni-
trogen is used satisfactorily in some FDA laboratories. An external switch on the
chromatograph permits selection of pulse width and cell current to accommodate
whichever gas is predominant upon reaching the detector. Only reliable, high
purity grade gas should be used, with oxygen and moisture traps on all gases going
to the detector.
Often, the carrier gas and flow rate chosen for optimum column efficiency do not
result in the best detector operation. Most EC detectors are configured to allow
makeup gas to be added to the flow from the column so that detector operation
is enhanced. In the most common example, argon/methane is essential to opera-
tion and is added to helium column carrier gas to produce optimum EC detector
response and stability.
Detector Characteristics
The following basic characteristics of EC detectors must be understood for proper
application to pesticide residue determination:
Pesticide Analytical Manual Vol. I SECTION 503
503–5
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
-1 0123456
Log Chlorpyrifos (pg) Injected
0
100
200
300
400
500
600
Mean Response Factor
Sensitivity. Magnitude of response of a
63
Ni EC detector to a particular compound
is dependent on the electron-capturing ability of the compound. Under condi-
tions described below, the minimum detectable amount of an electron-capturing
compound to which the detector will respond is typically in the 1-10 pg range.
Reproducibility of response at this low
level is not as good as that for larger
amounts of the same compound. A se-
ries of injections of 1-300,000 pg chlor-
pyrifos was made monthly over a 6-mon
period [5]; variation in response factor
over that time is shown in Figure 503-b.
Standard deviations of response factors
for each weight indicate that response
stability was considerably better at levels
≥3 pg than at the 1 pg level, although
response to 1 pg was reproducible over
the course of any one day. Although this
experiment was performed with an HP
5713A instrument, minimum response is
expected to be equivalent with other
63
Ni
constant current detectors.
Selectivity. EC detectors respond to
molecules containing an electrophoric
group, i.e., a highly polar moiety that
provides an electron-deficient center in
the molecule; examples include halogen, sulfur, phosphorus, and nitro- and α-
dicarbonyl groups [2]. Because the response is not to a single element nor is it
proportional to the amount of an element in a molecule, statements on detector
selectivity can refer only to ranges or to relative responses to particular analytes.
Relative to its response to hydrocarbons, an EC detector may display 100-1000 fold
greater response to mono- and disubstituted halogens and up to 10
6
times greater
to polysubstituted halogens [6]. However, other molecules that contain only
carbon, hydrogen, and oxygen may also be electrophoric, and EC detector
response is far less selective to halogen relative to these molecules; examples
include quinones, cyclooctatetracene, 3,17-diketosteroids, o-phthalates, and conju-
gated diketones [2].
The relative lack of selectivity of the EC detector provides a bonus of applicability
to a variety of analytes; e.g., if an extract contains residues of pesticides containing
halogen and also other residues containing sulfur, use of EC-GLC permits simul-
taneous determination of each. Lack of selectivity is more often a detriment to
residue analysis, however; in practice, the EC detector’s value is dependent on
how free of interfering co-extractives the final extract is. Food co-extractives
or environmental contaminants with electrophoric characteristics compromise
the determination by causing responses that interfere with residues or that are
mistakenly interpreted as residues.
Many examples of the interfering substances have been documented during long
use of EC detectors. In addition to examples noted in Section 501 C, artifacts from
plastics, rubber products, hand lotions, and cleaning solutions have been seen.
Figure 503-b
Reproducibility
Mean and standard deviation of different
weights of chlorpyrifos injected monthly
(6 mon) into GLC with
63
Ni EC detector.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–6
SECTION 503 Pesticide Analytical Manual Vol. I
-10123456
Log pg Injected
100
200
300
400
500
600
Response Factor
ronnel
chlorpyrifos
p,p′-DDT
0123456
Log Chlorpyrifos (pg) Injected
0
100
200
300
400
500
Response Factor
Figure 503-c
Dynamic Response Ranges
Response range of
63
Ni EC detector to three
halogenated pesticides.
Figure 503-d
Linear Response Range
Response range of
63
Ni EC detector is linear
(±10%) to amounts of chlorpyrifos injected
over a limited portion of the dynamic response
range.
Certain fruits and vegetables also contain nonhalogenated substances that cause
EC response, and these are not always removed by Florisil column cleanup (Sec-
tion 303 C1, C2). Some of the sample types known to contain artifacts that cause
an EC response include cabbage, radishes, and lettuce (6% Florisil eluate), carrots
(15% Florisil eluate), and onions (both 6 and 15% eluates). Recommendations for
avoiding interferences are included in the recommended operating procedures
for EC, below.
Linearity. The mode of operation of
the
63
Ni constant current detector pro-
duces a dynamic response range of
greater than five orders of magnitude.
Experiments in which an automatic data
system was interfaced with the detector
[5] showed that detector response to
increasing quantities of injected mate-
rial was still increasing when the data
system became saturated. In order to
plot the dynamic response over such a
large range, the response (units in which
response is measured/pg injected) is
plotted vs log pg injected. A linear dy-
namic response range would produce a
straight line parallel to the x-axis in such
a plot. Plots of typical dynamic response
ranges of the HP
63
Ni detector to three
pesticides are shown in Figure 503-c.
Instead of straight horizontal lines, the
plots indicate a variation in response factor with amount of pesticide injected.
These plots show that the detector is not linear over its entire dynamic response
range. Within smaller segments of this range, however, the detector displays ac-
ceptable linear response.
The range over which response is con-
sidered linear is dependent on the defi-
nition of linearity chosen. For example,
in Figure 503-d the linear range is con-
stant within ±10%; detector response is
linear from approximately 30-500 pg
chlorpyrifos. With this same definition,
detector response can also be considered
linear over other segments of the dynamic
range. A change in definition (i.e., dif-
ferent % variation permitted) would
change the ranges for which detector re-
sponse is considered linear.
The general rule that detector response
to residue and reference standard should
match within 25% is especially important
when using the EC detector.
Pesticide Analytical Manual Vol. I SECTION 503
503–7
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
0 25 50 75 100 120 150
Flow Rate (mL/min)
0
5
10
15
20
25
Peak Area
12345
Log Chlorpyrifos (pg) Injected
100
150
200
250
300
350
Response Factor
200°
250°
300°
350°
Figure 503-e
Effect of Detector Temperature
Dynamic response range of
63
Ni EC
detector at four different temperatures.
Figure 503-f
Effect of Carrier Gas Flow Rate
Response range of
63
Ni EC detector to 100
pg chlorpyrifos at different carrier gas flow
rates.
Other Influences on Detector Performance
The following parameters were studied for HP
63
Ni constant current detectors [5]
and, in a more limited study, for a Tracor
63
Ni constant current detector [7].
Detector Temperature. Experiments with
the HP
63
Ni constant current detector [5]
documented its dynamic response range for
seven pesticides at four detector tempera-
tures; Figure 503-e displays results for
chlorpyrifos. Detector temperature caused
only slight changes in response to any par-
ticular amount of pesticide and caused no
consistent change over the whole dynamic
range. Thus, there is no reason to choose
detector temperature on the basis of en-
hanced response.
Manufacturers provide recommendations
for operating temperatures of detectors.
Varian recommends operating the detec-
tor at 30° above column temperature, HP
recommends 250-300° C, and Tracor rec-
ommends operation at 350° C. Detector
contamination by materials eluting from the
GLC column can be minimized with use of
higher detector temperatures, but the
63
Ni radioactive foil must not be operated
at >400° C.
Flow Rates. Column flow rate is usually chosen to optimize column efficiency and
permit reasonable analysis time. The effect of flow rate on detector operation
must also be considered, however, because response of the concentration-sensitive
EC detector decreases with increased flow rate.
Plots in Figure 503-f illustrate the effect of
flow rate on response for the HP
detector; a similar plot was obtained with
the Tracor detector. In each case, the
optimum flow rate range is that which
provides for response stability, i.e., flow at
which a small change in rate does not
cause a large change in response.
HP and Varian recommend a minimum
flow rate of 30 mL/min through their
detectors, and Tremetrics recommends
60-100 mL/min [6, 8, 9]. Each recom-
mends use of makeup gas to bring the
total flow through the detector to 60-100
mL/min. Makeup gas is usually needed
for capillary columns (Section 502 C) and
may be either the same as the carrier gas
or different. For example, when hydrogen
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–8
SECTION 503 Pesticide Analytical Manual Vol. I
or helium carrier gas is used to obtain maximum resolution and efficiency on the
column, makeup gas of argon/methane or nitrogen must be used for proper
detector operation.
Recommended Operating Procedures
The following steps should be taken for a new detector and after each time a clean
detector is installed. (See above for special requirements for radioactive sources.)
?
63
Ni constant current detectors are usually delivered already installed in
the chromatograph. To re-install after cleaning, follow manufacturer’s
directions for setting up the instrument and conditioning the column [6,
8, 9]. Ensure that all heaters, temperature sensors, and electrical connec-
tors are properly positioned. Never connect a column to a cool detector.
? Heat detector until it reaches the maximum operating temperature
recommended by the manufacturer, then attach column and equilibrate
overnight at operating temperature and flow rate.
? Determine the instrument attenuation required to cause 40-80% FSD in
response to 1.5 ng chlorpyrifos.
? Determine detector dynamic response range to chlorpyrifos and other
standards of interest by plotting response factor (response/unit weight) vs
weight injected on a semilogarithmic scale. Do not operate instrument
outside the dynamic response range.
Earlier Tracor models allow for determining the pulse frequency profile and
saturation current. To operate these models, refer to the instructions in the
manufacturer’s operation manual. Newer models have these parameters preset, so
adjustments are not necessary or possible.
To minimize interferences that can occur during determination with EC detectors,
follow these rules:
? Exercise extreme caution to avoid introduction of contaminants from
reagents, apparatus, and environmental sources; routine inclusion of
reagent blanks in laboratory quality assurance procedures will monitor
success of these precautions.
? Employ suitable cleanup procedures for extracts that will be examined by
EC detectors. Elution through Florisil is usually required before EC deter-
mination, though even this is not a guarantee that artifacts from foods
will not cause response.
? Always confirm the identity of residues that have been tentatively identi-
fied by EC GLC; confirmation may include GLC with element-selective
detectors or other techniques (Section 103).
Pesticide Analytical Manual Vol. I SECTION 503
503–9
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
503 C: FLAME PHOTOMETRIC DETECTOR
The flame photometric detector (FPD) is the detector of choice for determination
of organophosphorus residues and the only practical detector for organosulfur
compounds. In its phosphorus-selective mode (FPD-P), the detector is one of the
most element-selective GLC detectors available, although large amounts of sulfur
will cause a response. The sulfur-selective mode (FPD-S) offers less selectivity (phos-
phorus can interfere) and is linear only on a semilogarithmic scale, but it provides
useful confirmation for sulfur compounds tentatively identified by EC determina-
tion. Neither nitrogen nor chlorine cause any practical interference in either
mode. In addition, the ratio of FPD-P and FPD-S responses can be used to calcu-
late an analyte’s P:S ratio for confirmatory purposes.
Methods designed for use with FPD determination sometimes include only mini-
mal cleanup. However, column contamination can be caused by repeated injec-
tions of extracts from such methods, and the cautions outlined in Section 501 C
must be observed.
The majority of FPD detectors in use in FDA laboratories were produced by three
manufacturers: HP, Tremetrics (formerly Tracor), and Varian Associates. Discus-
sion of detector characteristics in this section is limited to these models.
Principles
GLC column effluent is burned in a flame fed by a mixture of hydrogen and air.
Characteristic optical emissions are produced when compounds containing phos-
phorus or sulfur are decomposed in the flame, and these emissions are viewed by
a conventional photomultiplier tube through a narrow bandpass (interference)
filter of appropriate wavelength. Choice of filter determines whether emissions
produced by phosphorus or sulfur reach the photomultiplier tube. A filter with
maximum transmittance at 526 nm, corresponding to the emission wavelength of
HPO, permits detection of phosphorus compounds, while one with maximum
transmittance at 394 nm, the emission wavelength of S
2
, detects sulfur compounds.
A single optical filter and photomultiplier tube may be used, or two filters and
photomultiplier tubes can be assembled to permit response to both phosphorus
and sulfur simultaneously.
Design
Figures 503-g and 503-h are
diagrams of single and dual
flame FPDs, respectively. In
each, column effluent enters
the detector from the bottom,
where it is mixed with hydro-
gen gas. Air is added before
the effluent and hydrogen
emerge from the jet or at the
same time they emerge. The
emission from the resulting
flame is measured by a photo-
multiplier tube after passing
through the proper filter.
Figure 503-g
Single-Flame FPD
[Reprinted with permission of John Wiley & Sons, Inc., from Detectors
for Capillary Chromatography (Copyright ?1992) Hill, H.H., and McMinn,
D.G., ed., Chapter 9, by Hutte, R.S., and Ray, J.D., Figure 9.1, p. 196.]
Photomultiplier tube
Interference filter
From GC
Hydrogen
inlet
Air inlet
Flame
Emission region
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–10
SECTION 503 Pesticide Analytical Manual Vol. I
Photomultiplier tube
Interference
filter
Flames
Emission
region
Vent
H
2
inlet
Air #1
inlet
Air #2
inlet
From
GC
The Varian design differs
from the others in that two
air-hydrogen jets are used,
“to separate the region of
sample decomposition from
the region of light emission”
[10]. Sample decomposition
occurs in the presence of air
and hydrogen in the first
combustion. Above that area,
additional air is added to the
combusted mixture; this
supports a second flame,
whose optical emission is
filtered and measured as
detector response.
Apparatus and Reagents
Section 501 B provides gen-
eral information on appara-
tus and reagents required for GLC. Nitrogen or helium, of at least 99.998% purity,
are usually used as carrier gases for columns connected to FPDs. Hydrogen and
air are used as reactant gases in the detector flame. Manufacturers’ recommenda-
tions for hydrogen purity vary from 99.995-99.999%. Air purity should be zero
grade (maximum total hydrocarbon <2 ppm) or CGA Grade E [10-12].
Detector Characteristics
Sensitivity. The minimum amount of phosphorus detectable by the FPD-P is
about 0.01 ng; for the FPD-S, about 0.04 ng sulfur. Detector sensitivity is greatly
dependent on the condition of the photomultiplier tubes. Response varies among
tubes, and the use of a variable voltage output with the power supply (or a variable
voltage power supply) makes precise attainment of specific sensitivities easier to
accomplish. Light leaking into a photomultiplier tube will increase the noise level
and decrease the detector’s effective sensitivity by making it less able to detect
small amounts of analyte.
Response of the FPD-S (394 nm filter) to sulfur is proportional to the square of
the concentration of sulfur. When the 526 nm filter is used (FPD-P mode), re-
sponse to sulfur is also proportional to the square of the concentration. This
relationship affects both selectivity ratios and linearity of the detector in both the
P and S modes.
Selectivity. The response of the FPD-P detector is about 10
5
times greater to
phosphorus than to carbon. The selectivity of phosphorus to sulfur in the FPD-P
mode varies with the amount of sulfur present because of the square root
relationship of response to sulfur concentration. For the Varian detector, P:S
selectivity varies from >10
4
for very low concentrations of sulfur to about 50 for
very high concentrations. Preliminary experience with the Varian detector indicates
that it has a greater P:S selectivity than the other models. It is assumed that this
increased selectivity is due to the stacked flame arrangement of this detector.
Figure 503-h
Dual-Flame FPD
[Reprinted with permission of John Wiley & Sons, Inc., from Detectors
for Capillary Chromatography (Copyright ?1992) Hill, H.H., and McMinn,
D.G., ed., Chapter 9, by Hutte, R.S., and Ray, J.D., Figure 9.2, p. 197.]
Pesticide Analytical Manual Vol. I SECTION 503
503–11
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Selectivity of the FPD-S also varies with concentration of sulfur because of the
square root relationship. The sulfur-to-carbon ratio varies from >10
6
at high sulfur
concentration to 10
4
at low concentration. The S:P selectivity varies from >10
4
at
high sulfur concentrations to 10 at low concentrations [10].
Selectivity of the FPD-S varies with the ability of the individual filter to screen out
wavelengths associated with phosphorus. Experimentally, FPD-S detector response
to a mixture of pesticides was compared using several different 394 nm filters.
When conditions for each system were set to cause equal response to propargite
(sulfur only), all filters permitted FPD-S response to methamidophos, chlorpyrifos,
acephate, and omethoate (both phosphorus and sulfur), and most permitted
response to monocrotophos (phosphorus only). Only one filter of those tested did
not permit FPD-S response to monocrotophos; i.e., the particular filter was far
more selective to sulfur than the others. Further examination of the spectrum of
light passed by the different 394 nm filters showed a distinct difference in the
amount of absorbance at 526 nm; as expected, the filters that permitted FPD-S
detection of monocrotophos passed much more 526 nm light than did the filter
that did not detect it [13].
Linearity. Response of the FPD-P (526 nm filter) to phosphorus is linear over
about four orders of magnitude.
Because of the square root relationship, response of the FPD-S to sulfur can be
plotted as a straight line only if semilog paper is used. Most newer instruments
provide a switch that automatically converts the detector output signal to its square
root for an apparently linear response. However, quantitation using this converted
signal is accurate only if the signal is carefully “zeroed,” and the detector response
is less sensitive at this setting. Many laboratories choose to measure the uncon-
verted signal and plot response vs weight injected on semilog paper. Quantitation
using the FPD-S is always less reliable than with other detectors.
Other Influences on Detector Performance
Detector Temperature. Each of the three manufacturers recommends a mini-
mum detector operating temperature of 120° C. Recommended maxima range
from 250° C (Tremetrics) to 350° C (Varian), with HP intermediate at 300° C.
Physical deterioration of parts of the detector can occur or be accelerated at
higher temperatures. Both O rings and the casing for the photomultiplier tube
have been seen to deteriorate at high temperatures. During routine operation, O
rings should be changed periodically (about every 6-12 mon); all O rings in a
particular detector should be changed at the same time. All manufacturers warn
against continued operation at maximum temperature. Normal detector operat-
ing temperature should be about 20° C above that of the column, usually ≤250°
C.
Gas Flow Rate. Optimum gas flow rate varies among detectors, and directions
provided by the manufacturer of the specific detector should be followed. Addi-
tional experimentation may be required to optimize flow rates for any particular
detector.
Detector Voltage. Satisfactory operation of the FPD requires use of a highly
stabilized voltage power supply. Depending on the manufacturer, voltage may
range from 350-850 V and may be obtained from either a variable or set voltage
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–12
SECTION 503 Pesticide Analytical Manual Vol. I
supply source. Detectors are usually shipped with the recommended voltage preset
at the factory. Manufacturers’ operations manuals give specific instructions for
varying voltage when this is an option.
Sensitivity of the detector can be increased by increasing the voltage, but an upper
limit is imposed by the simultaneously increasing noise. Optimum voltage can be
determined by comparing detector response to an amount of compound as the
voltage is varied.
Adequate Chromatography. Nonlinear response of FPD-P to oxygen analogs
of organophosphorus pesticides (P=O compounds) is often noted. Because the
detector (526 nm filter) is linear over a wide range for P=S compounds (i.e., most
parent organophosphorus pesticides), the difficulty is assumed to be caused by
degradation of P=O compounds. Once attributed to a defect in detector design,
this problem is now considered to be caused by poor chromatography of these
polar compounds, and thus a column problem. Use of wide bore capillary col-
umns (Section 502 C) minimizes the effect.
Recommended Operating Procedures
FPD-P. The following steps should be taken for detector operation:
? Install detector if necessary, according to instructions provided in
manufacturer’s manual [10-12]. FPD usually comes installed in chromato-
graph.
? Set detector temperature as recommended by manufacturer, at least
20° C above column temperature.
? Establish flow rate of column carrier gas as suitable for proper column
operation (Section 502). Set flows of hydrogen and air as recommended
by detector manufacturer or as determined from experimentation to pro-
vide optimum operation. Connect column to detector.
? If voltage is set by user, follow manufacturer’s directions.
? Ignite flame after all instrument temperatures are equilibrated and with
carrier gas flowing into detector.
? Turn on air.
? Depress ignitor and hold.
? Slowly turn on hydrogen.
? Release ignitor after hydrogen has been turned completely on. Slight
increase in signal should occur when flame is ignited. Alternatively, check
for lighted flame by holding mirror or other shiny object at exhaust end
of detector. Presence of condensed moisture indicates that flame is present.
? If flame does not light, turn off hydrogen and repeat previous steps.
Increasing air flow and/or decreasing carrier flow may help in igniting
flame.
Pesticide Analytical Manual Vol. I SECTION 503
503–13
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
? Turn on auxiliary gas if needed.
? Determine detector response by injecting 1.5 ng chlorpyrifos. Adjust
electrometer sensitivity so that 1.5 ng gives about 50% FSD (40-80% is
satisfactory). Adjust voltage to change response, if variable power supply
is available. At given voltage, changes in flow rates may improve sensitiv-
ity and chromatography.
FPD-S.
? Follow procedures above for FPD-P, but use 394 nm filter.
? If more than one 394 nm filter is available, test to determine which is
most selective to sulfur over phosphorus by injecting mixture of
methamidophos, chlorpyrifos, acephate, omethoate (each containing
S and P), and monocrotophos (P only). A filter that does not permit
response to monocrotophos, or that permits least response to it, is the
best choice for sulfur selectivity.
? For greatest sensitivity, do not use electrometer square root function;
instead, plot response vs amount injected on semilog paper and quan-
titate from that calibration. FPD-S is sufficiently insensitive that it should
be set up to provide the greatest sensitivity possible while still maintain-
ing reasonable baseline noise; this will vary from instrument to instru-
ment.
Troubleshooting
Consult the manufacturer’s operation and service manual for recommendations
specific to detector model being used. Note the following additional suggestions:
Symptom Possible Solution
Noisy and/or Install flow controllers to prevent gas flow fluctuation;
wandering baseline normal baseline is very straight with <1% noise in
P mode and <2% in S mode.
Check by shining flashlight on detector. Recorder
will show response if leak exists. Replace O rings.
If this does not work, seal light leaks with black tape
or other material. Photomultiplier tube should never
be exposed to light when connected to power supply
or it will burn out.
Clean detector.
Low sensitivity Check for photomultiplier tube light leaks as above.
Peak broadening or Improve chromatography by changing to capillary
tailing, poor response column or other column suited to chemistry of
reproducibility analyte.
Rejuvenate capillary column.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–14
SECTION 503 Pesticide Analytical Manual Vol. I
503 D: ELECTROLYTIC CONDUCTIVITY DETECTOR
The electrolytic conductivity detector (ElCD) is capable of operating in modes
selective to halogen, nitrogen, or sulfur. ElCDs can also be configured for selective
detection of nitrosamines or esters or for the nonspecific detection of carbon-
containing compounds.
For pesticide residue determination, the ElCD is most often used in the halogen
mode (ElCD-X), where it exhibits much greater selectivity to halogen than does
the EC detector and yet responds to <1 ng of most organohalogen pesticide resi-
dues in foods. Although ElCD in the nitrogen mode (ElCD-N) has been shown
suitably sensitive for use in residue determination, it is used routinely for that
purpose in only a few laboratories, because adequate operation is more difficult
to establish and maintain. In addition, problems associated with interferences
from nitrogen-containing commodity co-extractives (Section 503 A) apply to ElCD-
N.
Only ElCDs in the halogen and nitrogen modes are discussed in detail in this
section.
Presently, two manufacturers market ElCDs, Tremetrics and OI Corp., College
Station, TX. The Tremetrics “Hall 1000” and “Hall 2000” replaced the original
“Hall 700A,” which was marketed by Tracor Inc. (now Tremetrics); the latter
model is no longer commercially available but continues to be used in many
residue laboratories. OI markets the “4420” and a newer “5200.” FDA experience
is limited to the Hall 700A, Hall 1000, and OI 4420, so only these models will be
discussed in this chapter.
Principles
GLC column effluent is pyrolyzed in a nickel reaction tube at >800° C in the
presence of hydrogen reactant gas. Heat causes most of the compounds in the
reaction tube to be pyrolyzed to their elemental form, but the presence of reactant
gas results in other chemical reactions. Products formed during reaction of the
analytes are either removed by appropriate scrubbers prior to entering a conduc-
tivity cell or are swept into the conductivity cell via carrier gas where they are
dissolved in a circulating conductivity solvent (electrolyte).
In the conductivity cell, electrolyte conductivity is constantly monitored for changes
caused by dissolution of the reaction products. Change in conductivity is con-
verted to a voltage signal that produces an electrical peak at the detector output.
ElCD hardware is configured into various operating modes by appropriate selec-
tion of reactant gas, electrolyte, ion exchange resin, and chemical scrubbers used
to remove interferences. Detector sensitivity is affected by reaction conditions as
well as by electrolyte flow rate and reactant gas flow rate.
Design
Figure 503-i displays a block diagram of a GLC system with ElCD; terminology is
generalized to display the basic system arrangement that applies to all models of
ElCD; some design differences exist between models. Figure 503-j displays a sim-
plified diagram of the ElCD reactor and conductivity cell.
Pesticide Analytical Manual Vol. I SECTION 503
503–15
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
To
amplifier
Sample
Reaction
gas
Reactor
Heating
element
Detector
base
Vent
valve
Pump
Conductivity
solvent
Ion
exchange
cartridge
Conductivity cell
Transfer line
Reaction tube
Cell electrodes
Reaction
gas
Carrier
gas
Gas
chromatograph
Reaction
gas
control
Vent
timer
Temperature
controller
Reactor
Conductivity
amplifier
Conductivity
cell
Data
output
device
Resin
cartidge
Solvent
pump
Solvent
reservoir
Pump
controller
Figure 503-i
Block Diagram of the ElCD
[Reprinted with permission of John Wiley & Sons, Inc., from Detectors for Capillary Chromatography (Copyright ?1992)
Hill, H.H., and McMinn, D.G., ed., Chapter 6, by Hall, R.C., Figure 6.1, p. 111.]
Figure 503-j
ElCD Reactor and Conductivity Cell
[Reprinted with permission of John Wiley & Sons, Inc., from Detectors for Capillary Chromatography (Copyright ?1992)
Hill, H.H., and McMinn, D.G., ed., Chapter 6, by Hall, R.C., Figure 6.2, p. 112.]
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–16
SECTION 503 Pesticide Analytical Manual Vol. I
Each model ElCD includes four major units:
1) Reactor unit, including reactant gas inlet, nickel reaction tube, and
solvent vent to prevent injection solvent from entering the reaction tube.
2) Electrolyte unit, including solvent reservoir, ion exchange tube, and sol-
vent pump. The OI 4420 uses resin in two stages, a “roughing” resin
within the electrolyte reservoir and a “finishing” resin through which the
electrolyte passes before entering the conductivity cell.
3) Conductivity cell.
4) Signal processing unit, which converts conductivity cell signal for display
and provides power for other detector components.
The Hall 700A differs from other ElCDs in that it has a reference cell measuring the
conductivity of the electrolyte without any dissolved reaction products; electrolyte
conductivity is subtracted from that of the analytical cell, so that background signal
is removed from the final measurement. The Hall 1000 and OI 4420
measure absolute conductivity with no subtraction of signal related to electrolyte
conductivity.
ELCD-X
Principles
ElCD response in the halogen mode results from formation of HF, HCl, HBr, or
HI by catalytic reduction of analytes containing fluorine, chlorine, bromine, or
iodine, respectively. The heated nickel reaction tube provides all the ingredients
for reaction: a chamber for mixing analyte and hydrogen gas, heat, and the nickel
surface for catalysis.
Carrier gas transports the acid formed in the reaction tube into a conductivity cell.
The acid dissolves in deionized n-propanol electrolyte, increasing electrolyte con-
ductivity and producing a measurable response (peak) at the detector output.
To prevent neutralization of the acid formed in the reaction tube, pH of the n-
propanol electrolyte must also be slightly acidic. Electrolyte acidity is maintained
by circulation through ion exchange resin.
Apparatus and Reagents
Section 501 B provides general information on apparatus and reagents required
for GLC. Consult appropriate instrument manuals for purchasing information and
proper procedure for replacing the following reaction tubes, resins, scrubbers, and
electrolyte:
additional nickel reaction tubes. Tubes are purchased from the instrument
manufacturer or from other suppliers of nickel tubing. Use of tubing not
designed for analytical purposes will probably require additional cleaning
and/or polishing to be suitable [14]; laboratory prepared tubing has not been
successfully used in the OI 4420 models.
Pesticide Analytical Manual Vol. I SECTION 503
503–17
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
gases, helium, ultra high purity, 99.999%, used as column carrier gas; hydro-
gen, ultra high purity, 99.999%, used as reactant gas
gas filters, capable of removing oxygen and water from carrier gas
n-propanol, distilled from all-glass apparatus, for electrolyte
resin materials, to replace resin tube when selectivity deteriorates. Some
extra resin material should be provided with the detector and more can be
ordered from the manufacturer, or a system that replaces the whole unit
can be purchased to avoid the inconvenience of replacing the resin [15-17].
Consult appropriate detector manual for instructions on replacing resin.
(Resin and electrolyte are sold as a unit by Tremetrics, Inc., because the
resin bed is contained within the conductivity reservoir.)
Resins are subject to deterioration and are dated; they may be unsuitable
for use after storage, even if refrigerated as directed.
Detector Characteristics
Sensitivity. An FDA interlaboratory trial involving eight Hall 700A ElCDs in the
halogen mode, each operated at the same basic parameters, showed that each
detector was different in terms of the minimum amount of halogenated material
to which it would respond [18]. In this study, the most responsive detector was
10-25 times more sensitive than the least responsive detector to the same amount
of the same compound. However, each properly functioning detector was capable
of detecting 0.05 ppm lindane in the presence of sample extract. The more sen-
sitive detectors could readily measure 0.01 ppm lindane. No comparable study has
been performed with the other model ElCDs.
The following parameters affect ElCD sensitivity:
1) Reactant gas purity. Impurities in the reactant gas can undergo chemical
reaction in the reaction tube, and resulting products may be soluble in
the electrolyte. If this occurs, conductivity may be raised sufficiently
to obscure measurement of small amounts of analyte (i.e., the signal-
to-noise ratio will be reduced). Adherence to manufacturers’ purity
recommendations is critical. Gases meeting manufacturers’ specifications
occasionally contain traces of hydrochloric acid that can destroy the
detectors; use of appropriate gas filters is required even on high purity
gases.
2) Reactant gas flow rate. ElCD-X response was shown to increase with
increasing reactant gas flow rate, up to about 60 mL/min, in a study of
the Hall 700A ElCD [19]. Above that flow rate, response remained
essentially constant up to 100 mL/min. Manufacturers suggest reactant
gas flow rate of 50-75 mL/min for the Hall 700A [15], 25 mL/min for
the Hall 1000 [16], and 100 ± 10 mL/min for the OI 4420 [17]. FDA
laboratories routinely use 60-80 mL/min for the Hall 700A and the Hall
1000 and 100 mL/min for the OI 4420.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–18
SECTION 503 Pesticide Analytical Manual Vol. I
3) Reaction tube condition. The nickel reaction tube catalyzes the reaction
and must be free of contamination. After a period of use, nickel tubes
become contaminated because of fouling by bad gases, sample reaction
products, septum or column bleed, or other causes; chromatograms at
this point characteristically show a slow return to baseline after venting,
peak tailing, and loss of response. Once contaminated to this degree, the
tube must be replaced, because no successful reconditioning process has
been developed.
4) Reaction temperature. In a study using the Hall 700A in the halogen
mode, reaction temperature was not found to affect detector sensitivity
significantly. No significant differences in detector response occurred
when chlorinated compounds were injected at temperature control po-
tentiometer settings of 850, 730, and 630° C [19]. However, a minimum
potentiometer setting of 900° C is recommended for ElCD-X, because it
is suspected that potentiometer setting is not an accurate reflection of
the actual temperature of the reaction tube and because it is reasonable
to assume that more efficient reduction of halogen occurs at higher
temperatures. Reaction tubes whose operation is compromised by other
problems (e.g., contamination from samples, column bleed, or poor qual-
ity nickel tube) may show fluctuations in sensitivity with changes in
temperature.
Reaction furnaces (“reactors”) of OI 4420 detectors have been subject to
repeated burnout, requiring replacement. The manufacturer offers a
smaller, cartridge-style heating element as a reactor replacement in an
upgrade to the detector; this model is also expected to have a limited
lifetime but will be easier and less expensive to replace than the original
reactors.
5) Electrolyte flow rate. Electrolyte flow rate significantly affects detector
sensitivity by affecting the length of time the dissolved reaction products
spend in the conductivity cell. As the electrolyte flow decreases, response
increases. Below a certain flow rate, however, baseline noise increases and
further decrease in flow rate results in no additional improvement in the
signal-to-noise ratio.
Respective manufacturers recommend 0.5 mL/min electrolyte flow rate
for the Hall 700A, 0.6 mL/min for the Hall 1000, and 0.02-0.05 mL/min
for the OI 4420. FDA laboratories usually use 0.35 ± 10% mL/min flow for
the Hall 700A and the Hall 1000 and 0.035-0.050 mL/min for the OI 4420.
Selectivity. The ElCD is made selective to halogens by using hydrogen reactant gas
and n-propanol electrolyte. HX, formed by pyrolysis of halogenated compounds in
the presence of hydrogen, is readily soluble in n-propanol. Other compounds
formed in the reactor, such as H
2
S and NH
3
, do not usually cause detector re-
sponse because they are not ionized in n-propanol and therefore cannot change
solvent conductivity. No scrubber is needed to remove interfering reaction prod-
ucts from the gas flow in the halogen mode. Large quantities of nitrogen and
possibly carbon dioxide may cause a response, however.
(An optional oxidative mode operation for halogen selectivity, using oxygen as
reactant gas, is far less selective, produces greater noise, and requires use of scrub-
bers capable of removing SO
2
/SO
3
from the reaction products. This operation has
Pesticide Analytical Manual Vol. I SECTION 503
503–19
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
never been implemented for residue determination because it is far less preferable
than the reductive mode.)
Because detector selectivity to halogen is dependent on reaction conditions and
on solubility of reaction products in the electrolyte, all the following factors affect
selectivity:
1) Gas purity. Impurities in column carrier gas and/or hydrogen reactant
gas can introduce other chemical species that may interfere with halo-
gen detection; only ultrapure gases are acceptable. Gases must be free
of any level of halogenated compounds.
2) Reaction tube condition. The nickel reaction tubes that contain and
catalyze the reaction are known to vary from one another in their ability
to convert halogen to HX. A tube that initially produces an acceptable
response may deteriorate over a period of use because of contamination.
As previously discussed, the tube must be replaced when this occurs.
3) Reaction temperature. Reaction temperature may affect selectivity by
influencing the degree to which reduction of halogen to HX occurs.
Reaction temperature setting of 900° C is recommended to achieve
efficient reduction.
4) Ion exchange resin. The ion exchange resin affects selectivity by control-
ling pH of the electrolyte and by continuously removing ionized reac-
tion products from the electrolyte. The resin used for the halogen mode
maintains the n-propanol electrolyte at a slightly acidic pH. Presence of
the ionized HX then produces a measurable change in solvent conduc-
tivity. When the resin fails to control pH of the electrolyte appropriately,
certain species other than HX are also able to ionize, and detector
selectivity deteriorates. Failure to maintain slightly acidic electrolyte
results in negative or “V”-shaped peaks. Replacement of the resin with
fresh material re-establishes the necessary selectivity.
For unknown reasons, addition of n-propanol to the reservoir of the OI 4420, to
replace evaporated solvent, can cause severe damage to the resin. When this
occurs, resin and electrolyte must be replaced [20].
Linearity. Linear dynamic range of the Hall 700A ElCD-X varies with the com-
pound and with the individual detector [19]. Moreover, the typical degree of
linearity and length of linear range are not sufficiently reliable to eliminate the
need for matching peak heights of residue and standard when quantitating. Each
system should be tested to measure its linear range. For accurate quantitation of
residues, peak sizes must be within 25% of one another.
Other Influences on Detector Performance
Solvent Venting. ElCD reactor units include a vent line positioned just before the
heated reaction tube. The relatively large volume of injection solvent, eluting
through the column prior to the analytes, is diverted through this port to prevent
its entry into the reaction tube. Venting prevents combustion of hydrocarbon
solvent in the reaction tube and thus protects the tube from carbon deposition
that decreases catalytic performance, nickel tube lifetime, and detector response.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–20
SECTION 503 Pesticide Analytical Manual Vol. I
Contamination of the reaction tube and transfer line is most severe when acetone
is used and then not completely vented.
Efficiency of venting is affected by several factors, including reactant gas flow rate,
combustion tube diameter, and the percentage of flow that is vented. Both in-
creased flow of reactant gas and decreased combustion tube id improve vent
efficiency by increasing back pressure.
The percentage of total effluent that is vented is critical. While efficient venting
of most solvent is necessary, not all solvent can be vented, because lack of positive
pressure permits electrolyte to flow from the reservoir and enter the nickel reac-
tion tube, contaminating it. Vent rate is preset by the manufacturer, but the OI
4420 includes a vent system with a pressure-regulated “T” that permits the user to
adjust the vent rate by turning a threaded rod. Manufacturer directions specify
adjusting the rate so that about 50-60 mL/min total gas flow exits the vent; fre-
quent monitoring of the flow is necessary, because the vent split ratio fluctuates.
Many of the recurring problems with the OI 4420 detector were traced to the vent
system. It is now recommended that the original vent flow valve be replaced with
a constant flow port, which is capable of maintaining a constant vent flow while
the vent is open; the detector upgrade offered by the manufacturer includes this
replacement. Even this change, however, does not vent most of the solvent, and
its combustion in the reaction tube may cause subsequent problems [20].
Position of Capillary Column. Results observed during evaluation of a capillary
column with OI 4420 ElCD-X indicate that the most critical element for successful
operation is proper positioning of the column outlet in the reactor [21, 22]. When
the capillary column is installed as directed in the detector manual (i.e., column
outlet placed about 0.5" into the nickel reaction tube) a noisy baseline with
frequent “spiking” is observed.
These studies demonstrate that positioning the column outlet between the solvent
vent and the reactant gas inlet produces optimal results. In this position, the
column is outside the nickel tube and away from the extremely high temperatures
of the reactor. This positioning also ensures efficient venting, because the reactant
gas takes the path of least resistance and flushes the injection solvent through
the vent rather than through the small id transfer line to the combustion tube;
the possibility of tube contamination is thus reduced. A steady baseline is main-
tained when the column is installed between the vent and the reactant gas inlet.
This same positioning is also optimal when wide bore capillary columns are used.
A redesigned base for mounting the detector on the chromatograph may mini-
mize the importance of user attention to positioning the column. Operation of
ElCDs with packed columns is not as sensitive to column position.
Transfer Line Cleanliness. More often than not, broad, tailing peaks are caused
by a dirty transfer line between the reactor and the conductivity cell. Contami-
nants (i.e., unreacted hydrocarbons) can deposit in the transfer line and produce
active (adsorptive) sites. When hydrochloric or other gaseous acids pass through
the transfer line, they may be adsorbed; this phenomenon causes tailing or, in
severe cases, total loss of detector response. Transfer lines can be rinsed with the
injection solvent being used [23]; the ease with which this can be performed varies
with the detector model.
Pesticide Analytical Manual Vol. I SECTION 503
503–21
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Recommended Operating Procedures
Except for positioning the column in the reactor (above), follow directions
from the appropriate detector manual to set up and operate the GLC system with
ElCD. The following additional recommendations are based on experiences in
FDA laboratories:
? Reaction tube temperature: Set potentiometer that controls reaction
tube temperature to 900° C to ensure that temperature is high enough
to completely reduce analytes. To protect reaction tube from possible
deactivation by column bleed, do not allow reactor temperature to drop
below that of column.
? Reactant gas flow: Maintain hydrogen flow of about 60-100 mL/min
through reaction tube to ensure complete reduction of sample. Measure
flow rate using bubble meter and stopwatch, either at point where gases
enter conductivity cell (with column carrier gas turned off) or at return
line to solvent reservoir (with both column carrier gas and solvent pump
turned off). Column temperature should be reduced to room tempera-
ture if carrier gas is off for any length of time.
? Solvent flow rate: For optimum performance, pump n-propanol electro-
lyte through conductivity cell at 0.35 ± 10% mL/min for Hall 700A and
Hall 1000 and 0.035-0.050 mL/min for OI 4420 ElCD-Xs. Measure
flow rate by placing line that usually carries solvent to reservoir into a
graduated cylinder and measuring accumulation over known time.
System Suitability Test
Monitor detector selectivity by regularly injecting aliquot of mixed standard solu-
tion containing the following: 100 ng diisobutyl phthalate, 100 ng ethion, 1 ng
chlorpyrifos, 100 ng methyl palmitate, 100 ng caffeine, and 2 μg octadecane.
Properly operating detector will respond only to chlorpyrifos and possibly to caf-
feine. If response to caffeine is seen, calculate selectivity ratio as:
If selectivity ratio for chlorpyrifos:caffeine is <500:1 or if any response to other
compounds is seen, improve selectivity by following the suggestions for trouble-
shooting, below.
Routinely monitor detector response to halogen by injecting solutions containing
at least lindane, chlorpyrifos, and p,p′-DDT. If response decreases, follow direc-
tions in troubleshooting section to determine cause. While monitoring halogen
response, also note peak shapes on chromatograms. Deteriorating shape (i.e.,
increased tailing) of all peaks may be caused by various factors covered in trouble-
shooting section. Breakdown of p,p′-DDT (evidenced by smaller peak plus appear-
ance of another peak at retention time of p,p′-TDE) has been found to be caused
by prior injection of extracts prepared by the method of Section 302. This condi-
tion disappears over time if no further extracts from that method are injected.
detector response to chlorpyrifos × 100
detector response to caffeine
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–22
SECTION 503 Pesticide Analytical Manual Vol. I
Troubleshooting
Each detector manual [15-17] contains a section on troubleshooting that should
be consulted when problems occur. Another reference [24] also contains excel-
lent information on operation of the ElCD and potential problem areas. Beyond
the advice offered in those references, these additional suggestions for optimum
detector performance are offered:
Symptom Possible Solution
Selectivity of chlorpyrifos: Replace resin in cartridge and n-propanol
caffeine is <500:1 electrolyte.
Change nickel reaction tube. Several different
tubes may have to be tried, because each converts
halogen to HX to different extent.
Elevate reaction tube temperature to determine
whether sample is being completely reacted.
Replace column if liquid phase contains halogen
or nitrogen.
Loss of detector sensitivity Verify purity of gases; use only ultra high purity
gases.
Remove transfer line and rinse with injection
solvent or replace transfer line.
Replace column if liquid phase contains halogen
or nitrogen.
Breakdown of p,p′-DDT Remove first 1-2" of packing from GLC column
to p,p′-TDE and replace with fresh, conditioned packing, i.e.,
clean front end of column (Section 502).
Broad, tailing peaks Replace or clean transfer line.
Clean front end of column.
Replace nickel reaction tube.
Replace column if liquid phase contains halogen
or nitrogen.
Slow return to baseline Replace nickel reaction tube.
after venting
Breakdown of analytes Clean front end of column.
(but normal return to
baseline after venting,
normal peak shape)
Pesticide Analytical Manual Vol. I SECTION 503
503–23
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
ElCD-N
Principles
ElCD response in the nitrogen mode results when organic nitrogen is pyrolyzed
to ammonia with hydrogen reactant gas at high temperature. Acidic or sulfur
gases, resulting from pyrolytic reduction of halogen and sulfur compounds in the
reactor, are selectively removed by quartz threads coated with potassium hydrox-
ide. Carrier gas transports the ammonia into the conductivity cell, where it dis-
solves in an aqueous electrolyte to form ammonium hydroxide:
NH
3
+ H
2
O —> NH
4
+
+ OH
–
As a weak base, ammonium hydroxide readily ionizes in the aqueous electrolyte
and becomes a conducting species. The change in electrolyte conductivity caused
by dissolution/ionization of ammonia produces a measurable response (peak) at
the detector output.
Apparatus and Reagents
Section 501 B provides general information on apparatus and reagents required
for GLC. The section above on ElCD-X provides additional information about
replacement nickel reaction tubes, gas purity, and gas filters. Consult appropriate
instrument manuals for purchasing information and proper procedure for replac-
ing the following reaction tubes, resins, scrubbers, and electrolyte:
stainless steel, nickel, or copper gas lines from gas cylinders to the
instrument
water, HPLC grade, prepared from water purification equipment, or
equivalent commercial product; 18 megaohm resistance required
fresh electrolyte, prepared from reagent grade t-butanol and HPLC grade
water
additional scrubbers for use in nitrogen mode. Nitrogen scrubber generally
lasts 3-6 mon. However, under certain circumstances scrubber may last
<1 mon.
resin, may be ordered from detector manufacturer, as described above for
ElCD-X. Nitrogen mode operation is particularly sensitive to failures caused
by resin deterioration, so use of fresh resin is critical.
Detector Characteristics
Sensitivity. The ElCD-N detector is capable of producing as much as 50% FSD
response to 1 ng carbaryl or 0.5-1 ng chlorpyrifos. As with the ElCD-X operation,
sensitivity of the ElCD-N depends on reaction conditions, reactant gas flow rate,
and electrolyte flow rate. In addition, the condition of the chemical scrubber used
in the nitrogen mode will affect detector sensitivity.
The control module (signal processing unit) of the OI 4420 offers several sensi-
tivity settings, labelled according to the different modes of operation. Despite the
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–24
SECTION 503 Pesticide Analytical Manual Vol. I
label, the setting for the halogen mode should be used during nitrogen mode
operation for greatest sensitivity.
Selectivity. Selectivity for N:P is much better than that of the N/P detector
and is the chief virtue of the ElCD-N; selectivity of the OI 4420 in the nitrogen
mode was found to be 2400:1 for chlorpyrifos (molecular formula
C
9
H
11
Cl
3
NO
3
PS):bromophos (C
8
H
8
BrCl
2
O
3
PS) [25]. Selectivity for N:X is also very
high as long as the scrubber is efficiently removing HX from the reaction products
prior to dissolution in electrolyte; N:X selectivity is higher in this mode than is X:N
selectivity in the halogen mode. Selectivity was found to be 4800:1 for chlorpyrifos:
aldrin (C
12
H
8
Cl
6
)[25]. As with the ElCD-X operation, selectivity of the ElCD-N is
affected by reaction conditions and by parameters that affect ionization of the
reaction products in the electrolyte.
Any parameter that affects the conversion of organic nitrogen to NH
3
influences
selectivity. The most important of these parameters are reactant gas purity, condi-
tion of the reaction tube, and reaction temperature. In addition, the chemical
scrubber (i.e., quartz threads coated with potassium hydroxide), placed between
the reactor and conductivity cell, prevents interferences from reaching the electro-
lyte, dissolving in it, and causing detector response.
Conditions that affect ionization of the reaction products, including nature of the
ion exchange resin and electrolyte type and pH, also influence detector selectivity.
Electrolyte pH must be slightly basic to prevent neutralization of the basic ammo-
nium hydroxide; appropriate pH is maintained by passing the electrolyte through
an ion exchange resin before it enters the conductivity cell. Aqueous electrolyte
is used in the nitrogen mode because water is one of the few neutral solvents
capable of ionizing a weak electrolyte like ammonia. Water purity is critical to
detector selectivity; the high purity water specified above in Apparatus and Re-
agents is necessary.
Incorporation of carbon dioxide in the electrolyte affects its pH and thus detector
selectivity. Carbon dioxide can enter the electrolyte from improper venting or
from permeation through tubing during transfer to the conductivity cell. Early
versions of the OI 4420 included a permeation chamber filled with ammonia in
water, through which the Teflon tubing transferring the electrolyte passed. This
arrangement was used only for nitrogen mode operation, with the intent of per-
mitting ammonia to permeate the electrolyte and keep it sufficiently basic, but it
was not satisfactory and the permeation chamber is no longer included with the
system. The Hall 2000 uses a stainless steel transfer line to minimize permeation
of gases into the electrolyte. Purging the electrolyte with hydrogen or helium for
1 hr, after t-butanol and water are mixed, may also be used to dispel carbon
dioxide and air and improve detector performance [26].
Linearity. The linearity of ElCD-N detector response to any particular chemical
is approximately three orders of magnitude, within the range of 10 pg-100 ng.
Response to each chemical, depending on its chromatography and percentage
nitrogen, has a lower threshold of linearity; below that level, response can be
measured but is not linear. Response to an amount beyond the upper limit of
linearity often appears as a double peak [20].
Other Influences on Detector Performance
Factors that influence operation of the ElCD-X detector, i.e., venting efficiency
and transfer line cleanliness, are also important parameters in ElCD-N operation,
Pesticide Analytical Manual Vol. I SECTION 503
503–25
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
although use of a scrubber in the nitrogen mode minimizes transfer line contami-
nation. As with the ElCD-X, the optimal position for a capillary column in the OI
4420 detector was found to be between the vent line and the reactant gas
inlet [25]. In addition, the detector operations described below are critical to
acceptable detector operation.
Composition of Electrolyte. An important parameter for acceptable performance
is electrolyte composition [25]. During early studies of the OI 4420, 50% n-
propanol/water was recommended as a replacement for 0.1% hexanol/deionized
water, originally recommended by the manufacturer. In the meantime, however,
10% t-butanol/water has become the recommended electrolyte for OI 4420 nitro-
gen mode operation; 50% n-propanol/water is recommended for Hall 700A and
Hall 1000 detectors. As discussed above, water purity is critical in the nitrogen
mode.
Condition of Scrubber. The scrubber can become exhausted and must be re-
placed when the detector begins responding to halogenated compounds. Solvents
containing halogen or sulfur should not be used in the nitrogen mode because
they will rapidly deplete the scrubber.
Recommended Operating Procedures
Follow the explicit directions from the appropriate detector manual to set up and
operate the GLC system with ElCD, and incorporate special directions discussed
above, including use of high purity (18 megaohm resistance) water and hydrogen
purging of the mixed electrolyte prior to use, whenever electrolyte is changed.
System Suitability Test
Currently, there is no standardized system suitability check performed by
FDA laboratories for ElCD-N detectors. Suggested system suitability tests may be
found in detector operation manuals. It is recommended that a solution contain-
ing at least one compound containing nitrogen as the only heteroatom, one
halogenated compound, and one hydrocarbon be injected into the system; a prop-
erly functioning system should show no response to the halogenated compound
or to the hydrocarbon and should have no inverted (below baseline) peaks.
Troubleshooting
Detector operations manuals and Reference 24 each contain sections on trouble-
shooting. In addition, the following suggestions for optimum detector
performance are based on the FDA evaluation of wide bore column and OI 4420
ElCD-N for determination of nitrogen-containing pesticide residues in food:
Symptom Possible Solution
Peak tailing Replace scrubber.
Replace nickel reaction tube.
Replace older OI 4420 detector base.
Poor linearity Replace scrubber.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–26
SECTION 503 Pesticide Analytical Manual Vol. I
Replace electrolyte.
Replace older OI 4420 detector base.
Excessive noise Replace electrolyte.
Replace gas line filters.
Replace gas.
Check for temperature fluctuations and correct as necessary.
Replace nickel reaction tube.
Check for and correct gas flow instabilities.
Remove bubbles in OI cell by turning pump switch off for
1-2 sec, then turning on, or increase pump speed to maximum
for 1-2 min.
General Precautions for ElCDs
The following precautions should be followed to ensure optimum performance of
the ElCD in both halogen and nitrogen modes:
? Avoid column liquid phases that contain halogen or nitrogen, because
the phase may bleed and de-activate the reaction tube and/or raise the
conductivity of the electrolyte.
? Avoid injecting standards or sample extracts in solvents containing
halogen or nitrogen. Even though the solvent is vented, traces may re-
main and affect detector operation. This effect becomes critical in cases
where detector selectivity is already poor.
? Maintain constant carrier gas and reactant gas flow at all times. Reducing
gas flows overnight to conserve gas may result in diminished responses
when detector conditions are re-established the next day.
? When carrier gas flow must be interrupted, e.g., to change columns or
septa, cool reactor furnace first. Exposure of nickel reaction tube to
oxygen at high temperature invariably damages performance and usually
requires subsequent replacement of tube. Before reheating furnace, thor-
oughly purge system with carrier gas; 15 min is sufficient when capillary
columns are used.
? Do not allow solvent return line to dip below surface of solvent in reser-
voir. Violating this rule will lead to backup of solvent into reaction tube
anytime gas flow is inadvertently stopped.
? Vent injection long enough to ensure removal of solvents or volatile sample
co-extractives that can interfere with determination. Adequate venting also
protects reaction tube and conductivity cell. Vent time of 0.5-0.75 min is
adequate for wide bore column and Hall 700A detector; 0.75->1.3 min is
required for OI 4420.
Pesticide Analytical Manual Vol. I SECTION 503
503–27
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
503 E: NITROGEN/PHOSPHORUS DETECTOR
The nitrogen/phosphorus (N/P) detector is selective to residues containing
nitrogen and/or phosphorus atoms. Modern N/P detectors evolved from Kolb
and Bischoff’s 1974 design [27], itself an evolution of the potassium chloride
thermionic detector (KClTD); the KClTD, introduced in the mid-1960s, was the
first selective detector for phosphorus residues [28]. Most N/P detectors are more
responsive to phosphorus than to nitrogen, but this section emphasizes use as
a nitrogen-selective detector, because the FPD-P (Section 503 C) is preferred for
phosphorus residues.
Although N/P detectors are selective and sensitive, problems associated with their
reliability and performance have deterred their routine application for pesticide
residue determination in FDA laboratories. In addition, the N/P’s ability to distin-
guish residues from sample matrix unequivocally is hindered by the presence of
nitrogen in many commodity co-extractives, a dilemma common to all nitrogen
detectors. Despite these shortcomings, an N/P detector, optimized for nitrogen
selectivity, can play a valuable role in examining extracts for residues; many pes-
ticides contain no other heteroatom than nitrogen. Response of the N/P detector
also provides complementary evidence about element(s) present in a residue,
information often needed for confirmation of identity (Section 103, Table 103-a).
Several different manufacturers produce N/P detectors. Among these are:
Chrompack, Inc., Raritan, NJ; DETector Engineering & Technology, Inc., Walnut
Creek, CA; Hewlett-Packard Company, Wilmington, DE; Perkin Elmer Corpora-
tion, Instrument Division, Norwalk, CT; Shimadzu Scientific Instruments, Inc.,
Columbia, MD; Tremetrics, Inc., Austin, TX; and Varian Instrument Division, Walnut
Creek, CA.
Principles
GLC column effluent impinges onto the surface of an electrically heated and
polarized alkali source in the presence of an air/hydrogen plasma; ionization
occurs and the flow of ions between plasma and an ion collector is amplified and
recorded. Detector response to analytes results from the increased ionization that
occurs when compounds containing nitrogen or phosphorus elute from the
column. At gas flow rates used for N/P operation, the degree of ionization
of compounds containing nitrogen or phosphorus is >10,000 times greater than
for hydrocarbons. Mechanisms that explain the enhanced response to nitrogen
and phosphorus are not yet fully understood and are beyond the scope of this
manual; both gas phase ionization and surface ionization processes have been
proposed [29].
Design
An N/P detector is similar to an FID to which an electrically heated source of
alkali has been added between the jet and the ion collector; Figure 503-k provides
a schematic diagram of typical components. Commercially available N/P designs
vary considerably, with different collector electrodes, collector polarity, and
optimum potential between jet and collector; Figure 503-l displays several of these
variations. The most important component, the alkali source, is usually manufac-
tured by impregnating a glass or ceramic matrix with an alkali metal salt. Varia-
tions among alkali source designs represent attempts to optimize selectivity to
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–28
SECTION 503 Pesticide Analytical Manual Vol. I
Figure 503-k
N/P Detector Components
Sample
inlet
Thermionic
ionization
source
Air
H
2
(4 – 6 mL/min) + sample
Heating
current
Bias
voltage
Electrometer
GRND
Ion
collector
–
[Reprinted with permission of John Wiley & Sons, Inc., from Detectors for Capillary Chromatography (Copyright ?1992)
Hill, H.H., and McMinn, D.G., ed., Chapter 7, by Patterson, P. L., Figure 7.1, p. 142.]
nitrogen (selectivity to phosphorus over hydrocarbons is adequate for most de-
signs), detector operating stability, and source ruggedness for extended operating
life. Some but not all detector models permit adjustment of the alkali source
height above the jet for optimization of sensitivity and selectivity.
All N/P detectors provide electronic heating of the alkali source to 600-800° C.
The plasma in the region of the salt is sustained by flows of hydrogen and air. The
alkali source exhibits longer operating life and more stable and reproducible
response under these conditions than in the presence of a flame.
Apparatus and Reagents
Section 501 B provides general information on apparatus and reagents required
for GLC.
Detector Characteristics
Sensitivity. N/P detectors are capable of producing detectable peaks in response
to as little as 5-10 pg nitrogen-containing compounds or to 1-5 pg phosphorus-
containing compounds [30]. FDA experience indicates that the pesticide for which
the greatest N/P response occurs is diazinon, which contains two nitrogen atoms
and one phosphorus atom; 25 pg diazinon should cause a response of approxi-
Pesticide Analytical Manual Vol. I SECTION 503
503–29
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
A
C
S
I
B
C
outer
C
inner
S
I
C
C
S
I
(C-ion collector, S-thermionic source, I-sample inlet)
D
S
I
C
Figure 503-l
N/P Detector Configurations
[Reprinted with permission of John Wiley & Sons, Inc., from Detectors for
Capillary Chromatography (Copyright ?1992) Hill, H.H., and McMinn,
D.G., ed., Chapter 7, by Patterson, P. L., Figure 7.2, p. 144.]
mately 5% FSD; tebuthiuron,
with four nitrogen atoms (no
phosphorus), requires about 50
pg for 5% FSD response when
chromatographed on a wide
bore capillary column at stan-
dardized sensitivity (Appendix I,
PESTDATA).
N/P sensitivity is most influ-
enced by hydrogen flow and the
magnitude of heating current
supplied to the alkali source.
Alkali source position may affect
sensitivity, but not all designs
permit adjustment of source
height. Response diminishes
over the lifetime of the alkali
source. Variations in response
are also seen among individual
alkali sources.
Detector response to nitrogen is
most affected by hydrogen flow
rate, with response increasing as
hydrogen flow decreases; opti-
mum flow for the particular
detector must be determined
experimentally. Typical hydro-
gen flow rate for optimum ni-
trogen sensitivity is 1-5 mL/min.
Response also increases with
increasing alkali source current,
but little improvement is realized, because detector background can also increase.
Lifetime of the alkali source may also be shortened by operation at higher current.
Response to nitrogen compounds is not strictly proportional to the amount of
elemental nitrogen in the molecule; variations based on molecular structure oc-
cur. Although the reactions that occur within the detector plasma are effective in
decomposing analytes into common species, those compounds that easily decom-
pose to the cyano radical usually cause higher response than do amides or nitro
compounds [29].
Selectivity. Selectivity of the N/P detector is about 10
3
–10
5
for N:C response,
10
4
–5 × 10
5
for P:C response, and 0.1–0.5 for N:P response [29]. Factors that
affect detector sensitivity do not always affect selectivity similarly. While both
sensitivity and selectivity to nitrogen improve with decreasing hydrogen flow,
only sensitivity (but not selectivity) improves with source heating current, because
background noise and response to other elements increase simultaneously.
Linearity. Manufacturers of N/P claim linearity of response over four or five
orders of magnitude. No FDA studies have been done on modern N/P detectors
to measure detector linearity relative to amount of pesticides. Laboratories using
N/P detectors must evaluate the linear range, work within that range, and match
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–30
SECTION 503 Pesticide Analytical Manual Vol. I
peak sizes of residue and standard within 25% for accurate quantitative determi-
nation.
Other Influences on Detector Performance
Detector Temperature. Detector output is very sensitive to temperature changes
within the active zone where ionization occurs; for stability of operation, condi-
tions that permit variations in temperature should be avoided. Temperature of the
alkali source is controlled by the electrical current at which it is operated but is
also affected by hydrogen flow and, to a lesser extent, the rate of air and column
carrier gas flowing past the source. The detector walls are heated separately.
Stability is improved when the N/P detector itself is operated at a high tempera-
ture, because this minimizes the temperature gradient between the alkali source
and the surrounding wall; reducing the gradient minimizes change in source
temperature that occurs when high concentrations of analytes pass through the
detector [29].
Age of Alkali Source. Each alkali source has a finite lifetime; eventually each must
be replaced. Both sensitivity and selectivity decrease as the source ages, so regular
calibration of detector performance is required. Source activity can be conserved
by reducing hydrogen flow when the detector is not in use; however, manufacturer’s
instructions regarding source current and gas flow must be followed carefully
to avoid destruction of the source. Operation of the detector at the lowest source
current compatible with desired sensitivity is also recommended, as is mainte-
nance of the detector at 100-150° C when not in use to prevent water condensa-
tion. Because degradation occurs more rapidly with higher source heating
current, increasing the electrometer sensitivity to maintain constant detector
sensitivity is preferable to increasing source heating current [29].
Replacement of the alkali source and re-establishment of optimum operating
conditions can be troublesome and time-consuming with some detector designs.
Design quality is at least partly judged by the stability of the source itself and even
more so by the ease with which the source can be replaced and stable operation
re-established.
Gas Flow Stability. Stable flow of hydrogen and air is critical for constant and
linear response. High precision gas flow valves, standard equipment on some
chromatographs, may be required for acceptable operation.
Position of Column Outlet. For maximum sensitivity and optimal peak shape,
the GLC column should be positioned about 1-3 mm from the tip of the detector
jet. The column should not protrude into the flame, because the polyimide coat-
ing on capillary columns will decompose and the resulting nitrogen products
cause high background signal and noise. If the column outlet is too far below the
tip, peaks may tail and/or be reduced in size because of the dead volume between
the column and the alkali source [30].
Solvents and Reagents. Use of certain materials can have a detrimental effect on
efficient operation of N/P detectors and should be avoided. For example, injection
of extracts containing even trace amounts of acetonitrile can cause large detector
response and preclude examination of the early eluting portion of the chromato-
gram; such extracts must be evaporated or azeotroped to remove all acetonitrile
before injection. In addition, halogenated solvents may destroy the alkali source
Pesticide Analytical Manual Vol. I SECTION 503
503–31
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
and thus should not be injected. (Some N/P detectors are designed to permit use
of halogenated solvents, but this must be ascertained prior to injection.)
Packed column stationary phases containing cyano groups (e.g., OV-225) are un-
acceptable for use with N/P detectors. Equivalent bonded phase capillary
columns have little bleed and may be acceptable, however. Other materials to
avoid include those known to cause problems in many GLC systems, e.g., septa not
designated for high temperature use, impurities in gases, and leak-detecting
solutions.
Certain common materials can appear as contaminants in determinations using
N/P detectors. Nicotine is usually detected when cigarette smoking occurs in the
vicinity; if phosphate detergents are used to wash glassware, or if the GLC column
or glass wool is treated with phosphoric acid, trace amounts remain and are de-
tectable during determination.
Recommended Operating Procedures
The following directions, adapted from the instrument manual for one N/P de-
tector [31], have not yet been tested within FDA but are proposed as a way of
optimizing detector operation:
? Follow manufacturer’s directions for installation and operation. Pay
particular attention to recommendations related to the alkali source,
including situations that should be avoided to prevent its destruction.
Use of wide bore capillary column with retention gap (Section 502 C)
is recommended; makeup gas should not be necessary if column carrier
flow rate of 10-20 mL/min is used.
? Follow manufacturer’s directions to establish detector operation selec-
tive to nitrogen. Adjust detector parameters and instrument attenuation
so that 1.0 ng chlorpyrifos causes 50% FSD.
? Prepare test solution containing 2.0 ng/μL azobenzene (containing 310
pg N), 2.0 ng/μL parathion-methyl (110 pg N and 230 pg P), 4.0 ng/
μL malathion (380 pg P), and 4 μg/μL n-heptadecane (3.4 μg C) in
isooctane.
? Inject 1 μL test solution, and adjust detector attenuation and range to
keep peaks on scale. Examine relative responses of detector to four
components; negative deflection of pen is normal in area of solvent
peak.
? Experiment with effect of hydrogen flow on detector selectivity to
nitrogen by re-injecting test solution after changing hydrogen flow rate
in increments of 0.5 mL/min.
? Based on experimental results, use hydrogen flow rate that produces
greatest ratio of response for parathion-methyl:malathion, as long as
azobenzene peak is ≥4 times heptadecane peak at that flow; malathion
peak can be expected to always be larger than parathion-methyl peak.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–32
SECTION 503 Pesticide Analytical Manual Vol. I
References
[1] Standard Practice for Gas Chromatography Terms and Relationships (1992) ASTM
E 355-77, reapproved 1989, ASTM, Philadelphia, PA
[2] Standard Practice for Use of Electron-Capture Detectors in Gas Chromatography (1992)
ASTM E 697-91, ASTM, Philadelphia, PA
[3] Radiation Safety Handbook for Ionizing & Nonionizing Radiation (Oct. 1976)
HEW Publication (FDA) 77-8007, Washington, DC
[4] Gaeta, N., et al. (Dec. 1977) “Servicing Tritium and Nickel-63 Electron Cap-
ture Detectors,” LIB 2151, FDA, Rockville, MD
[5] Clower, M.G., Jr. (March 1980) “Characterization of a Constant Current
63
Ni
Electron Capture Detector for Application to Pesticide Residue Analysis,”
Pittsburgh Conference, Atlantic City, NJ
[6] HP 5890 Series II Operating Manual 05890-90260 and HP 5890 Series II Reference
Manual 05890-90270 (Jan. 1990) Hewlett-Packard Co., Avondale, PA
[7] Sawyer, L.D. (May 1980) private communication, FDA, Minneapolis, MN
[8] Tracor Linearized Electron Capture Detector System Operation Manual, 115314F
(Feb. 1981) Tremetrics Inc, Austin, TX
[9] Varian Electron Capture Detector Manual 03-914087-00:B (Sept. 1986) Varian
Associates, Inc., Walnut Creek, CA
[10] Varian Flame Photometric Detector Manual 03-914088-00 (July 1986) Varian Asso-
ciates, Inc., Walnut Creek, CA
[11] Tracor Flame Photometric Detector Operation Manual 116538C (Aug. 1986)
Tremetrics, Inc., Austin, TX
[12] HP 5890A Gas Chromatograph Accessory 19256A Flame Photometric Detector, HP
5890A Option 240, 19256-90100 (Nov. 1986) Hewlett-Packard Co., Avondale,
PA
[13] Krick, F. (1992) private communication, FDA, Bothell, WA
[14] Ward, P.M. (Sept. 1986) “A Cleaning Procedure for Nickel Reaction Tubes
Used in the Hall Electrolytic Conductivity Detector,” LIB 3086, FDA, Rockville,
MD
[15] 700A Hall Electrolytic Conductivity Detector: Operations Manual 115448B (May
1979) Tracor Instruments, Austin, TX
[16] Tremetrics Model 1000 Hall Detector System: Operation and Service Manual 117451P
(Nov. 1990) Tremetrics, Inc., Austin, TX
[17] OI 4420 Electrolytic Conductivity Detector: Operating and Service Procedures (July
1988) OI Corp., College Station, TX
Pesticide Analytical Manual Vol. I SECTION 503
503–33
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
[18] Carson, L.J. (June 1982) “Hall 700A Electrolytic Conductivity Detector
Interlaboratory Study,” LIB 2473A, FDA, Rockville, MD
[19] Carson, L.J. (Jan. 1981) “Hall Electrolytic Conductivity Detector Halogen
Mode: Evaluation,” LIB 2473, FDA, Rockville, MD
[20] McGill, B. (1993) private communication, FDA, Dallas, TX
[21] Gilvydis, D.M., et al. (July 1986) “Preliminary Evaluation of O.I.C. Model 4420
Electrolytic Conductivity Detector in Halogen Mode with Capillary GC,” LIB
3064, FDA, Rockville, MD
[22] Fehringer, N.V., et al. (Dec. 1986) “Evaluation of the O.I. Corporation Model
4420 Electrolytic Conductivity Detector with High Resolution Gas Chroma-
tography in the Splitless Mode,” LIB 3100, FDA, Rockville, MD
[23] 4420 ELCD Troubleshooting: Transfer Line Rinsing (1991) OI Newsworthy No:
03080891, College Station, TX
[24] Anderson, R.J. (1982) The Hall Book, Tracor Instruments, Austin, TX
[25] Fehringer, N.V., et al. (1992) J. High Resolut. Chromatogr. 15, 124-127
[26] Froberg, J. (1993) private communication, FDA, Los Angeles, CA
[27] Kolb, B., and Bischoff, J. (1974) J. Chromatogr. Sci. 12, 625-629
[28] Giuffrida, L. (1964) J. Assoc. Off. Agric. Chem. 47, 293-300
[29] Patterson, P.L. (1992) “The Nitrogen-Phosphorus Detector,” in Detectors for
Capillary Chromatography, Hill, H.H., and McMinn, D.G., ed., Wiley, New York
[30] Rood, D. (1991) A Practical Guide to the Care, Maintenance, and Troubleshooting
of Capillary Gas Chromatographic Systems, Huethig Buch Verlag, Heidelberg,
Germany
[31] Varian Thermionic Specific Detector Manual 03-914089-00 (1987) Varian Associ-
ates, Inc., Walnut Creek, CA
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)503–34
SECTION 503 Pesticide Analytical Manual Vol. I
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 504–1
Pesticide Analytical Manual Vol. I SECTION 504
504: QUANTITATION
504 A: INTRODUCTION
Accurate quantitation of pesticide residues identified by GLC is always of critical
importance. Whether the analysis has been performed for purposes of monitoring
or for enforcement of regulations, the consequences always have potential long
term impact. All analyses that claim to produce quantitative results must be calcu-
lated in a consistent, reliable manner.
Accurate quantitation depends on use of accurate reference standards, use of a
GLC system whose response is linearly proportional to the weight of chemical
being detected (or for whose nonlinear response adjustment can be made), and
use of proper technique for measuring detector responses. Given these conditions,
quantitation is based on a simple proportion equation, i.e.:
Quantity of analyte, the unknown value, is readily calculated from the known
quantity of standard and the measured detector responses.
This section assumes that the first two conditions for accurate quantitation, i.e.,
accurate reference standards and a linear GLC system, are met. Only techniques
for measuring detector response are discussed here.
Measurement of detector response for use in the above formula has traditionally
involved manual measurement of the peak that represents detector response on
a chromatogram drawn by a strip chart recorder. Section 504 B provides directions
for the most practical ways of manually measuring peaks.
Modern automated data handling systems electronically integrate the detector
output signal and produce a numerical representation of peak size. Step-by-step
directions for such systems are not included in this manual, however, because each
is unique; analysts using electronic integration must follow the directions provided
by the manufacturer. Section 504 C provides general guidance to the appropriate
application of electronic integration and advice about avoiding pitfalls that can
occur.
Whether the detector response (peak) is measured manually or electronically,
proper positioning of the baseline below the peak is critical. Accuracy of the
measurement depends in part on how well the detector’s response to the residue
can be distinguished from its response to sample co-extractives and co-eluting
residues. Typically, a residue peak in a sample chromatogram may occur on a
sloping baseline, on top of another peak, or incompletely separated from another
peak; in contrast, the reference standard solution usually causes a single symmetri-
cal peak. Quantitative accuracy is sacrificed if the residue peak’s baseline is not
properly delineated. To measure peaks manually, the analyst must literally draw
the baseline on the chromatogram before measuring; to use automated measure-
ment, the analyst must configure the system to include only that part of the signal
that can reasonably be assumed to represent the residue.
quantity of analyte
detector response to analyte
quantity of standard
detector response to standard
=
Pesticide Analytical Manual Vol. ISECTION 504
504–2
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Appropriate setting of the baseline is integral to the directions below for measur-
ing the peaks. In some cases, choice of appropriate baseline for particular residues
will be shown by example.
504 B: MANUAL QUANTITATION
Many methods of quantitating gas chromatographic peaks have been presented
in the literature, but through the years most laboratories that perform manual
measurements of peaks have relied on the two simplest approaches: measurement
of peak height and measurement of area of the triangle that best fits the peak
(“triangulation”). Peak area is the more accurate representation of detector
response, but peak height is a justifiable approximation of area when peak shape
makes height proportional to area. Advances in column techniques (Section 502)
have resulted in improved peak symmetry and resolution, thus encouraging use
of peak height for quantitation.
Other techniques for manual measurement of peaks have been described in
various chromatographic texts; these include calculation of the product of peak
height and width at half height, product of retention time and peak height, weight
of peaks cut from chromatogram, peak area measured by a planimeter, and peak
area measured by a mechanical integrator attached to the recorder. Comparison
of results among some of these techniques indicated their validity [1], but none
are described in this section because they are time-consuming and more difficult
and offer no significant advantage over those presented here.
Measurement of Peak Height
Peak height measurements are recommended for early eluting peaks, peaks of
width <10 mm, and very small peaks. If analyte and reference standard peaks are
narrow and approximately the same size, comparison of peak heights is less sub-
ject to measurement error than is triangulation. Peak height measurements are
very sensitive to changes
in operating conditions,
so operating parameters
must be closely con-
trolled for accurate
quantitation.
To measure peak height,
construct a baseline be-
neath the peak and mea-
sure the length of the
perpendicular from peak
apex to midpoint of the
constructed baseline. In
Figure 504-a, this is rep-
resented by line AB on
Peak 1.
Measurement of Area by Triangulation
Measurement by triangulation involves drawing a triangle that approximates a
peak’s dimensions and calculating the area of the triangle. This method requires
EFD
2
C
B
A
1
Figure 504-a
Manual Peak Measurement
Measurement by: (1) peak height and (2) triangulation.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 504–3
Pesticide Analytical Manual Vol. I SECTION 504
Figure 504-b
Triangulation of Peak on Sloping Baseline
extreme care in construction of the triangle and in measuring its dimensions.
Special treatment is required for peaks on sloping baselines and for skewed (asym-
metrical) peaks. The technique is subject to error when the peak is narrow but is
preferred over measurement of peak height when the peak is >10 mm wide at the
base.
To construct the triangle, draw a baseline below the peak and draw inflection
tangents to the peak, as shown in Figure 504-a, Peak 2. Drop a perpendicular
bisect from the constructed apex to the baseline. Measure triangle height (length
of bisect from baseline to constructed apex, CF in the figure) and the base (length
of baseline between its intersection
with the tangents, DE in the figure).
Calculate area as 1/2 (base × height),
i.e., 1/2 (DE × CF).
When the chromatogram baseline
slopes under a peak, the line
dropped from the intersection of
tangents does not serve as an accu-
rate measure of height because it is
not perpendicular to the baseline;
e.g., Figure 504-b, line AD in triangle
ABC. To measure the area of such a
triangle, draw a perpendicular to
one of the tangents (CE in Figure
504-b). Then use its length as the
triangle height and the length of the
tangent (AB) as the base. Calculate
area from these values using the standard formula, i.e., 1/2 (AB × CE).
Skewed peaks present another challenge to the validity of area measurement by
triangulation. As a peak becomes more skewed, less and less of its area is included
within the triangle drawn to approximate it. Skewed peaks may be tailing or fronting,
depending on what physical phenomena caused the poor chromatography.
The preferred solution to quantitation of skewed peaks is to improve chromatog-
raphy sufficiently to cause peaks to be symmetrical. Use of a more polar column,
changing column or inlet temperature, or optimizing the injection system may
effect the improvement.
If manual quantitation must be performed on a skewed peak, measurement of the
area using the formula for calculating area of a trapezoid is preferred [2]. In this
system, peak widths at 15 and 85% of height are measured and used in the for-
mula: area = 1/2 × height × (width at 15% + width at 85%). Calculations
performed in this way have been shown to accurately represent peak area even for
increasingly skewed peaks [3].
504 C: ELECTRONIC INTEGRATION
Electronic integration devices provide laboratories with powerful tools to accom-
plish their work more efficiently. Over the years, technology has progressively
improved from simple desktop integrators to software programs operated by
computers at all capability levels. The more powerful “automated data handling
systems” can automate the entire determinative step, including monitoring of
instrument temperatures and flow rates, operation of autoinjectors, acquisition of
A
E
B
D
C
Pesticide Analytical Manual Vol. ISECTION 504
504–4
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
retention times and detector responses, and interpretation of those values for
residue identification and quantitation. Unattended operation of instruments is
common when automated systems are available; some systems are capable of si-
multaneous management of multiple instruments. Automated data handling is
often incorporated into computerized laboratory information management sys-
tems capable of producing both the final laboratory report and the documenta-
tion necessary for quality assurance requirements (Section 206).
Discussion of entire systems is beyond the scope of this section. The focus is
instead on measurement of peaks by electronic integration of the signal produced
by a detector, i.e., summation of the change in electronic signal per unit time.
Beyond that generalized description of the integration process, each system oper-
ates under a unique “integration algorithm” that specifies how it will choose what
part of the signal to integrate. The accuracy with which the system can measure
detector response to a particular analyte depends on the algorithm itself, on the
configuration options available to the user, and on the user’s conscientiousness in
choosing appropriate options. If an electronic integrator is properly configured,
its measurement of peaks is the fastest, most accurate, and most reproducible
available.
Major pitfalls exist, however, in the uncritical acceptance of results generated
by electronic integration. Proper configuration of the algorithm, to the extent
permitted by the system, is critical. After chromatograms have been run and re-
sults presented, review by a competent analyst is essential, because no integration
algorithm can ever handle perfectly all the variations that occur in the chromato-
graphic environment. The analyst must understand the concepts incorporated
into the algorithm, be able to interpret the visual display of the chromatogram
provided by the system, and evaluate whether integration was appropriately
performed.
Data systems that perform electronic integration vary in the amount of “memory”
available for storing data. Although simple integrators have only enough memory
to process one chromatogram at a time, computer-based systems can usually store
data associated with many chromatograms. In the latter case, when review of a
chromatogram suggests that the original integration was performed improperly,
the system can be reconfigured and a new calculation made from the stored data.
If the system lacks the memory required to permit recalculation, the sample must
be rechromatographed with the integrator reconfigured. Alternatively, the peak(s)
can be measured manually from a printed chromatogram.
Optimum quantitation accuracy with any electronic integrator is dependent on
the operator’s making complete use of options available within the integrator. The
following approach is recommended:
? Configure integrator for the GLC system. At the minimum, configure the
integrator for the particular GLC system in use, rather than operating
with default settings. Develop an integrator configuration for each GLC
system routinely used. Store integrator settings as a “program,” if the
system permits, or keep a written record if necessary.
? Optimize peak and baseline recognition by considering the typical chro-
matograms the GLC system can be expected to produce. Chromatographic
features and the conditions that determine them include: baseline noise,
varying with type of detector; expected peak widths, dependent on column
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 504–5
Pesticide Analytical Manual Vol. I SECTION 504
and conditions; and inclusion of a solvent peak, dependent on whether
or not solvent is vented. Most electronic integrators can be configured
to specify the following options:
1) Size (in whatever value the integrator generates) below which a
response is not recognized as a peak; sometimes called “area reject.”
2) Range of peak base widths within which detector response is recog-
nized as a peak.
3) Increase in baseline slope above which detector response is recog-
nized as a peak; referred to as “threshold.”
4) Appearance of multiple inflection points before the apex, used to
identify the existence of two or more peaks when no “valley” exists
between them. Some systems can classify such “shoulders” as front or
rear.
5) Slope of peak above which response is recognized as the solvent
peak; can be specified because solvent peak rises faster than most
other peaks; value depends on detector, sensitivity, and column effi-
ciency. May also permit recognition of peaks that appear on the
tailing edge of the solvent.
? Use integrator features that demonstrate its operation. Most electronic
integrators offer the option of displaying, on the chromatogram, an
indication of exactly where the measurement started and ended. Some
integrators can also be configured to show where the baseline was drawn.
The analyst should take advantage of these features by choosing the
option to print such indicators and should then use them in subsequent
comparison of integrator measurements to the chromatogram.
? Configure integrator to accommodate particular chromatograms. An
integrator configured by a pre-established program for a particular GLC
system may not measure peaks accurately if the chromatogram includes
responses to co-extractives or an unexpectedly complicated pattern of
residues. Choose other options for configuration if experience with the
commodity, method, or likely residues suggests in advance what type of
chromatogram can be expected. For example, if the chromatogram is
likely to contain isolated, symmetrical peaks on a flat, quiet baseline,
configure the integrator to match peak width selection to measured
width of the peak at half height, and set the threshold a few units below
the highest value still capable of detecting the peak. In contrast, if the
chromatogram is likely to contain peaks clustered together or with a
noisy or sloping baseline, configure the integrator to accommodate those
conditions. Table 504-a lists the effects produced when the two most
important integrator settings, peak width and threshold, are varied.
? Review integrator measurements and reconfigure for accuracy. The
analyst is ultimately responsible for accurate quantitation, so review
and evaluation of chromatograms and integrator reports are essential.
If examination reveals that the integrator inappropriately included or
excluded portion(s) of the chromatogram, the following integrator
options should be changed and the peak recalculated:
Pesticide Analytical Manual Vol. ISECTION 504
504–6
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Table 504-a: Effects of Changing Electronic Integrator Settings
Setting Result
Peak Width Threshold
High High Only major peaks detected; random noise eliminated
High Low Trace-level peaks detected; noise also recognized as peaks
Low High Peaks on sloping baselines detected; noise not detected
Low Low Narrow and broad peaks both detected (low peak width
permits recognition of narrow peaks, while low threshold
permits detection of broader peaks)
- Reposition baseline to appropriate base of the peak(s).
- Measure peak(s) appearing on top of much larger peak from baseline
constructed to represent remainder of larger peak; sometimes called
“tangent skim.”
- Identify point at which to split incompletely resolved peaks, i.e., where
to end integration of one response and start integration of the next;
sometimes called “split peak.”
- Delete one or more peaks from integration; this does not remove peak
from chromatogram.
- Integrate area within chromatogram as single number; useful when
multicomponent residues, such as toxaphene, are being measured.
504 D: SPECIAL CONSIDERATIONS FOR COMPLEX CHROMATOGRAMS
Chromatograms that display residues of multicomponent chemicals or mixtures of
two or more residues challenge the chemist to perform accurate measurement of
peak size. Quantitative accuracy is further challenged when the residue has under-
gone degradation and the pattern of peaks does not match that of the most
appropriate reference standard. The following procedures for quantitation of cer-
tain difficult residues have been developed during years of practical experience.
BHC (also known as HCH, hexachlorocyclohexane)
Technical grade BHC is a mixture of six chemically distinct isomers and one or
more heptachlorocyclohexanes and octachlorocyclohexanes [4]; as a practical
matter, the isomers α, β, γ, and δ are the only ones ever reported by FDA. The γ
isomer is also known as lindane and is marketed as a separate pesticide. Currently,
U.S. tolerances for BHC have been revoked, but residues are still found in
imported commodities; U.S. tolerances for lindane remain on several commodi-
ties.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 504–7
Pesticide Analytical Manual Vol. I SECTION 504
BHC
Isomer
Residues of BHC can be expected to vary in relative amounts of the individual
isomers for several reasons: (a) Separate use of both BHC and lindane is possible,
(b) commercial formulations vary in the percentage of individual BHC isomers
present, and (c) isomers undergo different rates of metabolism or environmental
degradation; e.g., the elimination rate of isomers fed to rats was 3 weeks for the
α, γ, and δ isomers and 14 weeks for the β isomer [5]. This difference in animal
metabolism rates explains the typical finding of β isomer as the predominant BHC
residue in dairy products.
Detector response to the same amount of different isomers may also vary. When
BHC isomers were chromatographed individually on a wide bore methyl silicone
column, relative response of an electroconductivity (halogen mode) detector
(ElCD-X) to each isomer ranged from 0.58-1.00, while
63
Ni electron capture (EC)
detector responses at the same conditions varied from 0.43-1.30 (Table 504-b).
Both detectors responded less to ?-BHC than to the other three isomers [6].
Hexachlorobenzene, an industrial chemical and impurity associated with the pes-
ticide quintozene, elutes near the BHC residues on all commonly used GLC sys-
tems. Although hexachlorobenzene has only occasionally been found in the same
sample as BHC, it is important to ascertain that it is not present before BHC
residues are quantitated. Several packed columns were once cited as capable of
separating hexachlorobenzene and the four important BHC residues from one
another [7, 8]. Among the GLC systems described in Section 302, the best choice
for separating these residues is DG18 (50% cyanopropylphenyl, 50% methyl silox-
ane column at 200° C, electron capture detector). The column of DG18 is not
compatible with ElCD-X, so DG22 (DEGS column at 180° C, ElCD-X) is recom-
mended for confirmation of BHC residues as long as β- and δ-BHC, which do not
separate, are not both present.
To quantitate BHC most accurately:
? Choose GLC system that separates residues in the sample from one
another; if possible, use a halogen-selective detector, such as ElCD-X.
? Quantitate each isomer separately against a standard of the respective
pure isomer.
Table 504-b: Response of Two Detectors to Four BHC Isomers
Ng Required for 1/2 FSD Response Relative to Lindane
63
Ni EC ElCD-X
63
Ni EC ElCD-X
Alpha 0.24 0.41 1.30 0.71
Beta 0.72 0.50 0.43 0.58
Gamma (lindane) 0.31 0.29 1.00 1.00
Delta 0.33 0.49 0.94 0.59
Pesticide Analytical Manual Vol. ISECTION 504
504–8
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Chlordane
Chlordane is a technical mixture of at least 11 major components and 30 or more
minor ones; Figure 504-c is a chromatogram of technical chlordane. Structures of
the many chlordane constituents have been elucidated using GLC-mass spectrom-
etry and nuclear magnetic resonance analytical techniques [9, 10]. The two major
components of technical chlordane are trans- and cis-chlordane (Figure 504-c
peaks E and F, respectively); the exact percentage of each in the technical material
is not completely defined and is inconsistent from batch to batch.
At one time, heptachlor, a component of technical chlordane, was also marketed
as a separate pesticide. When residues of heptachlor and its metabolite heptachlor
epoxide were found in the same commodity as chlordane, the source of the former
was in question. Currently, neither chlordane nor heptachlor is registered for use
on foods in the United States, and tolerances for both have been revoked. Most
residues that are now found occur in fish as a result of lingering environmental
contamination.
The GLC pattern of a chlordane residue may differ considerably from that of the
technical standard. Depending on the sample substrate and its history, residues of
chlordane can consist of almost any combination of constituents from the techni-
cal chlordane, plant and/or animal metabolites, and products of degradation
caused by exposure to environmental factors such as water and sunlight. Only
limited information is available on which GLC residue patterns are likely to occur
in which commodities (e.g., References 11 and 12), and even this information may
not be applicable to a situation where the route of exposure is unusual. For
example, fish exposed to a recent spill of technical chlordane will contain a resi-
due drastically different from a fish whose chlordane residue was accumulated
through normal food chain processes.
Figure 504-c
Technical Chlordane
Chromatogram of 1.8 ng technical chlordane, chromatographed on system 302 DG 1.
Labeled peaks are thought to represent, respectively: A, monochlorinated adduct of
pentachlorocyclopentadiene with cyclopentadiene; B, co-elution of heptachlor and alpha
chlordene; C, co-elution of beta chlordene and gamma chlordene; D, a chlordane analog;
E, trans-chlordane; F, cis-chlordane; G, trans-nonachlor; H, co-elution of cis-nonachlor and
“Compound K,” a chlordane isomer.
A
B
C
D
E
F
H
G
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 504–9
Pesticide Analytical Manual Vol. I SECTION 504
Because of this inability to predict a chlordane GLC residue pattern, no single
method can be described for quantitating chlordane residues. The analyst must
judge whether or not the residue’s GLC pattern is sufficiently similar to that of a
technical chlordane reference standard to use the latter as a reference standard
for quantitation, then:
? When the chlordane residue does not resemble technical chlordane,
but instead consists primarily of individual, identifiable peaks, quantitate
each peak separately against the appropriate reference standard. Refer-
ence standards are available for at least 11 chlordane constituents, me-
tabolites, or degradation products that may occur in the residue.
? When the GLC pattern of the residue resembles that of technical
chlordane, quantitate chlordane residues by comparing the total area of
the chlordane residue from peaks A through G (Figure 504-c) to the
same part of the standard chromatogram. To define appropriate mea-
surable area of chromatograms, adjust amount of extract injected so
that the major residue peaks are about 50% full scale deflection (FSD),
then inject an amount of reference standard that causes response within
±25% of that; peaks E and F in the two chromatograms should be about
the same size. Construct the baseline beneath the standard from the
beginning of peak A to the end of peak G. Use the distance from the
trough between peaks E and F to the baseline in the chromatogram of
the standard to construct the baseline in the chromatogram of the sample.
Peak H may be obscured in a sample by the presence of other pesticides. If H is
not obscured, include it in the measurement for both standard and sample. If the
heptachlor epoxide peak is relatively small, include it as part of the total chlordane
area for calculation of the residue. If heptachlor and/or heptachlor epoxide are
much out of proportion, as in Figure 504-d, calculate these separately and subtract
Figure 504-d
Chlordane, Heptachlor, Heptachlor Epoxide
Chromatogram of 1.8 ng technical chlordane, 0.1 ng heptachlor, and 0.3 ng heptachlor
epoxide, superimposed on chromatogram of technical chlordane only; system 302 DG1.
Heptachlor
Heptachlor epoxide
Pesticide Analytical Manual Vol. ISECTION 504
504–10
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
their areas from total area to give a corrected chlordane area. (Note that octachlor
epoxide, a metabolite of chlordane, can easily be mistaken for heptachlor epoxide
on a nonpolar GLC column.)
(When measurement of total peak area by integration was compared to addition
of peak heights for quantitation of chlordane residue in several samples, the re-
sults of the two techniques were reasonably close; results justify the use of the
“peak height addition” technique for calculating total chlordane when no means
of measuring total area is available. To quantitate by peak height addition, mea-
sure heights of peaks A, B, C, D, E, F, and G, in mm, from peak maximum of each
to the baseline constructed under the total chlordane area, then add heights. The
technique has inherent difficulties because not all the peaks are symmetrical and
not all are present in the same ratio in the standard and in the sample.)
PCBs
Polychlorinated biphenyls (PCBs) were manufactured for many years in the United
States by the Monsanto Co. and marketed under the trade name Aroclor. Each
Aroclor product was a mixture of chlorobiphenyl congeners into which 1-10 chlo-
rine atoms were substituted; 209 different congeners were possible. Common
commercial products included Aroclor 1221, 1242, 1248, 1254, 1260, and 1262,
with the last two digits in the name indicating the average percent chlorination in
the particular mixture; Aroclor 1016, purportedly a purified version of Aroclor
1242, was also marketed. Aroclors are no longer used or marketed in the United
States, but their residues remain in the environment, in foods like fish and shell-
fish, in animals, and in human tissue.
GLC chromatograms of PCB residues contain many peaks, and patterns vary con-
siderably, because residues can result from any combination of Aroclor mixtures.
Variations in residue patterns are also caused by degradation from weathering or
metabolism. Different congeners vary in the degree to which they are excreted by
or retained within an animal and by the degree to which they volatilize. This
multiplicity of potential PCB residue patterns makes the task of identifying and
quantitating residues extremely challenging. The presence in the extract of resi-
dues from chlorinated hydrocarbon pesticides further complicates the determina-
tion. Residues of p,p'-DDE are most likely to interfere in determination of PCBs,
because both residues are often present in the same commodity and because their
structural and behavioral similarities make them difficult to separate with normal
analytical methodology.
Quantitation of PCB residues is best achieved by following these steps:
? Isolate PCB residues from sample co-extractives and from other residues
to the degree possible before GLC determination. Certain cleanup step
options in Chapter 3 methods are designed to separate PCBs from pesti-
cide residues of similar structure; these options should be used to analyze
any commodity in which PCB residues are likely to occur, especially fish
and shellfish.
? Select the reference standard that most closely resembles the residue pat-
tern. A single Aroclor or, more often, a mixture of Aroclors that produce
the most similar pattern is used for quantitation. Judgment must be made
about what proportion of different Aroclors should be combined to pro-
duce the appropriate reference standard.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 504–11
Pesticide Analytical Manual Vol. I SECTION 504
? Use a GLC system that separates peaks efficiently. Packed or capillary
columns may be used; wide bore capillary columns provide the best
compromise of speed and efficiency.
? Choose from the following quantitation options the one that best suits
the residue pattern. Both have been successfully collaborated in
interlaboratory tests [13, 14]; choice depends on the degree to which
the residue and reference standard match:
1) When PCB residue pattern closely resembles that of a single Aroclor
or mixture of Aroclors, quantitate by comparing total area or height
of residue peaks to total area or height of peaks from appropriate
reference standard(s). Measure total area or height response from
common baseline under all peaks. Use only those peaks in the resi-
due that can be attributed to chlorobiphenyls. These peaks must also
be present in chromatogram of reference standards.
2) When PCB residue pattern is significantly different from that of any
Aroclor or mixture of Aroclors, quantitate by comparing area of each
peak in residue to peak at same retention time in a specially cali-
brated lot of Aroclor reference standard (Table 504-c). To each peak
thus measured, apply weight factor associated with that peak in par-
ticular reference standard, as listed in Table 504-c. Sum individual
peak values to obtain total ppm PCB. This option can also be used
when residue and reference standard chromatographic patterns match.
The special Aroclor reference standards were calibrated using the
separations effected by packed column chromatography, but the
weight factors are also valid with chromatography on the equivalent
wide bore capillary column operated in the packed column mode
(Section 502 C).
Other quantitation techniques are sometimes used. One system makes use of
capillary column chromatography, capable of separating most PCB congeners from
one another, and a precalibrated reference standard mixture for which identity
and weight percent of each congener have been established [15]. This “individual
congener” capillary column method is significantly more time-consuming than
measurement of individual peak areas or heights obtained by packed column
GLC, and results from the two approaches are not significantly different when
total PCBs are calculated [16, 17]. Several European countries use variations of
the individual congener method by measuring, in sample and standard, only se-
lected peaks [18]; in these countries, legal limits on PCB residues are defined in
terms of results from the established quantitation method.
Accurate quantitation of both p,p'-DDE and PCBs in the same sample is
possible only when chromatographed on a narrow bore capillary column.
Quantitation of only the PCB residue, when p,p'-DDE is present, can be accom-
plished by first eliminating p,p'-DDE with derivatization and column chromatog-
raphy [19].
Figure 504-e, a chromatogram of PCB residues isolated from chinook salmon,
demonstrates the challenge of PCB determinations. Quantitation was performed
by comparison to a mixture of Aroclors 1254 and 1260.
Pesticide Analytical Manual Vol. ISECTION 504
504–12
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Table 504-c: Weight Percent Factors for Individual Gas
Chromatographic Peaks in Aroclor Reference Standards
Aroclor
1016 1242 1248 1254 1260
R
DDE
(100x)
1
(77-029)
2
(71-696)
2
(71-697)
2
(71-698)
2
(71-699)
2
11 0.2
16 3.8 3.4 0.3
21 8.1 10.3 1.1
24 1.2 1.1 0.2
28 16.8 15.8 6.0
32 7.6 7.3 2.6
37 18.5 17.0 8.7
40 14.6 13.0 7.4
47 11.6 9.9 15.7 7.1
54 7.7 7.1 9.3 2.7
58 6.4 4.4 8.3 1.2
70 3.4 8.7 18.2 14.7 2.4
78 1.9 6.4
84 4.6 18.6 3.6
98 3.4 8.3
2.8
104 3.3 14.1
112 1.0
117 4.4
125 2.3 15.6 11.0
146 1.2 9.0 13.3
160 5.5
174 7.4 10.0
203 1.3 10.9
232-244 11.2
280 12.5
332 4.2
360-372 5.4
448 0.8
528 2.0
1
Peaks are identified by their retention times relative to p,p′-DDE=100 at conditions described in Section 302 DG1.
2
Food and Drug Administration Lot Nos. (Weight factors are valid only for these FDA Lot Nos.) Aroclor reference
standards are available from Food and Drug Administration, Division of Pesticides and Industrial Chemicals, HFS-337,
200 C Street SW, Washington, DC 20204. Aroclor 1016 (77-029) was referred to as KB-06-256 in J. Assoc. Off. Anal. Chem.
(1978) 61, 272-281.
}
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 504–13
Pesticide Analytical Manual Vol. I SECTION 504
p,p¢ -DDE
trans-Nonachlor
Hexachloraobenzene
010203040
Figure 504-e
PCBs in Chinook Salmon
1.23 ng Aroclor 1254 and 0.745 ng Aroclor 1260, chromatographed on same
system as above.
Hexachlorobenzene
′
25 mg chinook salmon, extracted and cleaned up by method 304 E1+C3,
chromatographed on system 302 DG1, with detector sensitivity greater than
normal to permit measurement of low levels. Injection represents petroleum
ether eluate from Florisil, which separates PCB residues from most but not all
pesticides. Total PCB is 0.087 ppm, calculated using total area measurement,
0.090 using factors of Table 504-c; comparison is to mixed Aroclor standard
below. Pesticides were identified but not quantitated.
010203040
Pesticide Analytical Manual Vol. ISECTION 504
504–14
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Toxaphene
Toxaphene is a complex mixture that results from the chlorination of camphene.
The GLC chromatographic pattern for toxaphene does not display any individual
peaks to simplify quantitation, but instead appears as a series of incompletely
resolved peaks (Figure 504-f). Presence of other residues in the same extract
as toxaphene requires estimation of baseline placement for quantitation. Reason-
able accuracy is possible, but no truly quantitative technique has been developed.
To quantitate residues of toxaphene:
? Adjust amount of sample injected so that major residue peaks are 10-30%
FSD.
? Inject amount of reference standard that causes response within ±25% of
that of residue.
? Construct baseline for standard toxaphene between its extremities.
? Construct baseline under residue peaks, using distances of peak troughs
to baseline on standard chromatogram as guide.
? Measure areas above baseline in sample and standard chromatograms for
calculating level of residue. Relative heights and widths of matching peaks
in the residue and reference standard will probably differ.
Figure 504-f
Toxaphene
3010
Chromatogram of 11.4 ng toxaphene, chromatographed on system 302 DG1.
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 504–15
Pesticide Analytical Manual Vol. I SECTION 504
References
[1] Gaul, J.A. (1966) J. Assoc. Off. Anal. Chem. 49, 389-399
[2] Condal-Bosch, L. (1964) J. Chem. Educ. 41, A235, passim
[3] Delaney, M.F. (Aug. 1981) “Manual Chromatographic Peak Area Measurement Using
the Condal-Bosch Trapezoidal Method,” LIB 2529, FDA, Rockville, MD
[4] Metcalf, R.L. (1955) Organic Insecticides: Their Chemistry and Mode of Action, Interscience
Publishers, Inc., New York
[5] Lehman, A.J. Summaries of Pesticide Toxicity, FDA, Washington, DC
[6] Parfitt, C.H. (Feb. 1993) private communication, FDA, Division of Pesticides and
Industrial Chemicals
[7] Laski, R.R. (Aug. 1969) “GLC Determination of BHC,” LIB 946, FDA, Rockville, MD
[8] Power, J.L., and Kenner, C.T. (May 1972) “GLC Determination of BHC Isomers and
HCB,” LIB 1449, FDA, Rockville, MD
[9] Cochrane, W.P., and Greenhalgh, R. (1976) J. Assoc. Off. Anal. Chem. 59, 696-702
[10] Sovocool, G.W., et al. (1977) Anal. Chem. 49, 734-740
[11] Lawrence, J.H., et al. (1970) J. Assoc. Off. Anal. Chem. 53, 261-262
[12] Zitko, V. (1978) Chemosphere 7, 3-7
[13] Sawyer, L.D. (1973) J. Assoc. Off. Anal. Chem. 56, 1015-1023
[14] Sawyer, L.D. (1978) J. Assoc. Off. Anal. Chem. 61, 272-281; 282-291
[15] Mullins, M. (June 1985) private communication, EPA Large Lakes Laboratory, Grosse
Ile, MI
[16] Maack, L., and Sonzogni, W.C. (1988) Arch. Environ. Contam. Toxicol. 17, 711-719
[17] Trotter, W.J. (1993) private communication, publication in preparation, FDA, Divi-
sion of Pesticides and Industrial Chemicals
[18] Maybury, R.B. (1986) Report of Codex Committee on Pesticide Residues Ad Hoc
Working Group on Contaminants, CX/PR 86/13 Appendix 3
[19] Trotter, W.J. (1975) J. Assoc. Off. Anal. Chem. 58, 461-465
Pesticide Analytical Manual Vol. ISECTION 504
504–16
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)
Pesticide Analytical Manual Vol. I
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92) 505–1
SECTION 505
505: BIBLIOGRAPHY
GENERAL TEXTS
Ettre, L.S. (1973) Practical Gas Chromatography, Perkin-Elmer Corp., Norwalk, CT
Grob, R.L. (1985) Modern Practice of Gas Chromatography, Wiley, New York
Hyver, K., ed. (1989) High Resolution Gas Chromatography, Hewlett-Packard Co.,
Wilmington, DE
Jennings, W. (1987) Analytical Gas Chromatography, Academic Press, Orlando, FL
Hargis, L.G. (1988) Analytical Chemistry Principles and Techniques, Prentice-Hall, New
York
Poole, C., and Poole, S. (1991) Chromatography Today, Elsevier, New York
Rood, D. (1991) A Practical Guide to the Care, Maintenance, and Troubleshooting of
Capillary Gas Chromatographic Systems, Huethig Buch Verlag, Heidelberg, Germany;
available through J&W Scientific, Folsom, CA
INLETS
Klee, M.S. (1991) GC Inlets—An Introduction, Hewlett-Packard Co., Wilmington, DE
COLUMNS
Jennings, W. (1980) Gas Chromatography with Glass Capillary Columns, 2nd ed.,
Academic Press, Orlando, FL
DETECTORS
Buffington, R., and Wilson, M.K. (1991) Detectors for Gas Chromatography—A Practical
Primer, Hewlett-Packard Co., Wilmington, DE
Hill, H.H., and McMinn, D.G., ed. (1992) Detectors for Capillary Chromatography,
Chemical Analysis Series Vol. 121, Wiley, New York
Pesticide Analytical Manual Vol. ISECTION 505
505–2
Transmittal No. 94-1 (1/94)
Form FDA 2905a (6/92)